Datasheet

11
Electrical Specifications
AC specifications for U1271A
Function Range Resolution
Accuracy ± (% of reading + counts of least significant digit)
45 Hz to
65 Hz
30 Hz to
1 kHz
1 kHz to
5 kHz
5 kHz to
20 kHz
True RMS AC
Voltage
1
300 mV 0.01 mV 0.7 + 20 1.0 + 25 2.0 + 25 2.0 + 40
3 V 0.0001 V 0.7 + 20 1.0 + 25 2.0 + 25 2.0 + 40
30 V 0.001 V 0.7 + 20 1.0 + 25 2.0 + 25 2.0 + 40
300 V 0.01 V 0.7 + 20 1.0 + 25 2.0 + 25
1000 V 0.1 V 0.7 + 20 1.0 + 25
LPF (low pass filter) enabled,
applicable for all voltage
ranges and resolution
0.7 + 20
1.0 + 25 @ < 200 Hz
5.0 + 25 @ < 440 Hz
Function Range Resolution
Accuracy ± (% of reading +
counts of least significant digit)
Burden voltage/Shunt
45 Hz to 2 kHz
True RMS AC
Current
2
300 µA 0.01 µA 0.9 + 25 < 0.04 V/100 Ω
3000 µA 0.1 µA 0.9 + 25 < 0.4 V/100 Ω
30 mA 0.001 mA 0.9 + 25 < 0.08 V/1 Ω
300 mA 0.01 mA 0.9 + 25
< 1.00 V/1 Ω
3 A 0.0001 A 1.0 + 25 < 0.1 V/0.01 Ω
10 A 0.001 A 1.0 + 25 < 0.3 V/0.01 Ω
1. Notes for voltage specifications:
• Overload protection: 1000 Vrms. For millivolt measurements, 1000 Vrms for short circuits with < 0.3 A current.
• Input impedance: 10 MΩ (nominal) in parallel with < 100 pF.
2. Notes for current specifications:
• Overload protection for 300 μA to 300 mA range: 0.44 A/1000 V; 10 × 35 mm 30 kA fast-acting fuse.
• Overload protection for 3 A to 10 A range: 11 A/1000 V; 10 × 38 mm 30 kA fast-acting fuse.
• Specification for 300 mA range: 440 mA continuous.
• Specification for 10 A range: 10 A continuous. Add 0.3% to the specified accuracy when measuring signals > 10 to 20 A for 30 seconds
maximum. After measuring currents > 10 A, cool down the multimeter for twice the duration of the measured time before proceeding with low
current measurements.