User manual

62 U2300A Series DAQ Service Guide
BTest Limits
Analog Output
Analog Output
U2351A, U2353A:
Offset/
Gain Error (V)
Setting Range (V) Test Point (V) Tes t L imit (V) Tes t Result
Polarity Min Max Location Value Lower Upper Reading
0.001 BIP –10 10 0 0.0000 –0.0010 0.0010 Offset = DMM reading – DAQ source voltage
0.004 BIP –10 10 Positive FS 9.9997 9.9957 10.0037 DMM reading – Offset
0.004 BIP –10 10 Negative FS –10.0000 –10.0040 –9.9960 DMM reading – Offset
0.001 UNIP 0 10 0 0.0000 –0.0010 0.0010 Offset = DMM reading – DAQ source voltage
0.002 UNIP 0 10 Positive FS 9.9998 9.9978 10.0018 DMM reading – Offset
Analog Output
U2355A, U2356A:
Offset/
Gain Error (V)
Setting Range (V) Test Point (V) Tes t L imit (V) Tes t Result
Polarity Min Max Location Value Lower Upper Reading
0.001 BIP –10 10 0 0.0000 –0.0010 0.0010 Offset = DMM reading – DAQ source voltage
0.004 BIP –10 10 Positive FS 9.9951 9.9911 9.9991 DMM reading – Offset
0.004 BIP –10 10 Negative FS –10.0000 –10.0040 –9.9960 DMM reading – Offset
0.001 UNIP 0 10 0 0.0000 –0.0010 0.0010 Offset = DMM reading – DAQ source voltage
0.002 UNIP 0 10 Positive FS 9.9976 9.9956 9.9996 DMM – Offset
Analog Output
U2331A:
Offset/
Gain Error (V)
Setting Range (V) Test Point (V) Tes t L imit (V) Tes t Result
Polarity Min Max Location Value Lower Upper Reading
0.0015 BIP –10 10 0 0.0000 –0.0015 0.0015 Offset = DMM reading – DAQ source voltage
0.004 BIP –10 10 Positive FS 9.9951 9.9911 9.9991 DMM reading – Offset
0.004 BIP –10 10 Negative FS –10.0000 –10.0040 –9.9960 DMM reading – Offset
0.0025 UNIP 0 10 0 0.0000 –0.0025 0.0025 Offset = DMM reading – DAQ source voltage
0.004 UNIP 0 10 Positive FS 9.9976 9.9936 10.0016 DMM reading – Offset