Keysight Technologies Digital Multimeters 34460A, 34461A, 34465A (6½ digit), 34470A (7½ digit) Data Sheet Truevolt DMMs for your next generation of insights
02 | Keysight | Digital Multimeters: 34460/61/65/70A - Data Sheet Keysight’s NEW Truevolt Digital Multimeters (DMMs) offer a full range of measurement capabilities and price points with higher levels of accuracy, speed, and resolution. Measure low-power devices The ability to measure very low current, 1 µA range with pico second resolution, allows you to make measurements on very low power devices.
03 | Keysight | Digital Multimeters: 34460/61/65/70A - Data Sheet Bar meter mode provides the number display along with an analog meter to provide a visual view of your measurements. Number mode provides the traditional “digits” view of measurements. Measure with Unquestioned Truevolt Confidence Worry about the quality of your design, not the quality of your measuments In a rack or on a bench real-world signals are never flat.
04 | Keysight | Digital Multimeters: 34460/61/65/70A - Data Sheet BenchVue Software Data capture simplied. Click. Capture. Done. BenchVue software for the PC makes it simple to connect, record and achieve results with your instruments with no programming. BenchVue is an all-in-one application that enables you to: Figure 1. See your measurements across instruments in one place to quickly correlate measurement activities and obtain actionable insights.
05 | Keysight | Digital Multimeters: 34460/61/65/70A - Data Sheet Measure with Unquestioned Truevolt Confidence What Truevolt technology means to you: You can measure your real-world signals, not instrument error Noise and injected current: Keysight Truevolt DMMs contribute less than 30% of the injected current than alternatives. Compared to some lower cost alternatives, Truevolt DMMs offer almost 100% less noise.
06 | Keysight | Digital Multimeters: 34460/61/65/70A - Data Sheet Move to The Next-Generation 34401A DMM with 100% Assurance Migrate with confidence: Everything you depend on with the 34401A and more Like most 34401A DMM owners, you rely on your DMM and you trust the answers it gives you. Now, with the Keysight Truevolt 34461A DMM, you can get all of the advantages of the 34401A and more. Now you can get faster answers and have even more confidence in your results.
07 | Keysight | Digital Multimeters: 34460/61/65/70A - Data Sheet Specifications 34460A 34460A accuracy specifications: ± (% of reading + % of range) 1 These specification are compliant to ISO/IEC 17025 for K = 2 Range 2 /frequency 24 hours 3 TCAL ± 1 °C 90 days TCAL ± 5 °C 1 year TCAL ± 5 °C 2 years TCAL ± 5 °C Temperature coefficient/°C 4 0.0040 + 0.0060 0.0070 + 0.0065 0.0090 + 0.0065 0.0115 + 0.0065 0.0005 + 0.0005 DC voltage 100 mV 1V 0.0030 + 0.0009 0.0060 + 0.0010 0.0080 + 0.0010 0.
08 | Keysight | Digital Multimeters: 34460/61/65/70A - Data Sheet Specifications 34460A 24 hours 3 TCAL ± 1 °C 90 days TCAL ± 5 °C 1 year TCAL ± 5 °C 2 years TCAL ± 5 °C Temperature coefficient/°C 4 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.015 + 0.006 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.030 + 0.006 3 Hz – 5 kHz 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.015 + 0.006 5 – 10 kHz (typ) 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.030 + 0.
09 | Keysight | Digital Multimeters: 34460/61/65/70A - Data Sheet Specifications 34461A 34461A accuracy specifications: ± (% of reading + % of range) 1 These specification are compliant to ISO/IEC 17025 for K = 2 Range 2 /frequency 24 hours 3 TCAL ± 1 °C 90 days TCAL ± 5 °C 1 year TCAL ± 5 °C 2 years TCAL ± 5 °C Temperature coefficient/°C 4 DC voltage 100 mV 0.0030 + 0.0030 0.0040 + 0.0035 0.0050 + 0.0035 0.0065 + 0.0035 0.0005 + 0.0005 1V 0.0020 + 0.0006 0.0030 + 0.0007 0.0040 + 0.0007 0.
| Keysight | Digital Multimeters: 34460/61/65/70A - Data Sheet Specifications 34461A Range 2 /frequency True RMS AC current 2, 6, 9 100 µA, 1 mA, 10 mA, and 100 mA ranges 3 Hz – 5 kHz 5 – 10 kHz (typ) 1 A range 3 Hz – 5 kHz 5 – 10 kHz (typ) 3 A range 3 Hz – 5 kHz 5 – 10 kHz (typ) 10 A range 8 3 Hz – 5 kHz 5 – 10 kHz (typ) Continuity 1 kΩ 24 hours 3 TCAL ± 1 °C 90 days TCAL ± 5 °C 1 year TCAL ± 5 °C 2 years TCAL ± 5 °C Temperature coefficient/°C 4 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.10 + 0.
| Keysight | Digital Multimeters: 34460/61/65/70A - Data Sheet Specifications 34465A 34465A accuracy specifications: ± (% of reading + % of range) 1 DC voltage and resistance. Automatic calibration (ACAL) capable. 24 hours 3 TACAL ± 1 °C 90 days TACAL ± 2 °C 1 year TACAL ± 2 °C 2 years TACAL ± 2 °C Non ACAL 7 Temperature coefficient/°C With ACAL 8 Temperature coefficient/°C 100 mV 0.0030 + 0.0030 0.0040 + 0.0035 0.0050 + 0.0035 0.0065 + 0.0035 0.0005 + 0.0005 0.0002 + 0.0005 1V 0.
| Keysight | Digital Multimeters: 34460/61/65/70A - Data Sheet Specifications 34465A Temperature Probe accuracy + 0.05 °C PT100 (DIN/ IEC 751) 13 5 kΩ thermistor Probe accuracy + 0.1 °C K,J,T,E,N thermocouples R thermocouples 14 14 (250 - 1760 °C) Probe accuracy + reference junction accuracy + 0.3 °C Probe accuracy + reference junction accuracy + 0.5 °C 24 hours 3 TCAL ± 1 °C 90 days TCAL ± 5 °C 1 year TCAL ± 5 °C 2 years TCAL ± 5 °C Temperature coefficient/°C 6 3 – 5 Hz 0.50 + 0.02 0.
| Keysight | Digital Multimeters: 34460/61/65/70A - Data Sheet Specifications 34465A DC and AC current burden voltage at full scale DC current range Burden voltage 1 µA 10 µA 100 µA 1 mA 10 mA 100 mA 1A < 0.0011 V < 0.011 V < 0.11 V < 0.11 V < 0.027 V < 0.27 V 3A < 2.0 V/0.15 V 22 < 0.5 V 10 A < 0.7 V/0.
| Keysight | Digital Multimeters: 34460/61/65/70A - Data Sheet Specifications 34470A 34470A accuracy specifications: ± (% of reading + % of range) 1 DC voltage and resistance. Automatic calibration (ACAL) capable. Range 2 24 hours 3 TACAL ± 1 °C 90 days TACAL ± 5 °C 1 year TACAL ± 5 °C 2 years TACAL ± 5 °C Non ACAL 11 Temperature coefficient/°C With ACAL 12 Temperature coefficient/°C DC voltage 100 mV 0.0030 + 0.0030 0.0040 + 0.0035 0.0040 + 0.0035 0.0045 + 0.0035 0.0005 + 0.0005 0.
| Keysight | Digital Multimeters: 34460/61/65/70A - Data Sheet Specifications 34470A Temperature Probe accuracy + 0.05 °C PT100 (DIN/ IEC 751) 13 5 kΩ thermistor Probe accuracy + 0.1 °C K,J,T,E,N thermocouples R thermocouples 14 14 (250 - 1760 °C) Probe accuracy + reference junction accuracy + 0.3 °C Probe accuracy + reference junction accuracy + 0.5 °C 24 hours 3 TCAL ± 1 °C 90 days TCAL ± 5 °C 1 year TCAL ± 5 °C 2 years TCAL ± 5 °C Temperature coefficient/°C 6 3 – 5 Hz 0.50 + 0.02 0.
| Keysight | Digital Multimeters: 34460/61/65/70A - Data Sheet Specifications 34470A DC and AC current burden voltage at full scale DC current range Burden voltage 1 µA < 0.0011 V 10 µA < 0.011 V 100 µA < 0.11 V 1 mA < 0.11 V 10 mA < 0.027 V 100 mA < 0.27 V 1A < 0.7 V/0.05 V 22 3A < 2.0 V/0.15 V 22 10 A < 0.
| Keysight | Digital Multimeters: 34460/61/65/70A - Data Sheet Measurement Characteristics (for all models except where noted) DC voltage Measurement method: A/D Linearity: 34460/61A 34465A 34470A Input resistance: 0.1 V, 1 V, 10 V range 100 V, 1,000 V range Input bias current: Input terminals: Input protection: True RMS AC voltage Measurement type: Keysight patented continuously integrating multi-slope IV A/D converter 0.0002% of reading + 0.0001% of range 0.0001% of reading + 0.0001% of range 0.
| Keysight | Digital Multimeters: 34460/61/65/70A - Data Sheet Operating Characteristics (for all models except where noted) Performance versus measurement speed For DC voltage, DC current, and resistance 1 (34460A & 34461A) 34460A Integration time 34461A Digits Readings/s Digits Readings/s 100 PLC/1.67 s (2 s) 6½ 0.6 (0.5) 6½ 0.6 (0.5) 0% of range 10 PLC/167 ms (200 ms) 6½ 6 (5) 6½ 6 (5) 0% of range 1 PLC/16.7 ms (20 ms) 5½ 60 (50) 5½ 60 (50) 0.001% of range 2 0.
| Keysight | Digital Multimeters: 34460/61/65/70A - Data Sheet System speeds (nom) DC voltage, DC current, resistance 1, 2 34460A 34461A 34465A/34470A Autorange time 3 < 30 ms < 30 ms <5 ms Maximum internal trigger rate 300/s 1000/s 5,000/s Maximum external trigger rate 300/s 1000/s 5,000/s ASCII readings to bus 300/s 1000/s 40,000/s (GPIB 8,000/s) 50/s 150/s 250/s Single reading transaction rate 4 AC voltage, AC current 5 Autorange time 3 10/s 10/s < 5 ms Maximum internal t
| Keysight | Digital Multimeters: 34460/61/65/70A - Data Sheet General Characteristics (for all models except where noted) Line power Power supply: Power line frequency: Power consumption: Environment Operating environment: Operating altitude: Storage temperature: Mechanical Rack dimensions: Bench dimensions: Weight: Regulatory Safety EMC Acoustic noise (nominal) Triggering conditions External input Delay: Jitter: Minimum pulse width: Maximum rate: Voltmeter complete output Polarity: Pulse width: Com
| Keysight | Digital Multimeters: 34460/61/65/70A - Data Sheet Options & Accessories (for all models except where noted) Options 34460A Digital multimeter, 6½ digit, basic Truevolt DMM LAN Rear panel LAN/LXI web interface, external triggering for 34460A – factory enabled SEC NISPOM and file security for Truevolt Series DMMs – factory enabled Z54 Certificate of calibration – ANSI/NCSL Z540.
| Keysight | Digital Multimeters: 34460/61/65/70A - Data Sheet myKeysight www.keysight.com/find/mykeysight A personalized view into the information most relevant to you. www.lxistandard.org LAN eXtensions for Instruments puts the power of Ethernet and the Web inside your test systems. Keysight is a founding member of the LXI consortium. Three-Year Warranty www.keysight.com/find/ThreeYearWarranty Keysight’s commitment to superior product quality and lower total cost of ownership.