Instruction Manual

72
E SR-750 UM
Timing OFF command
Send command LOFF
Response Normal process OK, LOFF
Abnormal process ER, LOFF, ee ee: Error code
Starts and finishes reading operations of the SR-750 Series. Same operations as
the timing ON/OFF signals from the IN terminal.
Timing ON/OFF commands can be changed, up to 8 characters.
When sending the trigger input command with the parameter bank number
added, the data is read only with the specified parameter bank.
Response is output upon receiving a send command.
When a code reading was successful while the SR-750 Series is in the Single
reading mode and the data transmission is set to Send after read, do not send a
timing OFF command. The response indicating the abnormal process is output.
When a code cannot be read and the trigger input OFF command is sent, a
reading error is output after response.
Timing OFF command cannot finish reading when the reading was started with
the IN terminal or the Test switch.
An error code is output when a timing ON command is received again during the
reading with a timing ON command. (Same for timing OFF commands)
If Specify response character is set to "SR-600 compatible", response is not sent.
If Specify response character is set to "User setting", the response command will
have the same character string as the one set for Success response character
and Failure response character.
"3-9 SR-600 Compatible Output Mode (Page 19)"
Preset Data Registration
Starting preset data registration
Send command PRON
Response Normal process OK, PRON
Abnormal process ER, PRON, ee ee: Error code
These are used to register preset data for the SR-750 Series.
When the preset data registration is completed, SR-750 outputs data in the
following format:
Finishing preset data registration
Send command PROFF
Response Normal process OK, PROFF
Abnormal process ER, PROFF, ee ee: Error code
These are used to finish or stop preset registration for the SR-750 Series.
After the SR-750 Series outputs preset registered data, do not send a PROFF
command.
When a PROFF command is sent to stop a preset registration, the following error
code will be output after the response:
If Specify response character is set to "SR-600 compatible", response is not sent.
"3-9 SR-600 Compatible Output Mode (Page 19)"
Reading Quick Setup Codes
Starting quick setup code reading
Send command RCON
Response Normal process OK, RCON
Abnormal process ER, RCON, ee ee: Error code
When a reading operation started and all quick setup codes were read, the
configuration will be saved in the ROM and applied to the operations.
Communication settings are also applied.
* Same for the quick setup code reading executed by the TUNE switch.
If Specify response character is set to "SR-600 compatible", OK is returned for
normal process and ERRee is returned for abnormal process.
"3-9 SR-600 Compatible Output Mode (Page 19)"
PR00 :
Preset registered data
PR01 : Read error
Finishing quick setup code reading
Send command RCOFF
Response Normal process OK, RCOFF
Abnormal process ER, RCOFF, ee ee: Error code
Finishes reading when a quick setup code reading is not completed.
When the SR-600 compatible output mode is "Enabled", OK and ERRee are
returned for normal process and abnormal process respectively.
"3-9 SR-600 Compatible Output Mode (Page 19)"
Quick setup code status check
Send command RCCHK
Response Normal process OK = RCCHK = mm…
mm = OK :
Quick setup code reading successful
READING : Reading quick setup code
NG : Quick setup code reading failed
NONE :
Quick setup code reading not executed
Abnormal process ER, RCCHK, ee ee: Error code
Checks the quick setup code reading status.
Test Mode Control
Starting reading rate test mode
Offline mode Send command Normal mode TEST1
Bank settings TEST1bb or TEST1, bb bb: parameter bank number (01 - 10)
Response Normal process OK, TEST1
Abnormal process ER, TEST1, ee ee: Error code
Online mode Send command Normal mode #TEST1
Bank settings #TEST1bb or #TEST1, bb bb: parameter bank number (01 - 10)
Response Normal process OK, #TEST1
Abnormal process ER, #TEST1, ee ee: Error code
Starting read time test mode
Offline mode Send command Normal mode TEST2
Bank settings TEST2bb or TEST2, bb bb: parameter bank number (01 - 10)
Response Normal process OK, TEST2
Abnormal process ER, TEST2, ee ee: Error code
Online mode Send command Normal mode #TEST2
Bank settings #TEST2bb or #TEST2, bb bb: parameter bank number (01 - 10)
Response Normal process OK, #TEST2
Abnormal process ER, #TEST2, ee ee: Error code
Starting code position test mode
Offline mode Send command Normal mode TEST3
Bank settings TEST3bb or TEST3, bb bb: parameter bank number (01 - 10)
Response Normal process OK, TEST3
Abnormal process ER, TEST3, ee ee: Error code
Finishing test mode
Offline mode Send command Normal mode QUIT
Response Normal process OK, QUIT
Abnormal process ER, QUIT, ee ee: Error code
Online mode Send command Normal mode #QUIT
Response Normal process OK, #QUIT
Abnormal process ER, #QUIT, ee ee: Error code
This outputs the results of each test after a response.
When a parameter bank number is appended at the end of the code, the test is
executed only for the specified parameter bank.
To finish the test mode, make sure to send an end command.
The code position test mode is used only in offline mode.
If Specify response character is set to "SR-600 compatible", OK is returned for
normal process and ERRee is returned for abnormal process.
"3-9 SR-600 Compatible Output Mode (Page 19)"