2-8 Installation
Installation Notes
This is a precision instrument. Be sure that the operating environment is safe and avoid placing
the device in locations like the following.
<Operating environment>
• Locations that receive direct sunlight or with a temperature range outside of 0
°C to 55°C
• Locations with a relative humidity higher than 35 to 85% or locations that are subject to condensation
due to rapid temperature change.
• Areas with explosive or corrosive gasses, dust, salt, iron powder or soot
• Locations where the device may be subject to vibrations or sudden impact
• Locations with a risk of exposure to airborne water, oil or chemical droplets
• Locations with strong magnetic or electrical fields
• Do not lay the cabling for this device in the same conduits that carry electrical power cables.
<For internal installation>
Make sure that the area to which the device will be installed meets the following conditions.
• There is enough space for air flow.
• If the operating temperature goes above between 0
°C and 55°C, a strong fan or a cooler will
regulate the temperature.
• Do not install this device inside any high voltage devices.
• Install this device as far away from power cords as possible.