96129E Digital Laser Sensor LV-N10 Series User's Manual Read this manual before use. Keep this manual in a safe place for future reference.
Introduction This manual describes the basic operations and hardware functions of the LV-N10 Series. Read this manual carefully to ensure safe performance and function of the LV-N10 Series. Keep this manual in a safe place for future reference. Ensure that the end user of this product receives this manual. Symbols The following symbols are used in this manual to alert you to matters concerning the prevention of injury and product damage. Always read these sections.
Safety Precautions General Precautions WARNING • This product is only intended to detect objects. Do not use this product for the purpose of protecting a human body or a part of the human body. • This product is not intended for use as an explosion-proof product. Do not use this product in a hazardous location and/or potentially explosive atmosphere. • This product uses DC power. Do not apply AC power. The product may explode or burn if an AC voltage is applied.
Safety Precautions * The laser classification for FDA (CDRH) is implemented based on IEC60825-1 in accordance with the requirements of Laser Notice No.50. Reference Laser Transmission OFF input When the laser transmission OFF input is set for external input, the laser transmission can be stopped by turning the external input ON (2 ms or longer). The transmission is stopped while the external input is ON. Once the external input is turned OFF, the laser transmission will resume within 20 ms.
Manual Organization 1 Before Using Outlines the package contents and identifies part names and functions. 2 Installation and Connection Provides procedures for installing sensor amplifiers and cables, as well as operating precautions. 3 Basic Operation Explains basic instructions for operating and setting the sensor amplifiers. 4 Settings for Advanced Functions Describes settings for advanced functions of the LV-N10 Series.
Table of Contents Safety Precautions............................................................................... 1 General Precautions .......................................................................... 1 Safety Precautions on Laser Product................................................. 1 Precautions on Regulations and Standards ....................................... 2 Manual Organization ........................................................................... 3 Table of Contents ..........
Table of Contents 3-4 3-5 3-6 3-7 3-8 Preset display mode (only when LV-N100/NH110/NH300 is connected)................................................................................. 3-11 Two-point Calibration..................................................................... 3-12 Maximum Sensitivity Calibration ................................................... 3-13 Full Auto Calibration ...................................................................... 3-14 Positioning Calibration.............
Table of Contents 4-7 Detection Mode for Output2 .......................................................... 4-33 Output Timer for Output2 .............................................................. 4-37 Settings Save/Recall .......................................................................... 4-38 Custom Save (Settings Save) ....................................................... 4-38 User Reset (Settings Recall) .........................................................
Before Using This chapter outlines the package contents and identifies part names and functions. 1-1 1-2 1 Checking the Package Contents ........................ 1-2 Part Names ........................................................
1-1 Before using the LV-N10 Series, make sure that the following equipment and accessories are included in the package. We have thoroughly inspected the package contents before shipment. However, in the event of missing, defective or broken items, contact your nearest KEYENCE office.
1-1 Checking the Package Contents LV-NH300* LV-S31 Transmitter (T) Receiver (R) Operation indicator (red) Mounting bracket set Transmitter F J F Receiver Adjustment screwdriver N N Mounting bracket × 1 Plate nut × 1 M3 × 15 screw × 2 Adjustment trimmer LV-S61 LV-S41/S41L Receiver Transmitter Operation indicator (red) Mounting bracket × 1 Operation indicator (red) Receiver Transmitter LV-S62 Transmitter/ Receiver Reflector Mounting (R-6) × 1 bracket set Mounting bracket × 1 Operation P
1-2 1 Part Names Sensor Amplifier Display/control unit Before Using Hold lock lever Hook Expansion connector*1 Expansion connector*1*2 Dust cover M8 connector (LV-N11Cغ/N12C)غ Connection cable (LV-N11غ/N12غ/N11MN) w/o cable (LV-N10) *1 When shipped from the factory, the expansion cover is installed. *2 This is not installed on the main unit (LV-N11N/N11P/N11CN/N11CP/N11MN).
1-2 Part Names Display/control unit (1)-1(10) (1)-2 (4) (5) (6)-1 1 (7) * Before Using (3) (2) (8) (9) (6)-2 * * Not available for the LV-N10. Item Description Operation indicator Indicates the current output (detection) status of chan(2-output type) nels 1 and 2 separately. Operation indicator (1)-2 Indicates the current output (detection) status. (1-output type) Used when setting sensitivity, etc.
1-2 Part Names MEMO 1 Before Using 1-6 - Digital Laser Sensor LV-N10 Series User's Manual -
Installation and Connection This chapter provides procedures for installing sensor amplifiers and cables, as well as operating precautions. 2-1 2-2 2-3 2 Installing Sensor Amplifiers................................. 2-2 Connecting the Sensor Amplifier to the Sensor Head..... 2-5 Mounting and Adjusting the Sensor Head..........
2-1 Installing Sensor Amplifiers Mounting the Sensor Amplifier Mounting on a DIN rail 2 1 Installation and Connection 2 Align the claw at the bottom of the main body with the DIN rail, as shown on the right. While pushing the main body in the direction of the arrow (1), push down in the direction of arrow (2). (3) (2) (1) To release the amplifier, raise the amplifier body in the direction of arrow (3) while pushing in the direction of arrow (1).
2-1 Installing Sensor Amplifiers Point 1 Remove the protection covers from the main unit and expansion unit(s). 2 Mount the main unit and expansion unit(s) on the DIN rail. (2) (1) 3 4 5 Slide the main unit and expansion unit(s) together. Engage the 2 claws of the expansion unit with the recesses on the main unit side until you hear/ feel a click. Attach the separately sold end units (OP-26751: a set of 2 units) to the DIN rail in the same manner as step (2).
2-1 Installing Sensor Amplifiers Wiring Diagrams for Sensor Amplifiers Point 2 • Be sure to turn off the power before wiring. • Insulate each input or output cable that will not be used.
2-2 Connecting the Sensor Amplifier to the Sensor Head Connecting the Sensor Amplifier to the Connector 1 Open the dust cover, and move the hold lock lever down. 2 Lift the hook up, and insert the connector completely. 3 Lower the hook to the position shown in the drawing, and pull up the hold lock lever. Point Installation and Connection 2 • When connecting to a different sensor head model, the settings for the previously connected model will be saved.
2-3 Mounting and Adjusting the Sensor Head LV-NH32/NH35/NH37/NH42/NH62/S63 2 Use the included mounting bracket. The tightening torques are shown below. LV-NH32/NH35/NH37/NH42/NH62: 0.3 N·m LV-S63: 0.6 N·m Installation and Connection LV-NH32 spot diameter adjustment Turn the focus ring to adjust the size of the spot diameter. After adjusting, tighten fix the focus lock screw. LV-NH100/NH110/NH300 Using the following optional brackets to mount the unit.
2-3 Mounting and Adjusting the Sensor Head 2 Adjust the beam axis. (a) (1) Adjust the beam axis upward by tightening the (a) screw, and downward by loosening the screw. Fix the angle setting with the (b) nut. (b) 2 (2) Adjust the horizontal angle by loosening the fixing section (c). (c) Installation and Connection LV-S31 1 Mount using the Included mounting bracket. F J N 2 Place a workpiece at the position to be set as the detection center.
2-3 Mounting and Adjusting the Sensor Head LV-S41/S41L Always use the included mounting bracket. 2 1 Installation and Connection 2 Attach the mounting bracket as shown on the right. Match section A of the fitting onto section B of the sensor head A B Attach onto a flat surface with M3 screws as shown on the right. (M3 screws are not included.) LV-S62 Mount with the optional L-mounting bracket (OP-84350), back surface mounting bracket (OP-84349) or horizontal mounting bracket (OP-84351).
2-3 Mounting and Adjusting the Sensor Head Back surface mounting bracket (optional: OP-84349) Included: Mounting bracket × 1 / M3 × 18 screw × 2 1 2 The tightening torque is 0.5 N·m or less. Adjust the beam axis. (1) The beam axis angle can be adjusted sideways by loosening screws (a), (b), and (c) and moving the bracket left and right. Always tighten screw (b) first. (2) The beam axis can be adjusted in the downward direction by tightening the bottom screw as shown with the arrow.
2-3 Mounting and Adjusting the Sensor Head LV-S71/S72 1 2 Installation and Connection 2 Attach so that the side on which the T (transmitter) and R (receiver) are printed faces upward. The operation indicator lights are on the printed side. Fixing nut Axis beam adjustment screw The tightening torque is 1.2 N·m or less. Adjust the beam axis. The beam axis can be adjusted in the downward direction by tightening the screw as shown with the arrow.
Basic Operation This chapter explains basic instructions for operating and setting the sensor amplifier. 3-1 3-2 3-3 3-4 3-5 3-6 3-7 3-8 3 Quick Reference................................................. 3-2 Switching Output ................................................ 3-4 Adjusting the Sensitivity ............................................... 3-5 Setting the Current Received Light Intensity to 0 (Zero Shift) .. 3-17 Loading the Recommended Settings (Recipe Function) ... 3-19 Initialization........
3-1 Quick Reference The main setting operations are explained according to purpose. Refer to Chapter 4 for information on advanced function settings and explanations not given below. Operation indicator (1-output type) Operation indicator (2-output type) 3 PST indicator [SET] button Basic Operation Purpose Changing the output Adjusting the sensitivity and integrating the display to "100.0" and ".0" Shifting the received light intensity to .
3-1 Quick Reference Operation procedures Reference page Loading the recommended settings 15 Load the recommended settings. (Recipe function) 1. Press and hold the [SET] button and [PRESET] button. 2. Display the LoAd screen with the ( ) button, and press the [MODE] button. 3. Select the recipe such as r-1 FALL with the ( ) button. 4. Press the [MODE] button to execute. 3-19 Initializing the settings 16 Initializing (Restore to factory default settings) 1.
3-2 Switching Output Output Switch (L-on/D-on) This function configures when the output turns ON. 1 3 When the current received light intensity is displayed, press the [MODE] button. The current output condition (L-on or Don) is displayed.*1 Basic Operation 2 Press the button to switch the output condition, and then press the [MODE] button. Select "D-on" if you want to output the ON signal when the beam is blocked (a workpiece is present.) for a thrubeam or retro-reflective model.
3-3 Adjusting the Sensitivity In this manual, the value at which the sensor amplifier's ON/OFF output switches is expressed as the "setting value". The process of adjusting the setting value is expressed as "adjusting the sensitivity". This section explains the method of adjusting the sensitivity. List of Sensitivity Adjustment Methods The methods of adjusting the LV-N10 series sensitivity is largely classified into 2 types. (2) Calibration The sensitivity can be adjusted with simple operations.
3-3 Adjusting the Sensitivity Preset Function Enabling the preset function When the PST indicator is not lit, press the [PRESET] button 3 The PST indicator lights in green. The current value is set to "100.0" and the setting value is set to "50.0". Basic Operation Green PST lights up Workpiece Setting value is " . " Current value is " . " Disabling the preset function When the PST indicator is lit, press and hold the [PRESET] button.
3-3 Adjusting the Sensitivity Work-Preset Function This function calibrates the current value to ".0". After the preset function has been executed with "100.0" displayed, and then executing this function with ".0" displayed, 2 random set points can be calibrated to "100.0" and ".0". Enabling the work-preset function Impotant The work-preset function can be used with the preset function (when preset is enabled). The received light intensity at that point is set to ".0". The value set to "100.
3-3 Adjusting the Sensitivity Maximum Sensitivity Preset Function This function calibrates the reference state to ".0" and the state at which the received light intensity is slightly higher as "100.0". This function is useful with the reflective model to detect while using the background as a reference.
3-3 Adjusting the Sensitivity Full Auto Preset Function This function automatically judges 2 states (workpiece presence/absence, etc.), and calibrates the current value to "100.0" and ".0". This is useful when the detector is moving at high speed, etc. Enabling the full auto preset function When the PST indicator is OFF, continue pressing the [PRESET] button until "Auto" flashes while the workpiece passes through. 2 After the workpiece has completely passed through, release the [PRESET] button.
3-3 Adjusting the Sensitivity Disabling the full auto preset function When the PST indicator is lit, press and hold the [PRESET] button. • The PST indicator turns off, indicating that the full auto preset function has been disabled. Point 3 Each preset function cannot be used when the following functions are set. Disable the function or change the setting before executing the preset function.
3-3 Adjusting the Sensitivity Preset display mode (only when LV-N100/NH110/NH300 is connected) The "100.0" (default value) displayed when preset is executed can be changed to "1000" or "3000" * according to the area width. The preset display mode "ArEA" is convenient for displaying the amount of light blocked in the area, etc. * "1000" is displayed when LV-N100/NH110 is connected, and "3000" when LV-NH300 is connected.
3-3 Adjusting the Sensitivity Two-point Calibration Two-point calibration is the most basic method of sensitivity setting. The setting value can be established by simply pressing the [SET] button once each when a workpiece is present. and when it is absent. 1 Press the [SET] button once when no workpiece is present. 2 Press the [SET] button once when a workpiece is present. Calibration is complete after the setting value flashes momentarily, and then stops (lights up).
3-3 Adjusting the Sensitivity Maximum Sensitivity Calibration This sensitivity setting method is useful if the received light intensity is reduced by dust or dirt. The setting value is calibrated to be slightly than the light intensity received when it is determined. 1 When a workpiece is absent for the reflective model, or when a workpiece is present for the thrubeam model/ retro-reflective model, press and hold the [SET] button for 3 seconds or more. Release the button when "SEt" flashes.
3-3 Adjusting the Sensitivity Full Auto Calibration This method automatically sets sensitivity using a moving workpiece. Sensitivity can be easily set by passing a workpiece without shutting down operating equipment. 1 While the workpiece passes through, press and hold the [SET] button until "SEt" flashes. 2 After the workpiece has completely passed through, release the [SET] button. Calibration is complete after the setting value flashes momentarily, and then stops (lights up).
3-3 Adjusting the Sensitivity Positioning Calibration This method is used when you want to position a workpiece. 1 Press the [SET] button once when no workpiece is present. 3 Position a workpiece such that its edge aligns with the center of the projecting beam. Then, press and hold the [SET] button for 3 seconds or more. Release the button when "SEt" flashes. Calibration is complete after the setting value flashes momentarily, and then stops (lights up). Press the button to adjust the setting value.
3-3 Adjusting the Sensitivity Other Calibration Methods Percentage calibration A setting value is established with a percentage in respect to the current received light intensity. When used together with external calibration input, percentage calibration can be performed from an external device, such as a PLC, allowing highly accurate detection of transparent workpieces and small workpieces, etc. 3 Refer to "Sensitivity Setting" (page 4-4) for details.
3-4 Setting the Current Received Light Intensity to 0 (Zero Shift) Zero Shift Function This function adjusts the current received light intensity display to "0". It is primarily used with reflective sensor heads. If the received light intensity cannot be set to "0" when a reflective model is installed, this function can be used to set the received light intensity to "0" with no workpiece present. This makes the difference in received light intensity easier to distinguish.
3-4 Setting the Current Received Light Intensity to 0 (Zero Shift) Point Reference The zero shift function cannot be used in combination with the preset function. To use the zero shift function, make sure that the preset function is disabled (the PST indicator is not lit). Zero shift of the expansion unit can be set from the main unit.
3-5 Loading the Recommended Settings (Recipe Function) Recommended settings (recipes) for each sensor head and application are preregistered in the LV-N10 Series. Selecting Recipe Point Press and hold the [SET] and [PRESET] buttons simultaneously for 3 seconds or more. The "rSt" display flashes. 2 Press the button to display "LoAd", and press the [MODE] button. 3 Press the button to select a desired recipe, and press the [MODE] button.
3-5 Loading the Recommended Settings (Recipe Function) List of Recipes and Recommended Sensor Heads Recipe 3 Fall (r-1 FALL) Basic Operation Percentage calibration (r-2 SEtP) Zero-Shift (r-3 0SEt) MEGA (r-4 MEGA) AREA (r-5 ArEA) 0 DATUM (r-6 0dtM) Reference 3-20 Recommended head Description To detect a falling workpiece. The falling edge of the received light intensity is Thrubeam model detected, and output as one-shot.
3-6 Initialization Initialization of Settings (Reset to Initial Values) The sensor amplifier can be reset to the factory default settings. 1 Press and hold the [SET] and [PRESET] buttons simultaneously for 3 seconds or more. The "rSt" display flashes. Press the [MODE] button. 3 Press the 4 Press the [MODE] button. Basic Operation 2 3 button to display "init". After the settings are initialized, the screen displays "oK", which is then replaced with the current received light intensity.
3-7 Locking in MEGA Mode MEGA Mode Lock (1-Output Type Only) This function is available for the 1-output type models (LV-N11C / N12C /N11MN) only. The sensor amplifier can be locked in MEGA mode, such that it always operates in MEGA mode regardless of the power mode selected in the basic setup. "Power Modes" (page 4-4) 3 1 Slide the power select switch to the "M" side.
3-8 Disabling the Key Operation Key Lock The key lock function disables all key operation to prevent unauthorized use. Activating key lock 1 Press and hold the [MODE] button and (or ) simultaneously for 3 seconds or more. 3 Basic Operation The screen displays "Loc", disabling key operation and displaying the current received light intensity. Deactivating key lock 1 Press and hold the [MODE] button and (or ) simultaneously for 3 seconds or more. The screen displays "unL", enabling key operation.
3-8 Disabling the Key Operation Key Lock with PIN Number A PIN number can be set when activating key lock. Only users knowing the PIN number can operate the unit. Activating key lock with a PIN number 3 1 Press (or ) 10 times while holding down the [MODE] button. Basic Operation The screen displays "Loc 0". 2 Press the button to set a desired number (up to 4 digits). 3 Press the [MODE] button. The screen displays "Loc", disabling key operation and displaying the current received light intensity.
Settings for Advanced Functions This chapter describes settings for advanced functions of the LV-N10 Series. 4-1 4-2 4-3 4-4 4-5 4-6 4-7 4 List of Settings.................................................... 4-2 Basic Settings..................................................... 4-4 Detection Settings (Func)................................... 4-7 Display Settings (diSP)..................................... 4-22 System Settings (SYS)..................................... 4-29 2-output Settings ( ) ........
4-1 List of Settings Basic Settings…(page 4-4) Display Settings (diSP)…(page 4-22) MODE Press and hold for 3 seconds or more 6WT$ 5W24 7.
4-1 List of Settings )*14…(page 4-33) 2-Output ( Light intensity detection mode …(page 4-37) Limit setting output mode *15 …(page 4-34) Warning output mode …(page 4-35) Counter output mode *16 …(page 4-35) Timer OFF …(page 4-37) Off-delay timer *3 …(page 4-37) On-delay timer *3 …(page 4-37) One-shot timer *3 …(page 4-37) Settings complete Return to normal display *1 hsp cannot be selected when LV-S31/S62/S63 is connected.
4-2 Basic Settings Power Modes Refer to page 4-2 "Basic Settings" for setting methods. The detection stability will increase when the response time is delayed. You can select from the following 6 power modes. The response time slows down in the order of hSP, FinE, turb, Supr, ULtr, MEGA.
4-2 Basic Settings Percentage calibration The setting value can be specified as a percentage in respect to the current received light intensity. Setting example When the percentage calibration target value is set to "-10%".
4-2 Basic Settings Zero shift calibration This sensitivity setting performs the zero shift function and basic calibration (two-point calibration / maximum sensitivity calibration / full auto calibration) simultaneously. The lower of the received light intensity values specified at the time of sensitivity setting will automatically be set to "0". Reference The light intensity received with a workpiece present will be applied with the same amount of correction as for that received with no workpiece.
4-3 Detection Settings (Func) Output Timer Refer to page 4-2 "Detection Settings (Func)" for setting methods. There are 3 types of timers available. Display Function Timing chart* Without workpiece With workpiece Without workpiece toFF on-d Shot ON Control output OFF Off-delay timer Without workpiece With workpiece Without workpiece Turns off the output at a specified time after the detection signal Control output ON OFF goes off.
4-3 Detection Settings (Func) Detection Mode Refer to page 4-2 "Detection Settings (Func)" for setting methods. The table below lists the detection modes that can be selected.
4-3 Detection Settings (Func) DATUM1 mode Refer to page 4-2 "Detection Settings (Func)" for setting methods. In DATUM1 mode, the light intensity received Received light intensity without a workpiece is always corrected to "100.0". The setting value is also corrected "100.0" according to the correction amount, so that the ratio of setting value and received light intensity Fixed proportion is kept constant, with the result of stable detection. The display of the setting value does "5.
4-3 Detection Settings (Func) Point • When the received light intensity stays lower than a given warning level, the received light intensity correction stops and the DTM indicator flashes. "Changing the warning output level" (page 4-13) • The light intensity correction stops if the raw received light intensity drops below 50 (200 when the light intensity expansion setting is set to FuLL).
4-3 Detection Settings (Func) DATUM2 mode Refer to page 4-2 "Detection Settings (Func)" for setting methods. In DATUM2 mode, the light intensity received Received light intensity without a workpiece is always corrected to ".0". The setting value is also corrected according to the correction amount, so that the ratio of setting value and received light Fixed proportion intensity is kept constant, with the result of stable detection. The display of the setting value does not change. ".0" "5.
4-3 Detection Settings (Func) Point • When the received light intensity stays higher than a given warning level, the received light intensity correction stops and the DTM indicator flashes. "Changing the warning output level" (page 4-13) • The light intensity correction stops if the raw received light intensity drops below 50 (200 when the light intensity expansion setting is set to FuLL).
4-3 Detection Settings (Func) Changing the warning output level Refer to page 4-2 "Detection Settings (Func)" for setting methods. A level for outputting a warning can be set in the range of 0P (0%) to 100P (100%). (Default: 50P (50%)). • 100P (100%) • 50P (50%) • 0P (0%) Reference Control output ON OFF : A warning is not output even if the received light intensity changes.
4-3 Detection Settings (Func) Area detection mode Refer to page 4-2 "Detection Settings (Func)" for setting methods. In this mode, the output signal is delivered (in N.O. condition) if the received light intensity is in the area defined by Hi (upper limit) and Lo (lower limit). Hi (Upper limit setting value) Lo (Lower limit setting value) 4 0 Settings for Advanced Functions Control output ON in N.O. condition OFF Control output ON in N.C.
4-3 Detection Settings (Func) Methods for setting upper and lower limits The upper and lower limits can be set using one of the 2 methods: ● Setting sensitivity using the percentage calibration mode See "Percentage calibration" (page 4-5). ●Sensitivity setting using other calibration modes Sensitivity is set separately for Hi and Lo. 1 Press to display "Hi" or "Lo". The "Hi" or "Lo" value, and the current setting values are displayed alternately.
4-3 Detection Settings (Func) Edge detection mode Refer to page 4-2 "Detection Settings (Func)" for setting methods. In the edge detection mode, the sensor amplifier detects changes of received light intensity over a fixed period. This mode is used to turn on the output only when a workpiece is moving into or out of the detection range of the sensor head.
4-3 Detection Settings (Func) External Input Refer to page 4-2 "Detection Settings (Func)" for setting methods. On sensor amplifiers with an external input option (LV-N11 /N12 /N11C / N12C / N11MN), the external input line can be controlled to utilize the following functions: Display oFF SEt PrSt*1 rSt Disables use of the external input function. (Default) External calibration input Uses the external input to perform calibration. Preset input Executes a preset at the rising edge of the external input.
4-3 Detection Settings (Func) Pause mode transition input Refer to page 4-2 "Detection Settings (Func)" for setting methods. During signal input, the output can be locked in a desired state regardless of received light input. There are 3 selections of output status that can be locked with the pause function.
4-3 Detection Settings (Func) Sleep mode transition input Refer to page 4-2 "Detection Settings (Func)" for setting methods. This function holds the sensor amplifier in the power save state (sleep mode) during input of the external signal. Once the sleep mode is entered, the following events occur. • The display of received light intensity and setting value goes out. • Light transmission stops.
4-3 Detection Settings (Func) Analog scaling (LV-N11MN only) Refer to page 4-2 "Detection Settings (Func)" for setting methods. The LV-N11MN is equipped with an analog voltage output. In the default state, an analog output can be made in the range of 1 to 5 V in respect to the current received light intensity 0 to 4000. The received light intensity corresponding to the 5 V analog voltage can be changed within the following range.
4-3 Detection Settings (Func) Analog scaling mode (LV-N11MN only) The analog output can be automatically scaled so that the 5 V is output in respect to the received light intensity "100.0". This function is used simultaneously with the various preset function. This function is useful when frequently using the preset function as the analog output can be constantly sequenced with the ".0 to 100.0" display, etc.
4-4 Display Settings (diSP) Display Reverse Refer to page 4-2 "Display Settings (diSP)" for setting methods. The current value and setting value displays can be inverted. 4 Display Function oFF(OFF) on(ON) Normal display (Default) Reversed display Normal display Reversed display Sub Display Refer to page 4-2 "Display Settings (diSP)" for setting methods.
4-4 Display Settings (diSP) Displaying a sub display After selecting the desired sub-screen display, press the [MODE] button twice in the normal display state (setting value / received light intensity). Press the [MODE] button once again to return to the normal display.
4-4 Display Settings (diSP) Excess gain (%) display Refer to page 4-2 "Display Settings (diSP)" for setting methods. Received light intensity is displayed as an excess Display example If the setting value is 1000 gain* relative to the setting value. On the screen, and the received light the received light intensity is displayed as a bar intensity is 2000 (200%): graph and a percentage number.
4-4 Display Settings (diSP) Timing chart for standard setting (Std) [Peak value] • Samples while the current received light intensity > setting value ((1), (3), etc.). • If the current received light intensity is lower than the setting value, the previous peak value is updated. Peak value Received light intensity Setting value Bottom value Received light intensity (1) (3) [Bottom value] • Samples while the current received light intensity < setting value ((2), (4), etc.).
4-4 Display Settings (diSP) Excess gain hold (%) display Refer to page 4-2 "Display Settings (diSP)" for setting methods. This function is the same as the "received-light intensity hold display", except that the value being held is displayed as excess gain instead of received light intensity. "Light intensity hold display" (page 4-24) L-on / D-on display 4 Refer to page 4-2 "Display Settings (diSP)" for setting methods.
4-4 Display Settings (diSP) Preset Saturation Function Refer to page 4-2 "Display Settings (diSP)" for setting methods. The saturation level can be set in the range of 100P (100%) to 200P (200%) when the preset saturation function is ON. For example, if this function is set to 150P, the received light intensity will be corrected to "150.0" when the preset function is enabled, and the digital display will display "100.0".
4-4 Display Settings (diSP) Operating principles of the preset function (Current received light intensity is 3300 in this example.) Preset input is used to register the light intensity as "100.0". The actual received light intensity (3300) is set to "110%" and all actual received light intensities exceeding "100%" are displayed as "100.0". 4 When light intensity falls below 100%, a value lower than "100.0" is displayed. (The percentage value 110% (i.e.
4-5 System Settings (SYS) Power Save Refer to page 4-2 "System Settings (SYS)" for setting methods. Power consumption can be reduced by turning off the digital display and output indicators. Function oFF (OFF) Disables the power save function. (Default) Enables the power save function. (A single segment pulses across the digital display.) Reduces power consumption. (The power save function is ON, and response time is 4 times longer.) Eco ALL (The digital display and output indicators are all off).
4-5 System Settings (SYS) Display Gain Refer to page 4-2 "System Settings (SYS)" for setting methods. Received light intensity can be increased 4 times without sacrificing response time. Display Function Std (Standard) FuLL (Full) 4 Normal light intensity display (Default) Full light intensity display (Hysteresis increases 4 times) The maximum received light intensity value which can be displayed differs according to the power mode setting and display screen.
4-5 System Settings (SYS) Interference Prevention Refer to page 4-2 "System Settings (SYS)" for setting methods. Malfunction may result from "interference", a phenomenon in which light transmission occurs simultaneously with any other sensor amplifier. However, when the number of sensor amplifiers increases, the light transmission timing automatically shifts and prevents this interference.
4-5 System Settings (SYS) Common Key-Operations Function Refer to page 4-2 "System Settings (SYS)" for setting methods. Some operations on the main unit are applied equally and simultaneously to all expansion units connected to it. Display oFF (OFF) on (ON) 4 Function Disables common key operations. (Default) Enables common key operations.
4-6 2-output Settings ( ) * This function is only available for the 2-output type models (LV-N11 / N12 ). Detection Mode for Output2 Refer to page 4-3 "2-Output Settings ( Display )" for setting methods. Function ALrt (Alert) "Limit setting output mode" (page 4-34) Warning output mode (Output 2 turns on when an error occurs). Cnt (Count) *3 "Warning output mode" (page 4-35) Counter output mode (Counts the number of output-1 outputs.
4-6 2-output Settings ( ) Limit setting output mode Refer to page 4-3 "2-Output Settings ( )" for setting methods. Setting output 2 to "LiMt" (limit setting output mode) enables detection of reductions in received light intensity for output 1. Specifically, if the cumulative minimum of peak value for output 1 falls below the limit setting value, it is recognized as a reduction in received light intensity and therefore output 2 is turned on.
4-6 2-output Settings ( ) Warning output mode Refer to page 4-3 "2-Output Settings ( )" for setting methods. If output 2 is set to "ALrt" (warning output mode), output 2 turns on (or off if the output condition is set to N.C.) when either of the following errors occur: • Correction error in DATUM1 or DATUM 2 mode • ErH Reference For error recovery procedures, see "Troubleshooting" (page 6-2). 4 Counter output mode )" for setting methods.
4-6 2-output Settings ( ) Normal channel 2 display during counter output mode The screens shown to the right are alternately displayed. Press to adjust the setting value. Alternate display Setting value Count value How to reset the count value • Press and hold the [MODE] and [SET] buttons simultaneously. • Execute reset using external input. • Cycle the power.
4-6 2-output Settings ( ) Output Timer for Output2 Refer to page 4-3 "2-Output Settings ( )" for setting methods. There are 3 types of timers available. Display Function Timing chart* Without workpiece With workpiece Without workpiece Uses no output timer. (Default) oFFd Off-delay timer Without workpiece With workpiece Without workpiece Turns off the output at a specified time after the detection signal Control output ON OFF goes off.
4-7 Settings Save/Recall Custom Save (Settings Save) The current setting value and advanced function settings can be saved in the sensor amplifier. "Settings for Advanced Functions" (page 4-1) Point 4 1 Executing a custom save will override the previously saved settings with the current settings. Press and hold the [SET] and [PRESET] buttons simultaneously for 3 seconds or more. Settings for Advanced Functions The "rSt" display flashes.
4-7 Settings Save/Recall User Reset (Settings Recall) The settings saved with the custom save function can be recalled. Point Reference 1 Executing a user reset will override all of the previous settings with the settings that were saved using the custom save function. The factory settings will be recalled if a user reset is executed without previously executing a custom save. 4 Press and hold the [SET] and [PRESET] buttons simultaneously for 3 seconds or more.
4-7 Settings Save/Recall MEMO 4 Settings for Advanced Functions 4-40 - Digital Laser Sensor LV-N10 Series User's Manual -
Specifications This chapter provides the specifications, circuit diagrams and dimensions of the LV-N10 Series. 5-1 5-2 5-3 5 Specifications ..................................................... 5-2 Circuit Diagrams ................................................. 5-5 Dimensions.........................................................
5-1 Specifications Sensor Head (1) Type Model Laser class Adjustable spot type Coaxial reflective type Super small spot type Wide area reflective type Coaxial retroreflective type LV-NH32 LV-NH35 LV-NH37 LV-NH42 LV-NH62 FDA (CDRH) Part1040.10 IEC 60825-1 5 Specifications MEGA 1200 750 1200 8000 ULTRA 1000 600 1000 7000 SUPER 750 450 750 6000 TURBO 500 300 500 5000 FINE 250 150 250 3500 HIGH SPEED 200 100 200 2000 Approx.
5-1 Specifications Sensor Head (2) Type Model Laser class Distance detection type Reflective type Reflective type, side view Coaxial retro-reflective type LV-S31 LV-S41 LV-S41L LV-S61 FDA (CDRH) Part1040.
5-1 Specifications Sensor Amplifier Type 2 output 1 output 0-line Cable M8 connector – Cable Expansion unit Main unit Cable / Connector Main unit / Expansion unit Main unit Expansion unit Main unit Expansion unit Model NPN LV-N11N LV-N12N LV-N11CN LV-N12CN PNP LV-N11P LV-N12P LV-N11CP LV-N12CP Input/output Control output 2 output External input 1 input Monitor output Response time Monitor output LV-N11MN LV-N10 1 output N/A 1 input N/A – 1 output 1 input N/A 1 outpu
5-2 Circuit Diagrams Cable Type Input/output Circuit Diagram LV-N11P / N12P Short-circuit current PLC, etc. 1 mA or less 0V Pink Blue*1 *1 LV-N11N only PLC, etc. Short-circuit current 2 mA or less Black (Control output 1) White (Control output 2) Load White (Control output 2) DC3.
5-3 Dimensions Sensor Amplifier LV-N11 / N12 (cable type), LV-N10 (0-line type), LV-N11MN (monitor output Type) ø3.9 5 pins × Brown and Blue 0.34 mm2*2 Black, White and Pink 0.18 mm2 When amplifier mounting bracket (OP-73880) is used When cover is opened: 108.7 max. Cable length: 2m *1 2-ø3.4 Counterbore: ø6.4 Depth: 2.7 Max. 170° Bottom surface of amplifier mounting bracket 5 2(4.4×3.4) Counterbore: ø7.2 Depth: 3.2 Min. *1 This is not installed on the main unit (LV-N11 ). Œ3.
Appendix This chapter provides troubleshooting instructions and initial settings (default values). 6-1 6-2 6-3 6-4 6 Troubleshooting .................................................. 6-2 Factory Default Setting (Default Value) List ........ 6-5 List of Recipe Function Settings......................... 6-6 Restrictions on Each Detection Mode ................
6-1 Troubleshooting Frequently Asked Questions Category Trouble Cause Action • Review the installation environment. • The received light intensity can be stably displayed at 100.0 by using the preset function. The received light intensity is inconsistent. How can it be stabilized? It may be affected by vibration, thermal characteristics caused by thermal changes, or the state of the workpiece surface, etc. The received light intensity is too low.
6-1 Troubleshooting Category Trouble A value less than "100.0" is displayed even when the [PRESET] button is pressed. Cause Action The raw received light intensity is 50 or less. * (200 when using the received light intensity expansion setting.) • Press and hold the [MODE] button and check the raw received light intensity level. Try adjusting the beam axis, removing contamination, or reviewing the installation environment. Take actions to raise the original received light intensity.
6-1 Troubleshooting Error Displays and Corrective Actions Error display Cause Solution • Check the load and return the current within the rated range. • Check that the output wire is not contacting any other wire or frame. ErC Overcurrent in the control output. ErH • Check that sensor head is connected. • Check whether head cable is not broken. Head cable is broken, or sensor head • Check the head cable connection to the connecis not connected. tor. • After checking, turn the power OFF and ON.
6-2 Factory Default Setting (Default Value) List The default settings are as follows.
6-3 Reference List of Recipe Function Settings Refer to "Selecting Recipe" (page 3-19) for details on selecting recipes. The following details are set using the recipe function.
6-3 List of Recipe Function Settings *1 *2 *3 *4 When the LV-S31 is connected and recipe r-1 is loaded, the rising edge detection mode is disabled, and operation takes place as Std. If the recipe r-3/r-4 is loaded with the zero-line type (LV-N10), the preset functions will be disabled. When using the preset function, read the application settings other than r-3/r-4. When the recipes are recalled, all setting items other than those described here are set to the default values.
6-4 Restrictions on Each Detection Mode Restrictions for Sensitivity Settings in Each Detection Mode The sensitivity setting is limited according to the selected detection mode. Refer to the chart below for details. Reading the table (Example) When detection mode is set to dtM1 (DATUM1 mode) When the sensitivity setting is set to Std, two-point calibration, maximum sensitivity calibration and full auto calibration are enabled.
6-4 Restrictions on Each Detection Mode MEMO 6 Appendix - Digital Laser Sensor LV-N10 Series User's Manual - 6-9
6-5 Index This index provides a list of terms used in this document in alphabetical order. Numerics D 2-output Settings................................. 4-33 Detection Mode for Output2 ....... 4-33 Output Timer for Output2 ........... 4-37 A 6 Appendix Adjusting the Sensitivity........................ 3-5 Full Auto Calibration................... 3-14 Full Auto Preset Function............. 3-9 List of Sensitivity Adjustment Methods .......................................
6-5 Index M T Mounting and Adjusting the Sensor Head .........................................2-6 LV-NH100/NH110/NH300 ............2-6 LV-NH32/NH35/NH37/NH42/ NH62/S63.....................................2-6 LV-S31 .........................................2-7 LV-S41/41L ..................................2-8 LV-S62 .........................................2-8 LV-S71/S72 ................................2-10 Troubleshooting ....................................6-2 Error Displays and Corrective Actions......
6-5 Index Operation indicator (1-output type) Operation indicator (2-output type) PST indicator [SET] button Purpose Changing the output Description 1 Shifting the received light intensity to ."0" Loading the recommended settings 6-12 Change the output. (L-on/D-on) [PRESET] button Channel select switch (2-output type) Power select switch (1-output type) Reference page Operation procedures 1. Press the [MODE] button. 2. Switch with the ( 3-4 ) button.
6-5 Index Purpose Description Reference page Operation procedures 16 Initializing (Restore to factory default settings) 1. Press and hold the [SET] button and [PRESET] button. 2. Press the [MODE] button while on the rSt screen. 3. Select init with the ( ) button. 4. Press the [MODE] button to execute. 3-21 Displaying the output 2 display screen with the 2-output type 17 Switch to the output 2 display screen. Set the channel switch to .
6-5 Index Basic Settings…(page 4-4) Display Settings (diSP)…(page 4-22) MODE Press and hold for 3 seconds or more 6WT$ 5W24 7.
6-5 Index )*14…(page 4-33) 2-Output ( Light intensity detection mode …(page 4-37) Limit setting output mode *15 …(page 4-34) Warning output mode …(page 4-35) Counter output mode *16 …(page 4-35) Timer OFF …(page 4-37) Off-delay timer *3 …(page 4-37) On-delay timer *3 …(page 4-37) One-shot timer *3 …(page 4-37) Settings complete Return to normal display *1 hsp cannot be selected when LV-S31/S62/S63 is connected.
Revision History Date of printing Edition December, 2010 First edition January, 2011 Second edition Description
WARRANTIES AND DISCLAIMERS (1) KEYENCE warrants the Products to be free of defects in materials and workmanship for a period of one (1) year from the date of shipment. If any models or samples were shown to Buyer, such models or samples were used merely to illustrate the general type and quality of the Products and not to represent that the Products would necessarily conform to said models or samples.
Copyright (c) 2010 KEYENCE CORPORATION. All rights reserved.