Datasheet

Page 8
Find us at www.keysight.com
Characteristics for Keysight passive probes
10070D 10073D 10074D
Bandwidth 20 MHz 500 MHz 150 MHz
Risetime (calculated) < 17.5 ns < 700 ps < 2.33 ns
Attenuation ratio 1:1 10:1 10:1
Input resistance 1 MΩ 2.2 M 10 M
Input capacitance Approx 70 pF Approx 12 pF Approx 15 pF
Maximum input (dc + peak ac) 300 Vpk mains isolated 500 Vpk mains isolated 500 Vpk mains isolated
150 Vpk CAT II 400 Vpk CAT II 400 Vpk CAT II
Compensation range None 6 to 15 pF 9 to 17 pF
Probe readout Yes Yes Yes
Cable length 1.5 m 1.5 m 1.5 m
N2142A N2140A
Bandwiidth 75 MHz at 10:1 200 MHz at 10:1
6 MHz at 1:1 6 MHz at 1:1
Risetime (calculated) 4.66 nsec at 10:1 1.75 nsec at 10:1
58.3 nsec at 1:1 58.3 nsec at 1:1
Attenuation ratio 1:1/10:1 switchable 1:1/10:1 switchable
Input resistance 10 MΩ at 10:1 10 MΩ at 10:1
1 MΩ at 1:1 1 MΩ at 1:1
Input capacitance 15 pF at 10:1 15 pF at 10:1
100 pF at 1:1 100 pF at 1:1
Maximum input (dc + peak ac) 300 Vrms CAT II at 10:1 300 Vrms CAT II at 10:1
150 Vrms CAT II at 1:1 150 Vrms CAT II at 1:1
Compensation range 15 pF to 40 pF 15 pF to 40 pF
Probe readout No No
Cable length 120 cm 120 cm
N2862B,41A/
N2863B,42A
N2889A N2890A/N2843A N2873A/N2894A N2853A N2840A
Bandwidth 150 MHz/
300 MHz
350 MHz (at 10:1),
10 MHz (at 1:1)
500 MHz 500 /700 MHz 350 MHz 50 MHz
Risetime (calculated) < 2.33 ns/
< 1.16 ns
<1 ns (at 10:1),
< 35 ns (at 1:1)
< 700 ps < 700 ps/
< 500 ps
< 1 ns < 7 ns
Attenuation ratio 10:1 1:1/10:1 switchable 10:1 10:1 10:1 10: 1
Input resistance 10 M 10 MΩ (at 10:1)
1 MΩ (at 1:1)
10 MΩ 10 MΩ 10 MΩ 10 MΩ
Input capacitance Approx 11 pF Approx 11 pF (at 10:1),
60 pF (at 1:1)
Approx 11 pF Approx 9.5 pF Approx 11 pF Approx 11 pF
Maximum input
(dc + peak ac)
300 V
mains isolated/
CAT II
300 V mains isolated/
CAT II (at 10:1), 150 V
mains isolated/CAT II
(at 1:1)
300 V
mains isolated/
CAT II
400 V mains
isolated,
300 V CAT II
300 V
mains isolated/
CAT II
300 V
mains isolated/
CAT II
Compensation range 5 to 30 pF 5 to 30 pF (at 10:1) 5 to 30 pF 10 to 25 pF 5 to 30 pF 5 to 30 pF
Probe readout Yes No Yes Yes Yes No
Cable length 1.2 m 1.3 m 1.3 m 1.2 m 2 m 1.2 m
Note: 700 MHz BW only available on DSOX/MSOX 4000 X-Series oscilloscopes with 1 GHz or 1.5 GHz bandwidth models.
Passive Probes (Continued)