Keysight 53131A/132A/181A Counters Data Sheet Recommended replacement products: 53200 Series RF and universal frequency counter/timers (Data sheet publication number: 5990-6283EN)
02 | Keysight | 53131A/132A/181A Counters - Data Sheet Overview –– 225 MHz bandwidth (optional 1.5, 3, 5, or 12.4 GHz) –– 10- or 12-digit resolution with 1 s gate time –– GPIB interface and IntuiLink connectivity software standard –– Data transfer rate of up to 200 fully formatted measurements/second A family of universal and RF counters to meet your needs The Keysight Technologies, Inc. 53131A/132A/181A high-performance counters give you fast, precise frequency measurements at an affordable price.
03 | Keysight | 53131A/132A/181A Counters - Data Sheet Keysight IntuiLink provides easy access to the counter’s data from your PC The Keysight 53131A/132A/181A counters, capture precise frequency and time measurements. IntuiLink software allows that data to be put to work easily. You work in a familiar environment at all times, using PC applications such as Microsoft Excel or Word to analyze, interpret, display, print, and document the data you get from the counter.
04 | Keysight | 53131A/132A/181A Counters - Data Sheet Instrument Inputs Input specifications Frequency range Voltage range and sensitivity (Sinusoid) 2 Channel 1 and 2 (53131A, 53132A) 1 Channel 1 (53181A) DC coupled AC coupled FM tolerance DC to 100 MHz 100 to 200 MHz 200 to 225 MHz Voltage range and sensitivity (Single-shot pulse) 2 4.5 to 10 ns pulse width > 10 ns pulse width Trigger level 2 Frequency range Range Accuracy Resolution 50 Ω 0 to 3.5 kHz, 1 MΩ 3.
05 | Keysight | 53131A/132A/181A Counters - Data Sheet Instrument Inputs (Continued) Input specifications 4 (Continued) External arm input specifications 5 Signal input range Timing restrictions Channel 3 (53131A, 53132A) Channel 2 (53181A) TTL compatible Pulse width Transition time Start-to-stop time Damage level > 50 ns < 250 ns > 50 ns 10 Vrms Impedance Input capacitance Start/stop slope 1 kΩ 17 pF Positive or negative Voltage range Damage level Frequency 200 mVrms to 10 Vrms 10 Vrms 1 MHz, 5 MHz,
06 | Keysight | 53131A/132A/181A Counters - Data Sheet For automatic or external arming: (and signals < 100 Hz using timed arming) LSD displayed: t res Gate time Frequency or period x t 2res + (2 x Trigger error) RMS resolution: 2 x Gate time 53131A Typical t res 650 ps 53132A t res 200 ps 53181A Frequency or period t res 650 ps See graphs for worst case resolution performance.
07 | Keysight | 53131A/132A/181A Counters - Data Sheet For time or digits arming: LSD displayed: RMS resolution (see Graph 2): 2 Gate time x 4x 2 x t res Number of samples t 2res + (2 x Trigger error Gate time x 2 ) x Gate time t jitter + Frequency or period x Frequency or period Gate time Number of samples 53131A/181A Typical t jitter + 53132A t res t jitter t res t jitter 500 ps 50 ps 225 ps 3 ps See graphs for worst case resolution performance.
08 | Keysight | 53131A/132A/181A Counters - Data Sheet Measurement Specifications Frequency (53131A, 53132A, 53181A) Channel 1 and 2 (53131A, 53132A) Channel 1 (53181A) Channel 3 (53131A, 53132A) Channel 2 (53181A) Period (53131A, 53132A, 53181A) Channel 1 and 2 (53131A, 53132A) Channel 1 (53181A) Channel 3 (53131A, 53132A) Channel 2 (53181A) Frequency ratio (53131A, 53132A, 53181A) Measurement is specified over the full signal range of each input.
09 | Keysight | 53131A/132A/181A Counters - Data Sheet Measurement Specifications (Continued) Gate time Auto mode 1 ms to 1000 s GPIB ASCII 200 measurements/s (maximum) Start measurement Stop measurement Time interval Delayed arming Free run, manual, or external Continuous, single, external, or timed 100 μs to 10 s (53131A) 100 ns to 10 s (53132A) Measurement throughput Measurement arming Arming modes Note that not all arming modes are available for every measurement function 5.
| Keysight | 53131A/132A/181A Counters - Data Sheet Time interval, pulse width, rise/fall time (53131A and 53132A only): RMS resolution: (t res ) 2 + Start trigger error 2 + Stop trigger error 2 Systematic uncertainty: ± (Time base error × Measurement) Trigger level timing error ± 1.
| Keysight | 53131A/132A/181A Counters - Data Sheet Auto arming Measurements are initiated immediately and acquired as fast as possible, using a minimum number of signal edges. Timed arming The duration of the measurement is internally timed to a user-specified value (also known as the “gate time”). Digits arming Measurements are performed to the requested resolution (number of digits) through automatic selection of the acquisition time.
| Keysight | 53131A/132A/181A Counters - Data Sheet Fractional time base error (see Graph 3) Time base error is the maximum fractional frequency variation of the time base due to aging or fluctuations in ambient temperature or line voltage: Time base error = ( ∆f f Aging rate + ∆f f Temperature + ∆f f Line voltage ) Multiply this quantity by the measurement result to yield the absolute error for that measurement. Averaging measurements will not reduce (fractional) time base error.
| Keysight | 53131A/132A/181A Counters - Data Sheet Differential channel error The differential channel error term stated in several systematic uncertainty equations accounts for channel-to-channel mismatch and internal noise. This error can be substantially reduced by performing a TI calibration (accessible via the utility menu) in the temperature environment in which future measurements will be made. Graph 1: 1E + 02 The graphs may also be used to compute errors for period measurements.
Graph 3: Frequency error Timebase error 7 10 kHz 10 µs 1 kHz 1 µs 100 Hz 100 ns 10 Hz 10 ns 1 Hz 1 ns 100 mHz 100 ps 10 mHz 10 ps 1 mHz 1 ps 100 µHz 100 fs 10 µHz 10 fs 1 µHz 1 fs 100 nHz Time error 14 | Keysight | 53131A/132A/181A Counters - Data Sheet 100 as 10 nHz 1 Hz 10 Hz 100 Hz 1 kHz 10 kHz 100 kHz 1 MHz 10 MHz 100 MHz 1 GHz 1 ns 10 ns 100 ns 1 µs 10 µs 100 µs 1 ms 10 ms 100 ms 1 sec 10 as Input signal frequency or time Remarks: High stability T.B.
| Keysight | 53131A/132A/181A Counters - Data Sheet Graph 5: E + 02 Keysight 53132A–Worst case RMS resolution 7 (Time or digits arming) 7. Frequency error (Hz) 1E + 00 Graphs 1, 2, 4 and 5 do not reflect the effects of trigger error.
| Keysight | 53131A/132A/181A Counters - Data Sheet Measurement Statistics Available statistics Number of measurements 2 to 1,000,000 Measurements Description Mean, Minimum, Maximum, Standard Deviation Statistics may be collected on all measurements or on only those which are between the limit bands. When the limits function is used in conjunction with statistics, N (number of measurements) refers to the number of in-limit measurements.
| Keysight | 53131A/132A/181A Counters - Data Sheet Ordering Information Model number Description 53131A 10 digit/s, 500 ps universal counter 53132A 12 digit/s, 150 ps universal counter 53181A 10-digit/s RF counter Accessories included Each counter comes with IntuiLink software, standard timebase, and power cord. CD with the following: IntuiLink software, Operating, Programming, Service and Getting Started Guides, a data sheet, and application notes.
| Keysight | 53131A/132A/181A Counters - Data Sheet 254.4 mm 103.6 mm 374.0 mm 212.6 mm 88.5 mm 348.
| Keysight | 53131A/132A/181A Counters - Data Sheet Evolving Since 1939 Our unique combination of hardware, software, services, and people can help you reach your next breakthrough. We are unlocking the future of technology. From Hewlett-Packard to Agilent to Keysight. myKeysight www.keysight.com/find/mykeysight A personalized view into the information most relevant to you. http://www.keysight.