7 Specifications and Characteristics
144 Agilent 1000 Series Oscilloscopes User’s Guide
AC coupled
1
: DSO102xA: 5 Hz to 200 MHz
DSO101xA: 5 Hz to 100 MHz
DSO100xA: 5 Hz to 60 MHz
Calculated rise time
(= 0.35/bandwidth in GHz):
DSO102xA: 1.8 ns
DSO101xA: 3.5 ns
DSO100xA: 5.8 ns
Range: 2 mV/div to 10 V/div (1 M)
Maximum Input: Maximum input voltage for analog inputs:
CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
with N2862A/N2863A 10:1 probe: CAT I 600 V (DC + peak
AC)
Offset Range: ±2 V on ranges <500 mV/div;
±40 V on ranges 500 mV/div to 5 V/div;
Dynamic range: ±6 div
Input impedance: 1 M ± 1% || 18 ±3 pF
Coupling: AC, DC, ground
BW limit: 20 MHz selectable
DC vertical gain accuracy
3
: 2 mV/div to 5 mV/div: ±4.0% full scale
10 mV/div to 5 V/div: ±3.0% full scale
Channel-to-channel isolation: DC to max bandwidth >40 dB
Noise peak-to-peak: 3% full scale or 4.5 mV, whichever is greater
Table 7 Vertical System Characteristics (continued)
Name Typical Value
1
20 MHz when vertical scale is set to < 5 mV (1X probe attenuation).
3
Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute
warm-up period and ±10° C from firmware calibration temperature.