Instruction Manual

22
E SR-750 UM
Modulation MOD (Modulation)
Description
Evaluates the variation degree in cell
brightness
Calculation
formula
Each cell's MOD = 2 ½ (abs (R-GT)/SC)
R : Brightness value
GT : Binarization threshold value
SC : Symbol contrast
Criterion
Reflectance margin RM (Reflectance Margin)
Description
Evaluates the variation degree in cell
brightness with black and white of the correct
cell considered.
Calculation
formula
[White cell]
R >= GT MARGIN = 2 ½ (R-GT)/SC
R < GT MARGIN = 0
[Black cell]
R < GT MARGIN = 2 ½ (GT-R)/SC
R >= MARGIN = 0
MARGIN : Margin of each cell
R : Brightness value
GT : Binarization threshold value
SC : Symbol contrast
Criterion
Fixed pattern damage FPD (Fixed Pattern Damage)
Description
Evaluates the degree of the fixed pattern
damage (area on the right) dependent on the
code type.
Calculation
formula
Criterion
Format information damage
FID
(Format Information Damage)
Description
Evaluates the format information damage
degree of QR code.
Calculation
formula
Criterion
Version information damage
VID
(Version Information Damage)
Description
Evaluates the version information damage
degree of QR code (Model 2, version 2 and
later versions).
Calculation
formula
Criterion
Axial nonuniformity AN (Axial Nonuniformity)
Description
Evaluates the distortion degree in vertical and
horizontal size of the code.
Calculation
formula
AN = abs (Xavg = Yavg)/((Xavg = Yavg)/2)
Xavg : Average cell size in horizontal
direction
Yavg : Vertical cell size in horizontal
direction
Criterion
A : less than 0.06
B : 0.06 to 0.08
C : 0.08 to 0.10
D : 0.10 to 0.12
F : more than 0.12
Grid nonuniformity GN (Grid Nonuniformity)
Description
Evaluates the maximum position slip of each
cell
Calculation
formula
GN = Hmax/X
Hmax : Maximum position slip amount
X : Cell size
Criterion
A : less than 0.38
B : 0.38 to 0.50
C : 0.50 to 0.63
D : 0.63 to 0.75
F : more than 0.75
Unused error correction
UEC
(Unused Error Correction)
Description
Evaluates the ratio of error correction unused
at the time of decoding.
Calculation
formula
UEC = 1.0 - ((e + 2t)/(d - p))
e : Number of code words that cannot be
read
t : Number of error code words
d : Number of error corrected code words
p : Number of error detected code words
Criterion
A : more than 0.62
B : 0.50 to 0.62
C : 0.37 to 0.50
D : 0.25 to 0.37
F : less than 0.25
Print growth (horizontal)
PGH
(Print Growth Horizontal)
Description
Evaluates the mark cell growth in the
horizontal direction.
Calculation
formula
(D - 0.5)/0.15
D : On the horizontal clock pattern
Ratio of the number of pixels of mark
cell
Criterion
A : -0.50 to 0.50
B : -0.70 to -0.50 or 0.50 to 0.70
C : -0.85 to -0.70 or 0.70 to 0.85
D : -1.00 to -0.85 or 0.85 to 1.00
F : less than -1.00 or more than 1.00
Print growth (vertical) PGV (Print Growth Vertical)
Description
Evaluates the mark cell growth in the vertical
direction.
Calculation
formula
(D - 0.5)/0.15
D : On the vertical clock pattern
Ratio of the number of pixels of mark
cell
Criterion
A : -0.50 to 0.50
B : -0.70 to -0.50 or 0.50 to 0.70
C : -0.85 to -0.70 or 0.70 to 0.85
D : -1.00 to -0.85 or 0.85 to 1.00
F : less than -1.00 or more than 1.00