Keysight 82357B USB/GPIB Interface User’s Guide
Notices Copyright Notice © Keysight Technologies 2006-2017 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technologies as governed by United States and international copyright laws.
Safety Summary The following general safety precautions must be observed during all phases of operation of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Keysight Technologies assumes no liability for the customer’s failure to comply with these requirements. Safety Notices WARNING CAUTION A WARNING notice denotes a hazard.
Safety Symbols The following symbols on the instrument and in the documentation indicate precautions which must be taken to maintain safe operation of the instrument.
Regulatory Markings The CE mark shows that the product complies with all the relevant European Legal Directives. ICES/NMB-001 indicates that this ISM device complies with Canadian ICES-001. The CSA mark is a registered trademark of the Canadian Standards Association. A CSA mark with the indicators “C” and “US” means that the product is certified for both the U.S. and Canadian markets, to the applicable American and Canadian standards.
General Safety Information WARNING – DO NOT use the device if it appears damaged or defective. – Observe all markings on the device before connecting any wiring to the device. – DO NOT operate the device in the presence of flammable gases or fumes. – DO NOT install substitute parts or perform any unauthorized modification to the device. CAUTION 8 – Use the device with the cables provided. – Repair or service that is not covered in this manual should only be performed by qualified personnels.
Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC This instrument complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste. Product category: With reference to the equipment types in the WEEE directive Annex 1, this instrument is classified as a “Monitoring and Control Instrument” product. The affixed product label is as shown below.
Environmental Conditions This instrument is designed for indoor use only. The table below shows the general environmental requirements for the product. CAUTION Environmental Cond itions Requirements Temperature 0 °C to 55 °C (Operating) –40 °C to +70 °C (Non-operating) Humidity Operating up to 90% at 40 °C (Non-condensing) Non-operating up to 90% at 65 °C (Non-condensing) This product is designed for use in compliance with: – IEC 61010-1:2001/EN 61010-1:2001 – USA: UL61010-1: 2004 – Canada: CSA C22.
Table of Contents Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Safety Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Regulatory Markings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 General Safety Information . . . . . . . . . . . . . . . . . . . . .
Multiple 82357B operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 SRQ operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 Setting Configuration Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 Changing configuration parameters . . . . . . . . . . . . . . . . . . . . . . . . . 36 Changing modes of operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Setting timeout floor values . . . . . . . . .
List of Figures Figure 1-1 Figure 1-2 Figure 1-3 Figure 1-4 Figure 1-5 Figure 1-6 Figure 2-1 Figure 2-2 Figure 3-1 Figure 3-2 Contents of Keysight 82357B USB/GPIB Interface . . .16 IO icon on the Windows taskbar notification area . . .18 82357B Hardware Features . . . . . . . . . . . . . . . . . . . . .19 Connecting the 82357B to Your PC . . . . . . . . . . . . . . .20 Connecting the 82357B to a USB Hub . . . . . . . . . . . .21 Connection from Single GPIB Instrument to the GPIB Port . . . . . . . . . . .
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
Keysight 82357B USB/GPIB Interface User’s Guide 1 Installing the Keysight 82357B This chapter shows a suggested five-step process to install the 82357B and the Keysight IO Libraries Suite, to connect the 82357B to your PC and to program GPIB instruments via the 82357B.
1 Installing the Keysight 82357B Step 1: Checking Your Shipment Your 82357B USB/GPIB Interface shipment should include the items shown in Figure 1-1. If any item is missing or damaged, keep the shipping materials and contact Keysight Technologies. See “Contacting Keysight” on page 50 later in this guide for addresses and telephone numbers.
Installing the Keysight 82357B 1 Step 2: Installing Keysight IO Libraries Suite NOTE You must have Administrator privileges to install Keysight IO Libraries Suite Connection Expert. This section describes how to install the Keysight IO Libraries Suite on your PC. The Keysight IO Libraries Suite is a collection of libraries and utilities that gives you the ability to use your instruments from instrument control software.
1 Installing the Keysight 82357B 4 After the IO Libraries suite is successfully installed, you will see the IO icon on the Windows taskbar notification area.
Installing the Keysight 82357B 1 Step 3: Connecting the 82357B After the Keysight IO Libraries Suite software has been installed, you can connect the 82357B to any USB port on your PC or you can connect the 82357B via standard USB hubs. NOTE If the Keysight IO Libraries Suite software have not been installed on your PC, STOP. Install the libraries (see Step 2: Installing Keysight IO Libraries Suite).
1 Installing the Keysight 82357B Connecting the 82357B to your PC This section shows steps to connect the 82357B to a USB port on your PC or to your PC via a USB Hub. 1 Connect to a USB Port. Make sure the PC is ON and plug the 82357B USB cable into any available USB port on your PC. Do not connect the 82357B to GPIB instruments at this time. Figure 1-4 Connecting the 82357B to Your PC 2 Observe the LEDs. Observe the LEDs on the 82357B for at least 10 seconds.
Installing the Keysight 82357B 1 Connecting the 82357B to a USB hub This section shows steps to connect the 82357B to a USB port on your PC via a standard USB hub. NOTE Any USB hub used with the 82357B MUST be self-powered (must not be bus-powered or powered from the USB bus). Also, be sure to check the applicable USB hub documentation for hub operating parameters, such as power requirements and maximum length of USB cables. 1 Plug the power adapter into the hub and into an electrical outlet.
1 Installing the Keysight 82357B 4 Observe the LEDs on the 82357B for at least 10 seconds. See Chapter 2, “Using the Keysight 82357B” for a description of the normal LED sequence during an initial installation of the 82357B. a Initially, only the red FAIL LED should be ON. After a few seconds, all three LEDs should be ON. When all three LEDs are ON, this shows that the 82357B has been successfully installed.
Installing the Keysight 82357B 1 Step 4: Connecting GPIB Instruments After the 82357B has been installed, the next step is to connect GPIB instruments to the 82357B. This step includes: – Connecting a Single GPIB instrument OR ... – Connecting Multiple GPIB Instruments CAUTION To avoid damage to the connectors, only finger-tighten the connectors. Connecting a single GPIB instrument Figure 1-6 shows connection from a single GPIB instrument to the GPIB port on an 82357B.
1 Installing the Keysight 82357B Connecting multiple GPIB instruments Figure 1-7 shows a typical way to connect three GPIB instruments to an 82357B. When you have made the connections for your system, go to Step 5: Programming via the 82357B. You may want to record the primary GPIB address of each attached instrument for future programming use. NOTE Although Figure 1-7 shows the connection from 82357B to GPIB Instrument 1, the connection can be to any GPIB instrument in the system.
Installing the Keysight 82357B 1 Step 5: Programming via the 82357B After you have connected your GPIB instrument(s) to the 82357B, the next step is to establish communication between your PC and the instruments using Interactive IO. After communication has been established, you can begin programming the instruments using VISA, VISA COM, or SICL. Establishing instrument communication When the Keysight IO Libraries Suite is installed on your PC, an IO utility called Interactive IO is also installed.
1 Installing the Keysight 82357B Figure 1-8 Keysight Interactive IO 5 When communication has been established to each GPIB instrument, you can begin to program the instruments using VISA, VISA COM, or SICL. See the next section, Programming GPIB instruments.
Installing the Keysight 82357B 1 Programming GPIB instruments This section provides an introduction to programming GPIB instruments via the 82357B USB/GPIB interface using the Keysight VISA, VISA COM, and SICL IO Libraries. You can program in various languages/applications, including Visual Basic, Visual C++, Keysight VEE, and National Instruments LabVIEW. See the applicable user’s guide(s), such as the Visual Basic User’s Guide, for programming guidelines.
1 Installing the Keysight 82357B Introduction to IO interface configuration An IO interface consists of a hardware interface and a software interface. The Connection Expert utility is used to associate a unique software interface ID with a hardware interface. The Keysight IO Libraries Suite uses an Interface ID or Logical Unit (LU) Number to identify an interface.
Installing the Keysight 82357B Figure 1-9 Keysight 82357B User’s Guide 1 Typical System Installation - 82357B USB/GPIB Interface 29
1 Installing the Keysight 82357B THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
Keysight 82357B USB/GPIB Interface User’s Guide 2 Using the Keysight 82357B This chapter describes normal operating states and modes for the 82357B and provides the guidelines to use the 82357B, including: – Initial 82357B operating states – Introduction to 82357B operating modes – Single 82357B operation – Multiple 82357B operation – SRQ operation 31
2 Using the Keysight 82357B Initial 82357B Operating States Figure 2-1 shows the sequence of initial operating states when the 82357B is first connected to a USB port on a PC or on a USB hub.
Using the Keysight 82357B 2 Introduction to 82357B operating modes The 82357B has two modes of operation. When only one 82357B is connected to a USB port within a system, we define the feature as the Single Mode Features. When up to four 82357Bs are connected at the same time to USB ports within a system, we define the feature as the Multiple Mode Features.
2 Using the Keysight 82357B Single 82357B operation When an 82357B is first installed, a default VISA Interface ID, SICL Interface ID, Logical Unit (LU) number and GPIB Address are automatically assigned to the serial number associated with this specific 82357B. For example, assume an 82357B with serial number MY12345678.
Using the Keysight 82357B 2 Multiple 82357B operation When two or more 82357Bs are attached to a system at the same time, we define the mode as the multiple mode of operation. In multiple mode of operation, each 82357B is “bound” to its related IO Configuration for that Serial Number. This is a different mode of operation than the single mode of operation in that the configuration is not reused if you replace an 82357B with another 82357B.
2 Using the Keysight 82357B Setting Configuration Parameters This section gives guidelines to change or set various configuration parameters for the 82357B, including: – Changing Configuration Parameters – Changing Modes of Operation – Setting Timeout Floor Value – Setting High-Performance Operation Changing configuration parameters To change the VISA or SICL Interface ID, the LU or GPIB Address, or if you want to check the values of these configuration parameters, highlight the USB/GPIB interface in the
Using the Keysight 82357B Figure 2-2 2 Keysight 82357B (High Speed) Interface - USB/GPIB Changing modes of operation If your system has multiple 82357Bs configured (multiple mode operation), the only way to change from multiple mode of operation to single mode operation is to perform the following: 1 Unplug all 82357Bs from the system. 2 Run Connection Expert (click the IO icon, then click Keysight Connection Expert).
2 Using the Keysight 82357B Setting timeout floor values The 82357B has a default timeout “floor” value that is an internal requirement to ensure reliable USB communication. The 82357B will not allow timeouts less than the floor value. (By default, VISA/SICL timeouts are set to infinite time). To programmatically determine the timeout floor, you can set the timeout to a very small value, such as 1 ms, then query for the actual timeout floor value.
Using the Keysight 82357B 2 Setting 82357B high-performance operation NOTE Changing the T1 delay as described in this section is an ad vanced feature and also requires attention to cable lengths and other system features. Introduction The GPIB transfer rate for 82357B writes using large (>1000 bytes) buffer size is affected by the Data Available (T1) delay time. (The transfer rates are not noticeably affected when the buffer size is <1000 bytes). The default delay time used by the 82357B is 800 ns.
2 Using the Keysight 82357B THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
Keysight 82357B USB/GPIB Interface User’s Guide 3 Troubleshooting the Keysight 82357B This chapter provides troubleshooting guidelines of the 82357B including hardware and software checks and also the service and support information for the 82357B.
3 Troubleshooting the Keysight 82357B Troubleshooting Flowchart Figure 3-1 shows a suggested sequence of steps to diagnose and troubleshoot 82357B problems, based on the LED states. You can use the LED states to help diagnose and troubleshoot the 82357B whenever the LED states do not match expected normal states. See Chapter 2, “Using the Keysight 82357B” for the normal LED sequence when the 82357B is initially connected to a USB port.
Troubleshooting the Keysight 82357B 3 START All LEDs OFF READY FAIL ACCESS Red Fail LED ON READY FAIL ACCESS ALL LEDs ON READY FAIL ACCESS Typical Cause Typical Cause Typical Cause No power on USB bus or device turned off by Windows Plug and Play Manager. Keysight IO Libraries not installed or 82357B USB drivers not installed. Improper Keysight IOLibraries configuration.
3 Troubleshooting the Keysight 82357B Hardware Checks If all LEDs are still off for 10 or more seconds after plugging the 82357B USB cable into a USB port, and all Windows Plug-and-Play Manager activity and the Connection Expert refreshes have ceased, start your troubleshooting sequence by performing hardware checks. If any action taken results in a change in the LED status, go to Software Installation Checks or Software Configuration Checks.
Troubleshooting the Keysight 82357B 3 Check Device Manager You can use the Windows Device Manager to reinstall the 82357B, as required. Go to Device Manager by selecting Start > Control Panel > System > Hard ware > Device Manager. For Windows 8, right-click the bottom-left corner of the Desktop and select Device Manager. From the Device Manager, select 82357 and then Properties. Tab to Driver and click Reinstall Driver.
3 Troubleshooting the Keysight 82357B Software Installation Checks When only the red FAIL LED is still on after 10 seconds, the 82357B has been detected by the host computer. Check suspend/resume operation Some Windows operating systems support Power Management which can suspend the PC while the 82357B is in operation. After a Suspend/Resume cycle, the 82357B may not properly resume operation. In this case, you may need to unplug/replug the USB cable to restore 82357B operation.
Troubleshooting the Keysight 82357B Figure 3-2 3 IO icon on the Windows taskbar notification area a If the IO icon is displayed, click the icon and select About Keysight IO Control to display the version. The version must be 15.0 or greater. b If the IO icon is not displayed, a version of the IO Libraries Suite may still be installed. To verify this, check for Keysight Connection Expert in the Start menu, or the Start Screen (Windows 8).
3 Troubleshooting the Keysight 82357B Software Configuration Checks If all three LEDs remain on for more than 10 seconds after the 82357B is connected to a USB port, the 82357B has been installed. Start your troubleshooting sequence by checking IO Control operation. If any action taken results in a change in the LED status, go to Software Installation Checks or Hardware Checks, where applicable. Checking IO Control operation When the Keysight IO Libraries Suite is installed, an IO Control is created.
Troubleshooting the Keysight 82357B 3 If any of these symptoms occur: 1 Unplug/Replug the 82357B. If unplugging then replugging the 82357B causes the Keysight 82357B USB/GPIB Interface Detected dialog box to appear or the Connection Expert window to refresh, the problem is solved. If not, go to Step 2. 2 Shut down and restart IO Control. Take these steps to shut down and then restart the IO Control. Taking these actions should initialize all attached 82357Bs and display only the green Ready LED.
3 Troubleshooting the Keysight 82357B Service and Support Information There are no user-serviceable parts for the Keysight 82357B USB/GPIB interface. If you suspect a hardware failure for the 82357B, contact Keysight for instructions to return the unit. See the following Contacting Keysight section for telephone numbers/website address.
Keysight 82357B USB/GPIB Interface User’s Guide 4 Product Specifications For the characteristics and specifications of the 82357B USB/GPIB Interface, refer to the datasheet at http://literature.cdn.keysight.com/litweb/pdf/5989-1889EN.pdf.
4 Product Specifications THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
Index B Before You Install, 16 C Changing Configuration Parameters, 36 changing configuration parameters, 36 Changing Modes of Operation, 37 changing modes of operation, 37 check device manager, 45 Check IO Control Operation, 48 Check PC, 44 Check Shipment, 16 Check USB Cables, 44 Check USB Interface, 44 Check USB Scanner, 49 checking shipment, 16, 44 configuration parameters, changing, 36 configuration parameters, setting, 36 connecting 82357A to USB hub, 21 Connecting GPIB Instruments, 23 Connection Exp
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
This information is subject to change without notice. Always refer to the Keysight website for the latest revision. © Keysight Technologies 2006-2017 Edition 10, June 1, 2017 Printed in Malaysia *82357-90003* 82357-90003 www.keysight.