Instruction Manual
3-22
- IV Series User’s Manual (Field Network) -
Data allocation in the cyclic communication
Error code list
The following shows the list of error codes that are generated in the IV series. 
Error 
code
Content Cause Countermeasure
0 No error --- ---
1 - 32
Program No. xx 
corruption error
A data error has 
occurred in program No. 
xx. 
The data corruption may 
have occurred due to a 
power-off while writing 
settings data and/or due 
to noise. 
Initialize the program No. xx. 
Switch on the power back ON. 
Do not switch off the unit while the 
settings are being saved. 
If the error persists, contact your 
nearest KEYENCE ofce. 
75
Non-volatile memory 
error
(EEPROM)
A data error has 
occurred. 
The data corruption may 
have occurred due to a 
power-off while writing 
settings data and/or due 
to noise. 
Initialize the settings of this unit. 
Switch on the power back ON. 
Do not switch off the unit while the 
settings are being saved. 
If the error persists, contact your 
nearest KEYENCE ofce. 
76
Non-volatile memory 
error
(FLASHROM)
81-128 System error
An error may have 
occurred in the sensor. 
Switch on the power back ON. 
If the error persists, contact your 
nearest KEYENCE ofce. 
If two or more errors are occurring at the same time, the error code of the highest priority error will be 
displayed. 
Greater the error code, higher the priority of the error will be. 
3
Cyclic communication