Instruction Manual
3-19
- IV Series User’s Manual (Field Network) -
Data allocation in the cyclic communication
Output Assembly (PLC → IV series)
Output Assembly are devices that write instructions from the PLC to the IV series. 
The device map of the data allocated for the Output Assembly is as shown below. 
These parameters are responsible for the control instructions for the IV series, clearing of warnings and 
handshake control. 
For details of each parameter, refer to   “Output Assembly parameter details” (Page 3-20). 
Address Bit7 Bit6 Bit5 Bit4 Bit3 Bit2 Bit1 Bit0
0 Reserved by system
Buffer 
clear 
request
Statistics 
reset 
request
Warning 
clear 
request
Program 
switching 
request
Master 
registration 
request
Trigger 
request
1 Reserved by system
2 Reserved by system
Result 
acquisition 
complete 
notication
3 Reserved by system
4 - 5 Program No.
6 - 11 Reserved by system
3
Cyclic communication