DATA SHEE T Digital Multimeters 34460A, 34461A, 34465A (6½ digit), 34470A (7½ digit) Truevolt DMMs for your next generation of insights
Table of Contents Measure with Unquestioned Truevolt Confidence.........................................................................................4 BenchVue Software........................................................................................................................................5 Measure with Unquestioned Truevolt Confidence.........................................................................................6 Move to The Next-Generation 34401A DMM with 100% Assurance.........
Keysight’s NEW Truevolt Digital Multimeters (DMMs) offer a full range of measurement capabilities and price points with higher levels of accuracy, speed, and resolution. Get more insight quickly Truevolt DMM's graphical capabilities such as trend and histogram charts offer more insights quickly. Both models also provide a data logging mode for easier trend analysis and a digitizing mode for capturing transients.
Bar meter mode provides the number display along with an analog meter to provide a visual view of your measurements. Histogram mode gives you a statistical view of your measurements. Number mode provides the traditional “digits” view of measurements. Measure with Unquestioned Truevolt Confidence Worry about the quality of your design, not the quality of your measurements In a rack or on a bench real-world signals are never flat.
BenchVue Software Data capture simplified. Click. Capture. Done. BenchVue software for the PC makes it simple to connect, control, capture and view Keysight's DMMs simultaneously with other Keysight bench instruments with no additional programming. Figure 1. See your measurements across instruments in one place to quickly correlate measurement activities and obtain actionable insights.
Measure with Unquestioned Truevolt Confidence What Truevolt technology means to you: You can measure your real-world signals,not instrument error Noise and injected current: Keysight Truevolt DMMs contribute less than 30% of the injected current than alternatives. Compared to some lower cost alternatives, Truevolt DMMs offer almost 100% less noise.
Move to The Next-Generation 34401A DMM with 100% 1 Assurance Migrate with confidence: Everything you depend on with the 34401A and more Like most 34401A DMM owners, you rely on your DMM and you trust the answers it gives you. Now, with the Keysight Truevolt 34461A DMM, you can get all of the advantages of the 34401A and more. Now you can get faster answers and have even more confidence in your results. The best news of all? You can migrate from the 34401A to the 34461A without a hassle.
Specifications 34460A 34460A accuracy specifications: ± (% of reading + % of range) 1. These specification are compliant to ISO/IEC 17025 for K = 2.
Specifications 34460A (Continued) Range 2/frequency True RMS AC current 2, 24 hours 3 TCAL ± 1 °C 90 days TCAL ± 5 °C 1 year TCAL ± 5° C 2 years TCAL ± 5 °C Temperature coefficient/°C 4 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.015 + 0.006 0.030 + 0.006 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.015 + 0.006 0.030 + 0.006 0.23 + 0.04 0.23 + 0.04 0.23 + 0.04 0.23 + 0.04 0.23 + 0.
Specifications 34461A 34461A accuracy specifications: ± (% of reading + % of range) 1. These specification are compliant to ISO/IEC 17025 for K = 2.
Specifications 34461A (Continued) Range 2/frequency True RMS AC current 2, 6, 9 100 µA, 1 mA, 10 mA, and 100 mA ranges 3 Hz to 5 kHz 5 to 10 kHz (typ) 1 A range 3 Hz to 5 kHz 5 to 10 kHz (typ) 3 A range 3 Hz to 5 kHz 5 to 10 kHz (typ) 10 A range 8 3 Hz to 5 kHz 5 to 10 kHz (typ) Continuity 1 kΩ Diode test 10 5V DC ratio (typ) 24 hours 3 TCAL ± 1 °C 90 days TCAL ± 5 °C 1 year TCAL ± 5 °C 2 years TCAL ± 5 °C Temperature coefficient/°C 4 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.
Specifications 34465A 34465A accuracy specifications: ±(% of reading + % of range) 1. DC voltage and resistance. Automatic calibration (ACAL) capable. Range 2 DC voltage 100 mV 1V 10 V 100 V 1000 V 9 Resistance 8 100 Ω 1 KΩ 10 KΩ 100 KΩ 1 MΩ 10 MΩ 100 MΩ 1000 MΩ Range 2 24 hours 3 TACAL ± 1 °C 90 days TACAL ± 2 °C 1 year TACAL ± 2 °C 2 years TACAL ± 2 °C Non ACAL 6 Temperature coefficient/°C With ACAL 7 Temperature coefficient/°C 0.0030 + 0.0030 0.0015 + 0.0004 0.0010 + 0.0003 0.0020 + 0.0006 0.
Specifications 34465A (Continued) Temperature PT100 (DIN/ IEC 751) 13 5 kΩ thermistor K, J, T, E, N thermocouples 14 R thermocouples 14 (250 to 1760°C) True RMS AC voltage 15, 16 100 mV, 1 V, 10 V, 100 V, and 750 V ranges 3 Hz to 5 Hz 5 Hz to 10 Hz 10 Hz to 20 kHz 20 kHz to 50 kHz 50 kHz to 100 kHz 100 kHz to 300 kHz True RMS AC current 16, 17 100 µA, 1 mA, 10 mA, 100 mA, 1 A ranges 3 Hz to 5 kHz 5 kHz to 10 kHz (typ) 3 A range 3 Hz to 5 kHz 5 kHz to 10 kHz (typ) 10 A range 4 3 Hz to 5 kHz 5 kHz to 10 kHz (
Specifications 34465A (Continued) DC and AC current burden voltage at full scale DC current range Burden voltage 1 µA < 0.0011 V 10 µA < 0.011 V 100 µA < 0.11 V 1 mA < 0.11 V 10 mA < 0.027 V 100 mA < 0.27 V 1A < 0.7 V/0.05 V 21 3A < 2.0 V/0.15 V 21 10 A < 0.5 V Digitizing 22 Typical performance for these conditions: Sample rate: 50 kHz (Aperture = 20 µS); Sine wave input: Vpeak = Full scale of range; Input frequency: 1 kHz/10 kHz Function: Range Spur-free range SFDR THD + noise SNDR BW (—3 db) DCV: 0.
Specifications 34470A 34470A accuracy specifications: ± (% of reading + % of range) 1. DC voltage and resistance. Automatic calibration (ACAL) capable. Range 2 DC voltage 100 mV 1V 10 V 100 V 1000 V 9 Resistance 8 100 Ω 1 KΩ 10 KΩ 100 KΩ 1 MΩ 10 MΩ 100 MΩ 1000 MΩ Range 2 24 hours 3 TACAL ± 1 °C 90 days TACAL ± 5 °C 1 year TACAL ± 5 °C 2 years TACAL ± 5 °C Non ACAL 6 Temperature coefficient/°C With ACAL 7 Temperature coefficient/°C 0.0030 + 0.0030 0.0010 + 0.0004 0.0008 + 0.0002 0.0020 + 0.0006 0.
Specifications 34470A (Continued) Temperature PT100 (DIN/IEC 751) 13 5 kΩ thermistor K, J, T, E, N thermocouples 14 R thermocouples 14 (250 to 1760 °C) True RMS AC voltage 15, 16 100 mV, 1 V, 10 V, 100 V, and 750 V ranges 3 Hz to 5 Hz 5 Hz to 10 Hz 10 Hz to 20 kHz 20 kHz to 50 kHz 50 kHz to 100 kHz 100 kHz to 300 kHz True RMS AC current 16, 17 100 µA, 1 mA, 10 mA, 100 mA, 1 A ranges 3 Hz to 5 kHz 5 to 10 kHz (typ) 3 A range 3 Hz to 5 kHz 5 to 10 kHz (typ) 10 A range 5 3 Hz to 5 kHz 5 to 10 kHz (typ) Capacit
Specifications 34470A (Continued) DC and AC current burden voltage at full scale DC current range Burden voltage 1 µA < 0.0011 V 10 µA < 0.011 V 100 µA < 0.11 V 1 mA < 0.11 V 10 mA < 0.027 V 100 mA < 0.27 V 1A < 0.7 V/0.05 V 21 3A < 2.0 V/0.15 V 21 10 A < 0.5 V Digitizing 22 Typical performance for these conditions: Sample rate: 50 kHz (Aperture = 20 µS); Sine wave input: Vpeak = Full scale of range; Input frequency: 1 kHz/10 kHz Function: Range Spur-free range SFDR THD + noise SNDR BW (—3 db) DCV: 0.
Measurement Characteristics (for all models except where noted) DC voltage Measurement method: A/D Linearity –– 34460/61A –– 34465A –– 34470A Input resistance –– 0.1 V, 1 V, 10 V range 100 V, 1,000 V range Input bias current Input terminals Input protection True RMS AC voltage Measurement type Keysight patented continuously integrating multi-slope IV A/D converter 0.0002% of reading + 0.0001% of range 0.0001% of reading + 0.0001% of range 0.00005% of reading + 0.
Operating Characteristics (for all models except where noted) Performance versus measurement speed For DC voltage, DC current, and resistance 1 (34460A and 34461A) Integration time 100 PLC/1.67 s (2 s) 10 PLC/167 ms (200 ms) 1 PLC/16.7 ms (20 ms) 0.2 PLC/3 ms (3 ms) 0.02 PLC/300 µs (300 µs) AC voltage, AC current 4, 5 34460A, 34461A, 34465A, 34470A Frequency, period 34460A, 34461A, 34465A, 34470A Frequency, period 34465A, 34470A 1. 2. 3. 4. 5. 6.
Operating Characteristics (Continued) (for all models except where noted) Noise performance for DC voltage, DC current, and resistance (34465A and 34470A) Integration time 100 PLC/1.67 s (2 s) 10 PLC/167 ms (200 ms) 1 PLC/16.7 ms (20 ms) 0.2 PLC/3 ms (3 ms) 0.06 PLC/1 ms (1 ms) 0.02 PLC/300 µs (300 µs) 0.006 PLC/100 µs (100 µs) 0.002 PLC/40 µs (40 µs) 0.0.001 PLC/20 us (20 µs) 4 Digits 1 34465/34470 6½ / 7½ 6½ / 7½ 6½ / 7 6½ / 6½ 6/6 6/6 5/5 5/5 4½ / 4½ Readings/s 0.06 (0.
System Speeds (Nom) DC voltage, DC current, resistance 1, 2 Autorange time 3 Maximum internal trigger rate Maximum external trigger rate ASCII readings to bus Single reading transaction rate 4 AC voltage, AC current 5 Autorange time 3 Maximum internal trigger rate Maximum external trigger rate ASCII readings to bus Single reading transaction rate 4 Frequency, period 6 Autorange time 3 Maximum internal trigger rate Maximum external trigger rate ASCII readings to bus Single reading transaction rate 4 1. 2. 3.
General Characteristics (for all models except where noted) Line power Power supply Power line frequency Power consumption Environment Operating environment Operating altitude Storage temperature Mechanical Rack dimensions Bench dimensions Weight Regulatory Safety EMC Acoustic noise (nominal) 100/120 (127)/220 (230)/240 ACV ± 10%, CAT II 50/60/400 Hz ± 10% 25 VA Full accuracy for 0 °C to 55 °C Full accuracy to 80% RH at 40 °C (non condensing) Full accuracy to 40% RH for 41 °C to 55 °C (non-condensing) U
General Characteristics (Continued) (for all models except where noted) Computer interfaces LXI (rev 1.4) 10/100Base-T Ethernet (Sockets, VXI-11 protocol, Web user interface) (Optional on 34460A) USB USB 2.0 (USB-TMC488 & MTP protocol) GPIB Optional GPIB IEEE-488 Language SCPI-1999, IEEE-488.2, 34401A compatible Front-panel USB host port (FAT32) Supports USB 2.
Options, Upgrades and Accessories (Continued) Accessories Accessories included 34460A Power cord Calibration certificate 34461A, 34138A test lead set with probes, fine tip probes 34465A, SMT grabbers and mini grabber attachments 34470A Power cord IO Libraries CD USB cable Calibration certificate Accessories available 11059A Kelvin probe set 11060A Surface-mount device probe 11062A Kelvin clip set 34131A Transit case 34133A Precision electronic test leads 34134A DC-coupled current probe 34136A High-voltage p