I/O Terminal
E LT-9001-IM
7-3
The signal names and descriptions of each terminal
The signal names and descriptions of each terminal are described as follows.
Refer to "Functions of the I/O Signal" (page 7-5) for the detailed functions.
24V DC (–) and COM1 are common via a choke coil. Be careful not to generate
a potential difference.
0 V for the analog monitor output is isolated from COM1. It is common to the FG
terminal for the AC power supply.
24V DC terminal is a dedicated power supply terminal for the monitor specified
by KEYENCE ONLY. Do not use to power other external devices.
The control I/O can be selected from the binary output of the measured value or
the tolerance judgment output depending on the setting.
Refer to "I/O" (page 5-20) for the setting details.
Matching connector specifications
HIROSE Electric HIF3B Series or equivalent
The pin layout is as follows.
Control I/O
Note
Terminal number Signal name Description
1 MON1 Analog monitor output1
2 MON2 Analog monitor output2
3 0V 0 V for analog monitor output
4 ZERO Auto-zero input
5 RESET Reset input
6 TIMING Timing input
7 COM1 COM for 4 to 6
8 24V DC (–) Power (–) terminal for the monitor (CA-MN80)
9 24V DC (+) Power (+) terminal for the monitor (CA-MN80)
17 16 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1
34 33 32 31 30 29 28 27 26 25 24 23 22 21 20 19 18
Terminal number