Datasheet
Page 17
Find us at www.keysight.com
High-voltage Differential Active Probes (Continued)
N2804A/N2805A high-voltage, high-speed differential probes
The N2804A and N2805A differential probes provide the superior general-purpose
differential signal measurements that are required for high-speed power measurements
such as measuring characteristics of switching power devices, DC-DC converters or
class D amplifiers, vehicle bus measurements, and high-speed digital system designs.
The N2804A 300-MHz differential probe offers 100:1 attenuation ratio, allowing it to be
used adequately for high voltage signal measurements. The differential probe has a
differential input resistance of 4 MΩ and low input capacitance of 4 pF to minimize
circuit loading. The probe comes with a pair of extension leads (30 cm long) with a
damping resistor built-in to damp out the in-band resonance and provide flat frequency
response even while using the extension leads and probe tip accessories.
The N2805A is a 200-MHz differential probe designed to provide superior differential
signal measurements with a long cable length (5 m), making it ideal in an environment
where extended cable length is required. This probe comes with an extensive set of
probe tip accessories for use with small and large components in tight spaces.
Characteristics for N2804A and N2805A
N2804A N2805A
Bandwidth 300 MHz (without extension leads) 200 MHz
120 MHz (with extension leads)
Attenuation ratio 100:1 50:1
CMRR —80 dB at 50/60 Hz —75 dB at 50/60 Hz
—75 dB at 1 MHz —80 dB at 1 MHz
Input impedance
(between inputs)
4 MΩ // 4 pF 4 MΩ // 4 pF
Max input voltage
(between two inputs)
± 300 V (DC+ peak AC) and
± 200 Vrms
± 200 V (DC+ peak AC) and
± 200 Vrms CAT II
Max input voltage ± 300 V (DC+ peak AC) and
± 200 Vrms CATII
± 500 V (DC+ peak AC) and
± 500 Vrms mains isolated
± 1000 V (DC+ peak AC) and
± 1000 Vrms CATI
± 300 V (DC+ peak AC) and
± 200 Vrms CATII
Cable length 1.2 m 5 m
Max number of probes
supported by InfiniiVision
Series
4 4
Compatible InfiniiVision
oscilloscopes
6000X Series, 4000X Series with software ver. 4.0 or higher, 3000X
Series with software ver. 2.38 or higher
For more information about the N7020A power rail probe, refer to the Keysight data
sheet with the publication number 5992-0141EN.
N2804A 300 MHz differential probe
N2805A 200 MHz differential probe