A-1 Error Messages
This section describes major error messages displayed during operations and the remedies when the
messages are displayed.
Error message Cause Remedy
Current SL-V system information mismatches
the last upload settings.
The last uploaded settings will be downloaded
or be verified with. Continue?
The system information sent to the
SL-V main unit and the actual
configuration of the SL-V is different.
Check the configuration
of the SL-V.
The amount of light received is not correctly
shown due to the unit mismatch.
The transmitter is not connected.
The transmitter and receiver are not
the same model.
Connect a transmitter.
Connect a transmitter
and receiver of the
same model.
Monitoring cannot be done due to the main-unit
mismatch.
The type and the number of beam
axes from the connected SL-V Series
(main unit) are different from the
settings of the main unit on the SL-V
Configurator.
Conform the settings of
the main unit on the SL-V
Configurator to the
settings on the SL-V.
Units mismatch. The system information on the SL-V
Configurator differs from the
information for the SL-V.
Conform the system
information on the SL-V
Configurator to the
configuration of the SL-V.
SL-V doesn't have an initial Password. A password has already been set on
the SL-V.
When you want to change
the password, refer to
"4-6 Changing the
Password" (Page 4-8). (If
you forget the password,
it cannot be changed.)