10. External input
You can select functions to be assigned to four external inputs (External input 1 to
4).
External input setting screen
When the user setting is set, each function of the external input 1 to 4 can be set.
Functions that can be set differ for each external input. (When the default is set,
these four screens are skipped.)
Press S / T button on each setting screen to specify functions
Reference
When the same function is selected for multiple external inputs, the
inputs function according to OR operation.
Zero shift input
When the zero shift input is set to ON, the internal measurement value (R.V.) is
shifted (offset) to the selected shift target value.
z Timing chart
z Minimum Input Time
ON time: 20 ms
OFF time: 20 ms
Reset input
When the reset input is turned ON while using the hold function, the judgment
value (P.V.) in the held state is canceled and the value becomes [-----].
z Timing chart
z Minimum Input Time
ON time: 20 ms
OFF time: 20 ms
T1 = Input response delay (20 ms max.) + reset time*
* Reset time (judgment output, analog output) = Sampling period × (averaging
rate + 2)
External input Description
Default
value
Default setting
Functions for the external inputs remain the same
as the factory default setting.
c
User setting
Functions for the external inputs are changed from
the factory default setting.
External Input 1 (Pink wire) External Input 2 (Yellow wire)
Setting
value
Description
Default
value
Setting
value
Description
Default
value
Shift
Zero shift input
c reset
Reset input
c
bnK-a
Bank A input
bnK-a
Bank A input
bnK-b
Bank B input
bnK-b
Bank B input
off
Not used
off
Not used
External input 3 (Pink/Purple wire) External Input 4 (Purple wire)
Setting
value
Description
Default
value
Setting
value
Description
Default
value
tiMnG
Timing Input
c bnK-a
Bank A input
c
bnK-a
Bank A input
bnK-b
Bank B input
bnK-b
Bank B input
off
Not used
off
Not used
10. in
dEFLt
Usr
HOLD
HI
V
kV
ALARM
BANK
ZERO SHIFT
TIMING
TEMP %RH
HI LO R.V.
0
1
2
3
LO
GO
HOLD
HI
V
kV
ALARM
BANK
ZERO SHIFT
TIMING
TEMP %RH
HI LO R.V.
0
1
2
3
LO
GO
HOLD
HI
V
kV
ALARM
BANK
ZERO SHIFT
TIMING
TEMP %RH
HI LO R.V.
0
1
2
3
LO
GO
HOLD
HI
V
kV
ALARM
BANK
ZERO SHIFT
TIMING
TEMP %RH
HI LO R.V.
0
1
2
3
LO
GO
HOLD
HI
V
kV
ALARM
BANK
ZERO SHIFT
TIMING
TEMP %RH
HI LO R.V.
0
1
2
3
LO
GO
HOLD
HI
V
kV
ALARM
BANK
ZERO SHIFT
TIMING
TEMP %RH
HI LO R.V.
0
1
2
3
LO
GO
10.External Input External input 1
setting screen
External input 2
setting screen
External input 3
setting screen
External input 4
setting screen
11.Bank switching
method
[MODE]
or X
[MODE]
or X
[MODE]
or X
[MODE]
or X
[MODE]
or X
Internal measurement
value (R.V.)
Max. 20 ms
Min. 20 msMin. 20 ms
R.V. before shift R.V. after shift
ON
OFF
Zero shift input
Judgment value
(P.V.) judgment
output
T1
Min. 20 ms
Min. 20 ms
Before resetting [-----]
ON
OFF
Reset input
Timing Input
When the timing input is turned ON while using the hold function, the judgment
value (P.V.) is held.
z Timing chart
When Sample hold is set
"Timing Input" (page 15) setting: "Level"
"Timing Input" (page 15) setting: "Edge"
When Peak hold, Bottom hold or Absolute value hold is set
"Timing Input" (page 15) setting: "Level"
"Timing Input" (page 15) setting: "Edge"
z Minimum Input Time
ON time: 2 ms
OFF time: 2 ms
Bank A input/Bank B input
By combining the ON/OFF status of bank A input and bank B input, you can
switch between banks.
z Timing chart
Reference
When the judgment value (P.V.) is held using the hold function, the
judgment value (P.V.) remains held even if the ON/OFF status of bank A
input or bank B input is switched. However, the judgment output will
change according to the HI/LO setting value of the new bank.
z Minimum Input Time
ON time: 20 ms
OFF time: 20 ms
Measuring period Measuring period
Max. 2 msMax. 2 msMax. 2 ms
Hold released
Sample hold
Min. 2 msMin. 2 ms
ON
OFF
Timing Input
Judgment value
(P.V.) judgment
output
Judgment value
(P.V.) judgment
output
Max. 2 msMax. 2 msMax. 2 ms
Sample holdSample hold Sample hold
Min. 2 msMin. 2 ms
ON
OFF
Timing Input
Max. 2 msMax. 2 ms
Min. 2 msMin. 2 ms
ON
OFF
Max. 2 ms
Sampling does not start Sampling period Sampling starts
Previous result
Judgment value
(P.V.) judgment
output
Timing Input
Judgment value
(P.V.) judgment
output
Max. 2 msMax. 2 ms
Min. 2 msMin. 2 ms
ON
OFF
Timing Input
Sampling period 1
Sampling period 1 result Sampling period 2 resultPrevious result
Sampling period 2 Sampling period 3
Judgment value
(P.V.) judgment
output
Max. 8 ms
Max. 20 ms
Max. 20 ms
Min. 20 ms
ON
OFF
Bank A input
ON
OFF
Bank B input
Max. 20 ms
Bank 1 Bank 2 Bank 3Bank 0
In this example, after 8 ms have passed, Bank A
and Bank B are judged as OFF and operation at
Bank 0 may be performed.