4-2 Basic Settings and Advanced Settings
4-27
IL-E
4
Setting Various Functions
The analog output scaling for the judgment value (P.V.) is set.
If the analog output is set to no output, this screen is skipped.
“3-1 Operation When the Power is Turned on for the First Time” (page 3-2)
“3-3 Initial Reset (Initialize)” (page 3-8)
Relation of voltage and analog output in default state (dEFLt)
The upper and lower limits for the different selections of analog outputs are as follows.
10. Analog Output Scaling
#P)
Analog output Description
Default
value
Default setting
Scaling is not performed.
c
Free range
Scaling is performed. The analog output range can be
changed by specifying the upper limit and lower limit of the
judgment value (P.V.).
Bank
A different scaling is provided for each bank. The analog
output range can be changed by designating an upper
limit and lower limit for each bank.
Example 1: Lower limit < Upper limit
(Analog output 1 to 5 V)
Example 2: Lower limit > Upper limit
(Analog output 1 to 5 V)
Reference
If Upper limit = Lower limit is set, the analog output is output as below.
Judgment value (P.V.)
≤
Upper limit: Analog output lower limit (When 1 to 5 V
→
1 V)
Judgment value (P.V.)
>
Upper limit: Analog output upper limit (When 1 to 5 V
→
5 V)
Lower limit Upper limit
5
1
Output upper
limit
Output lower
limit
Judgment
value
(P.V.)
(V)
(V)
5
1
Lower limit Upper limit
Judgment
value
(P.V.)
Output upper
limit
Output lower
limit
Model Upper limit Lower limit
IL-S025 5.000 - 5.000
IL-030 5.000 - 5.000
IL-S065/IL-065 10.000 - 10.000
IL-100 20.000 - 20.000
IL-300 140.00 - 140.00
IL-600 400.00 - 400.00
IL-2000 1000.0 -1000.0
Analog setting Output upper limit Output lower limit
OFF Not output
0-5v 5 V 0 V
-5-5v 5 V - 5 V
1-5v 5 V 1 V
AMPr 20 mA 4 mA