Datasheet

Page 11
Find us at www.keysight.com
Specifications 34461A (Continued)
Range
2
/frequency 24 hours
3
T
CAL
± 1 °C
90 days
T
CAL
± 5 °C
1 year
T
CAL
± 5 °C
2 years
T
CAL
± 5 °C
Temperature
coefficientC
4
True RMS AC current
2, 6, 9
Burden voltage
100 µA, 1 mA, 10 mA, and
100 mA ranges
< 0.011, < 0.11, < 0.05,
< 0.5 V
3 Hz to 5 kHz 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.015 + 0.006
5 to 10 kHz (typ) 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.030 + 0.006
1 A range < 0.7 V
3 Hz to 5 kHz 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.015 + 0.006
5 to 10 kHz (typ) 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.030 + 0.006
3 A range < 2.0 V
3 Hz to 5 kHz 0.23 + 0.04 0.23 + 0.04 0.23 + 0.04 0.23 + 0.04 0.015 + 0.006
5 to 10 kHz (typ) 0.23 + 0.04 0.23 + 0.04 0.23 + 0.04 0.23 + 0.04 0.030 + 0.006
10 A range
8
< 0.5 V
3 Hz to 5 kHz 0.15 + 0.04 0.15 + 0.04 0.15 + 0.04 0.15 + 0.04 0.015 + 0.006
5 to 10 kHz (typ) 0.15 + 0.04 0.15 + 0.04 0.15 + 0.04 0.15 + 0.04 0.030 + 0.006
Continuity
1 k 0.002 + 0.030 0.008 + 0.030 0.010 + 0.030 0.012 + 0.030 0.0010 + 0.0020
Diode test
10
5 V 0.002 + 0.030 0.008 + 0.030 0.010 + 0.030 0.012 + 0.030 0.0010 + 0.0020
DC ratio (typ)
(normalized input accuracy) + (normalized reference accuracy)
Temperature
11
PT100 (DIN/ IEC 751) Probe accuracy + 0.05 °C
5 kΩ thermistor Probe accuracy + 0.1 °C
Frequency: specification ± (% of reading)
12,
13
100 mV, 1 V, 10 V, 100 V, and 750 V ranges
14
3 to 10 Hz 0.100 0.100 0.100 0.100 0.100
10 to 100 Hz 0.030 0.030 0.030 0.030 0.035
100 Hz to 1 kHz 0.003 0.008 0.010 0.010 0.015
1 to 300 kHz 0.002 0.006 0.010 0.010 0.015
Square wave
15
0.001 0.006 0.010 0.010 0.015
Additional gate time errors ±(% of reading)
13
Frequency 1 second 0.1 second 0.01 second
3 Hz to 40 Hz 0 0.200 0.200
40 Hz to 100 Hz 0 0.060 0.200
100 Hz to 1 kHz 0 0.020 0.200
1 kHz to 300 kHz 0 0.004 0.030
Square wave
15
0 0 0
1. For DC: Specifications are for 60-minute warm-up, aperture of 10 or 100 NPLC, and auto zero on. For AC: Specifications are for 60-minute warm-up, slow
AC filter, sine wave.
2. 20% over range on all ranges, except 1,000 V DCV, 750 ACV, 10 A DC, 3 A AC, 10 A AC, and diode test.
3. Relative to calibration standards.
4. Add this for each °C outside TCAL ± 5 °C.
5. Specifications are for sinewave input > 0.3% of range and > 1 mVrms. 750 ACV range limited to 8 x 10 7 Volt–Hz.
6. Low-frequency performance: three filter settings are available: 3 Hz, 20 Hz, 200 Hz. Frequencies greater than these filter settings are specified with no
additional errors.
7. Specifications are for 4-wire ohms function or 2-wire ohms using math null for offset. Without math null, add 0.2 Ω additional error in 2-wire ohms function.
8. The 10 A range is only available on a separate front-panel connector. Add 2 mA base per amp or inputs > 5 A rms.
9. Specifications are for sinewave input > 1% of range and > 10 µA AC.
10. Specifications are for the voltage measured at the input terminals. The 1 mA test current is typical. Variation in the current source will create some
variation in the voltage drop across a diode junction.
11. Actual measurement range and probe errors will be limited by the selected probe. Probe accuracy adder includes all measurement and ITS-90 temperature
conversion errors. PT100 Ro settable to 100 Ω ± 5 Ω to remove the initial probe error.
12. Specifications are for 60-minute warm-up and sine wave input unless stated otherwise. Specifications are for 1-second gate time (7-digits).
13. Applies to sine and square inputs ≥ 100 mV. For 10 mV to < 100 mV inputs, multiply % of reading error x10.
14. Amplitude 10% to 120% of range and less than 750 ACV.
15. Square wave input specified for 10 to 300 kHz.