User guide

4-2 Basic Settings and Advanced Settings
4-14
4
Setting Various Functions
IL-E
According to the judgment value (P.V.), set the output state (N.O./N.C.) of judgment output
ON/OFF and edge check output.
There are three judgment outputs as below.
HIGH judgment output (Black wire)
GO judgment output (Gray wire)
LOW judgment output (White wire)
The judgment output is turned ON/OFF as below according to the tolerance setting value.
“3-4 Setting the Tolerance Setting Value” (page 3-10)
When Normally Open is set
When Normally Closed is set
*1 "Error Displays and Corrective Actions" (page A-4)
*2 When the judgment value (P.V.) is "-----".
5. Output State
QWV
Item Setting range Default value
Output state no (Normally Open), nC (Normally Closed) no
Judgment
Judgment output
HIGH GO LOW
HIGH ON OFF OFF
GO OFF ON OFF
LOW OFF OFF ON
Error
*1
ON OFF ON
"-----"
*2
OFF OFF OFF
Judgment
Judgment output
HIGH GO LOW
HIGH OFF ON ON
GO ON OFF ON
LOW ONONOFF
Error
*1
OFF ON OFF
"-----"
*2
ON ON ON
Reference
Regardless of the output state setting, the judgment indicator on the sensor
amplifier interlocks the judgment output when Normally Open is set.
“3-4 Setting the Tolerance Setting Value” (page 3-10)
The following is the default state of the judgment indicator. The lighting
status can be changed in the items for "18. Display Color".
"18. Display Color" (page 4-41)
Judgment
Judgment output
HIGH GO LOW
HIGH Red OFF OFF
GO OFF Green OFF
LOW OFF OFF Red
Error Red OFF Red
"-----" OFF OFF OFF