Instruction Manual

43
E SR-750 UM
(2) Image save mode
Default : Save final bank image
Setting range : Save final bank image, save specified number of images
from timing ON, save specified number of images before
timing OFF
Number of images saved
If the number of images to save has been specified from timing ON/OFF, then
specify this number.
Default : 10
Setting range : 1 to 10
FTP Settings
Set if Send by FTP is selected in saving images, and the SR-750 Series device is
operating as a client.
(3) IP Address
Set the IP address of the connection destination FTP server.
Default : 0.0.0.0
Setting range : 0.0.0.0 to 255.255.255.255
User Account/Password
Set the user account name and password of the destination FTP server.
Default : User Account: admin
Password: admin
Setting range : Specify up to 16 ASCII or hexadecimal characters
Passive mode
Default : Disabled (no check mark)
Setting range : Disable, Enable
Format
Default : JPEG
Setting range : Bitmap, JPEG
Quality (JPEG)
Set this when JPEG is selected as the format setting.
Default : 5
Setting range : 1 to 10
•Binning
Default : 1/4 skipping
Setting range : No binning, 1/4 skipping, 1/16 skipping, 1/64 skipping
(4) Edit image file name
When setting the edit image file name function, check "Use script".
* Valid only in FTP transmission
[Table] tab
On the list of this tab, you can set items that can be set in [Communication1],
[Communication2], [Operation], [I/O] and [Saving images] tabs.
Defaults and settings ranges are the same as those in each respective tabs.
Some functions can only be configured on the [Table] tab.
(1) Search box
Search whether the input words are present in the Table.
(1)
Communication1
RS-232C
Baud rate
Sets the baud rate.
Default : 115200 bps
Setting range : 9600, 19200, 38400, 57600, or 115200 bps
Data length
Sets the data bit length.
Default : 8 bits
Setting range : 7 or 8 bits
•Parity
Sets the parity.
Default : Even
Setting range : None, Even, or Odd
Stop bit
Sets the stop bit length.
Default : 1 bit
Setting range : 1 or 2 bits
•Protocol
Sets the output protocol for RS-232C read data.
Default : None
Setting range : None, PASS/RTRY, or ACK/NAK
Ethernet
IP address
Sets the IP address.
Default : 192.168.100.100
Subnet Mask
Sets the subnet mask.
Default : 24 (255.255.255.0)
Setting range : 8 to 30 (255.0.0.0 to 255.255.255.252)
Default Gateway
Sets the default gateway.
Default : 0.0.0.0
Communication2
Command port
Interface
Select the interface to perform command communication.
Default : RS-232C
Setting range : Not used, RS-232C, Ethernet
Port (Waiting)
Sets the port number to perform command communication. *Only when the
interface is set to Ethernet
Default : 9004
Setting range : 1024 to 65535
* 9013, 9014, 9015, 9016, 5900, 5910, and 44818 cannot be
used as they are reserved for the main unit system.
Keep Alive
Sets the keep-alive function for command communication. *Only when the
interface is set to Ethernet
Default : Enable
Setting range : Enable or Disable
Data port 1
Interface
Sets the interface to perform PC host output.
Default : RS-232C
Setting range : Not used, RS-232C, Ethernet
Remote IP address
Sets the IP address of the PC host. *Only when the interface is set to Ethernet
Default : 0.0.0.0 * When 0.0.0.0 is set, the SR-750 performs server
operations only.
Setting range : 0.0.0.0 to 255.255.255.255
Remote port
Sets the port number of the PC host. *Only when the interface is set to Ethernet
Default : 9004
Setting range : 1024 to 65535