Datasheet
Page 10
Find us at www.keysight.com
– Ideal for measuring up to 3.7 kV
– Up to 500 MHz bandwidth
– 100:1 attenuation
10076C high-voltage probe
The Keysight 10076C 3.7 kV 100:1 passive probe gives you the voltage and bandwidth
you need for making high-voltage measurements. Its compact design makes it easier to
probe today’s small power electronics components and its rugged construction means it
can withstand rough handling without breaking.
Characteristics 10076C
Bandwidth 500 MHz (–3 dB)
Risetime (calculated) < 0.7 ns
Attenuation ratio 100:1
Input resistance 66.7 MΩ (when terminated into 1 MΩ)
Input capacitance Approx 3 pF
Maximum input 3700 Vpk, 2650 Vrms, 2650 Vdc (mains isolated)
1000 Vpk CAT II (post receptacle mains)
Compensation 6 to 18 pF range
Probe readout Yes
Cable length 1.8 m
Operating temperature 0 °C to 50 °C
Ordering information for Keysight high-voltage probes
10076C High-voltage probe: includes one retractable hook tip, one ground-bayonet, one IC
probing tip, one alligator ground lead, one spring ground tip and a compensation
screwdriver
10077A Accessory kit for 10076A/C: includes one retractable
pincher tip, one ground lead, one insulation cap, two measuring pins, and two
colored tags
N2789A Spring ground tip for 10076B/C
10076C passive probe
10077A accessory kit for 10076C
10076C derating curve
High-voltage Passive Probes
N2789A spring ground clip for 10076B/C