Datasheet
Page 22
Find us at www.keysight.com
General Characteristics
(for all models except where noted)
Line power
Power supply 100/120 (127)/220 (230)/240 ACV ± 10%, CAT II
Power line frequency 50/60/400 Hz ± 10%
Power consumption 25 VA
Environment
Operating environment Full accuracy for 0 °C to 55 °C
Full accuracy to 80% RH at 40 °C (non
condensing)
Full accuracy to 40% RH for 41 °C to 55 °C
(non-condensing)
Operating altitude Up to 3,000 m
Storage temperature –40 to 70 °C
Mechanical
Rack dimensions (W x H x D): 212.8 mm x 88.3 mm x 272.3 mm
Bench dimensions (W x H x D): 261.2 mm x 103.8 mm x 303.2 mm
Weight 34460A: 3.68 kg (8.1 lb)
34461/65/70A: 3.76 kg (8.3 lb)
Regulatory
Safety EN 61010-1:2010 (3rd Edition)
ANSI/ISA-61010-1 (82.02.01) Third Edition
ANSI/UL 61010-1 Third Edition
CAN/CSA-C22.2 No. 61010-1 Third Edition
EN 61010-2-030:2010 (1st Edition)
ANSI/ISA-61010-2-030 (82.02.03) First Edition
ANSI/UL 61010-2-030 First Edition
CAN/CSA-C22.2 No. 61010-2-030 First Edition
Refer to Declaration of Conformity for current
revisions
Measurement Category II to 300 V
Other non MAINS circuits to 1,000 Vpk
Pollution Degree 2
EMC IEC 61326
EN 61326
CISPR
ICES-001
AS/NZS 2064.1
Refer to Declaration of Conformity for current
revisions
Acoustic noise
(nominal)
35 dBA
Triggering conditions
External input Low–power TTL compatible input
programmable edge triggered
– Delay < 1 µs
– Jitter < 1 µs
– Minimum pulse
width
1 µs
– Maximum rate Up to 1 kHz (34461A), up to 300 Hz (34460A)
Voltmeter complete
output
3.3 V logic output
– Polarity Programmable edge pulse
– Pulse width Approximately 2 µs