Keysight Technologies Digital Multimeters 34460A, 34461A, 34465A (6½ digit), 34470A (7½ digit) Data Sheet Truevolt DMMs for your next generation of insights
02 | Keysight | Digital Multimeters 34460A, 34461A, 34465A (6½ digit), 34470A (7½ digit) - Data Sheet Table of Contents Measure with Unquestioned Truevolt Confidence...................................................................................... 04 BenchVue Software......................................................................................................................................05 Measure with Unquestioned Truevolt Confidence.........................................................
03 | Keysight | Digital Multimeters 34460A, 34461A, 34465A (6½ digit), 34470A (7½ digit) - Data Sheet Keysight’s NEW Truevolt Digital Multimeters (DMMs) offer a full range of measurement capabilities and price points with higher levels of accuracy, speed, and resolution. Get more insight quickly Truevolt DMM's graphical capabilities such as trend and histogram charts offer more insights quickly.
04 | Keysight | Digital Multimeters 34460A, 34461A, 34465A (6½ digit), 34470A (7½ digit) - Data Sheet Bar meter mode provides the number display along with an analog meter to provide a visual view of your measurements. Histogram mode gives you a statistical view of your measurements. Number mode provides the traditional “digits” view of measurements.
05 | Keysight | Digital Multimeters 34460A, 34461A, 34465A (6½ digit), 34470A (7½ digit) - Data Sheet BenchVue Software Data capture simplified. Click. Capture. Done. BenchVue software for the PC makes it simple to connect, control, capture and view Keysight's DMMs simultaneously with other Keysight bench instruments with no additional programming. Figure 1. See your measurements across instruments in one place to quickly correlate measurement activities and obtain actionable insights.
06 | Keysight | Digital Multimeters 34460A, 34461A, 34465A (6½ digit), 34470A (7½ digit) - Data Sheet Measure with Unquestioned Truevolt Confidence What Truevolt technology means to you: You can measure your real-world signals,not instrument error Noise and injected current: Keysight Truevolt DMMs contribute less than 30% of the injected current than alternatives. Compared to some lower cost alternatives, Truevolt DMMs offer almost 100% less noise.
07 | Keysight | Digital Multimeters 34460A, 34461A, 34465A (6½ digit), 34470A (7½ digit) - Data Sheet Move to The Next-Generation 34401A DMM with 100% 1 Assurance Migrate with confidence: Everything you depend on with the 34401A and more Like most 34401A DMM owners, you rely on your DMM and you trust the answers it gives you. Now, with the Keysight Truevolt 34461A DMM, you can get all of the advantages of the 34401A and more. Now you can get faster answers and have even more confidence in your results.
08 | Keysight | Digital Multimeters 34460A, 34461A, 34465A (6½ digit), 34470A (7½ digit) - Data Sheet Specifications 34460A 34460A accuracy specifications: ± (% of reading + % of range) 1. These specification are compliant to ISO/IEC 17025 for K = 2.
09 | Keysight | Digital Multimeters 34460A, 34461A, 34465A (6½ digit), 34470A (7½ digit) - Data Sheet Specifications 34460A (Continued) Range 2/frequency True RMS AC current 2, 6, 8 100 µA, 1 mA, 10 mA, and 100 mA ranges 3 Hz to 5 kHz 5 to 10 kHz (typ) 1 A range 3 Hz to 5 kHz 5 to 10 kHz (typ) 3 A range 3 Hz to 5 kHz 5 to 10 kHz (typ) Continuity 1 kΩ Diode test 9 5V DC ratio (typ) 24 hours 3 TCAL ± 1 °C 90 days TCAL ± 5 °C 1 year TCAL ± 5° C 2 years TCAL ± 5 °C Temperature coefficient/°C 4 0.10 + 0.
| Keysight | Digital Multimeters 34460A, 34461A, 34465A (6½ digit), 34470A (7½ digit) - Data Sheet Specifications 34461A 34461A accuracy specifications: ± (% of reading + % of range) 1. These specification are compliant to ISO/IEC 17025 for K = 2.
| Keysight | Digital Multimeters 34460A, 34461A, 34465A (6½ digit), 34470A (7½ digit) - Data Sheet Specifications 34461A (Continued) Range 2/frequency True RMS AC current 2, 6, 9 100 µA, 1 mA, 10 mA, and 100 mA ranges 3 Hz to 5 kHz 5 to 10 kHz (typ) 1 A range 3 Hz to 5 kHz 5 to 10 kHz (typ) 3 A range 3 Hz to 5 kHz 5 to 10 kHz (typ) 10 A range 8 3 Hz to 5 kHz 5 to 10 kHz (typ) Continuity 1 kΩ Diode test 10 5V DC ratio (typ) 24 hours 3 TCAL ± 1 °C 90 days TCAL ± 5 °C 1 year TCAL ± 5 °C 2 years TCAL ±
| Keysight | Digital Multimeters 34460A, 34461A, 34465A (6½ digit), 34470A (7½ digit) - Data Sheet Specifications 34465A 34465A accuracy specifications: ±(% of reading + % of range) 1. DC voltage and resistance. Automatic calibration (ACAL) capable.
| Keysight | Digital Multimeters 34460A, 34461A, 34465A (6½ digit), 34470A (7½ digit) - Data Sheet Specifications 34465A (Continued) Temperature PT100 (DIN/ IEC 751) 13 5 kΩ thermistor K, J, T, E, N thermocouples 14 R thermocouples 14 (250 to 1760°C) True RMS AC voltage 15, 16 100 mV, 1 V, 10 V, 100 V, and 750 V ranges 3 Hz to 5 Hz 5 Hz to 10 Hz 10 Hz to 20 kHz 20 kHz to 50 kHz 50 kHz to 100 kHz 100 kHz to 300 kHz True RMS AC current 16, 17 100 µA, 1 mA, 10 mA, 100 mA, 1 A ranges 3 Hz to 5 kHz 5 kHz to
| Keysight | Digital Multimeters 34460A, 34461A, 34465A (6½ digit), 34470A (7½ digit) - Data Sheet Specifications 34465A (Continued) DC and AC current burden voltage at full scale DC current range Burden voltage 1 µA < 0.0011 V 10 µA < 0.011 V 100 µA < 0.11 V 1 mA < 0.11 V 10 mA < 0.027 V 100 mA < 0.27 V 1A < 0.7 V/0.05 V 21 3A < 2.0 V/0.15 V 21 10 A < 0.
| Keysight | Digital Multimeters 34460A, 34461A, 34465A (6½ digit), 34470A (7½ digit) - Data Sheet Specifications 34470A 34470A accuracy specifications: ± (% of reading + % of range) 1. DC voltage and resistance. Automatic calibration (ACAL) capable.
| Keysight | Digital Multimeters 34460A, 34461A, 34465A (6½ digit), 34470A (7½ digit) - Data Sheet Specifications 34470A (Continued) Temperature PT100 (DIN/IEC 751) 13 5 kΩ thermistor K, J, T, E, N thermocouples 14 R thermocouples 14 (250 to 1760 °C) True RMS AC voltage 15, 16 100 mV, 1 V, 10 V, 100 V, and 750 V ranges 3 Hz to 5 Hz 5 Hz to 10 Hz 10 Hz to 20 kHz 20 kHz to 50 kHz 50 kHz to 100 kHz 100 kHz to 300 kHz True RMS AC current 16, 17 100 µA, 1 mA, 10 mA, 100 mA, 1 A ranges 3 Hz to 5 kHz 5 to 10 k
| Keysight | Digital Multimeters 34460A, 34461A, 34465A (6½ digit), 34470A (7½ digit) - Data Sheet Specifications 34470A (Continued) DC and AC current burden voltage at full scale DC current range Burden voltage 1 µA < 0.0011 V 10 µA < 0.011 V 100 µA < 0.11 V 1 mA < 0.11 V 10 mA < 0.027 V 100 mA < 0.27 V 1A < 0.7 V/0.05 V 21 3A < 2.0 V/0.15 V 21 10 A < 0.
| Keysight | Digital Multimeters 34460A, 34461A, 34465A (6½ digit), 34470A (7½ digit) - Data Sheet Measurement Characteristics (for all models except where noted) DC voltage Measurement method: A/D Linearity –– 34460/61A –– 34465A –– 34470A Input resistance –– 0.1 V, 1 V, 10 V range 100 V, 1,000 V range Input bias current Input terminals Input protection True RMS AC voltage Measurement type Keysight patented continuously integrating multi-slope IV A/D converter 0.0002% of reading + 0.0001% of range 0.
| Keysight | Digital Multimeters 34460A, 34461A, 34465A (6½ digit), 34470A (7½ digit) - Data Sheet Operating Characteristics (for all models except where noted) Performance versus measurement speed For DC voltage, DC current, and resistance 1 (34460A and 34461A) Integration time 100 PLC/1.67 s (2 s) 10 PLC/167 ms (200 ms) 1 PLC/16.7 ms (20 ms) 0.2 PLC/3 ms (3 ms) 0.
| Keysight | Digital Multimeters 34460A, 34461A, 34465A (6½ digit), 34470A (7½ digit) - Data Sheet Operating Characteristics (Continued) (for all models except where noted) Noise performance for DC voltage, DC current, and resistance (34465A and 34470A) Integration time 100 PLC/1.67 s (2 s) 10 PLC/167 ms (200 ms) 1 PLC/16.7 ms (20 ms) 0.2 PLC/3 ms (3 ms) 0.06 PLC/1 ms (1 ms) 0.02 PLC/300 µs (300 µs) 0.006 PLC/100 µs (100 µs) 4 0.002 PLC/40 µs (40 µs) 4 0.0.
| Keysight | Digital Multimeters 34460A, 34461A, 34465A (6½ digit), 34470A (7½ digit) - Data Sheet System Speeds (Nom) DC voltage, DC current, resistance 1, 2 Autorange time 3 Maximum internal trigger rate Maximum external trigger rate ASCII readings to bus Single reading transaction rate 4 AC voltage, AC current 5 Autorange time 3 Maximum internal trigger rate Maximum external trigger rate ASCII readings to bus Single reading transaction rate 4 Frequency, period 6 Autorange time 3 Maximum internal trig
| Keysight | Digital Multimeters 34460A, 34461A, 34465A (6½ digit), 34470A (7½ digit) - Data Sheet General Characteristics (for all models except where noted) Line power Power supply Power line frequency Power consumption Environment Operating environment Operating altitude Storage temperature Mechanical Rack dimensions Bench dimensions Weight Regulatory Safety EMC Acoustic noise (nominal) 100/120 (127)/220 (230)/240 ACV ± 10%, CAT II 50/60/400 Hz ± 10% 25 VA Full accuracy for 0 °C to 55 °C Full acc
| Keysight | Digital Multimeters 34460A, 34461A, 34465A (6½ digit), 34470A (7½ digit) - Data Sheet General Characteristics (Continued) (for all models except where noted) Computer interfaces LXI (rev 1.4) 10/100Base-T Ethernet (Sockets, VXI-11 protocol, Web user interface) (Optional on 34460A) USB USB 2.0 (USB-TMC488 & MTP protocol) GPIB Optional GPIB IEEE-488 Language SCPI-1999, IEEE-488.2, 34401A compatible Front-panel USB host port (FAT32) Supports USB 2.
| Keysight | Digital Multimeters 34460A, 34461A, 34465A (6½ digit), 34470A (7½ digit) - Data Sheet Options, Upgrades and Accessories Options and upgrades Option (at purchase) GPB SEC LAN ACC DIG MEM Z54 Upgrades (post purchase) 3446GPBU 3446SECU 3446LANU 3446ACCU 3446DIGU 3446MEMU N/A Applicable models All All 34460A 34460A 34465/70A 34465/70A All Description Upgrade process Add GPIB interface, user-installable Enable NISPOM and file security Enable LAN interface and external triggering Add 34138A
| Keysight | Digital Multimeters 34460A, 34461A, 34465A (6½ digit), 34470A (7½ digit) - Data Sheet Download your next insight Keysight software is downloadable expertise. From first simulation through first customer shipment, we deliver the tools your team needs to accelerate from data to information to actionable insight. –– Electronic design automation (EDA) software –– Application software –– Programming environments –– Productivity software Learn more at www.keysight.