Datasheet

Find us at www.keysight.com Page 14
Performance characteristics (continued)
Trigger system
All Models
Trigger sources
Analog channels, line
6
, external, WaveGen, WaveGen modulation FM/FSK
Trigger modes
Normal (triggered): Requires trigger event for oscilloscope to trigger
Auto: Triggers on selected source or automatically triggers (asynchronously) in absence of a valid trigger event
Single: Triggers only once upon detection of a valid trigger event
Force: Front panel button that forces an asynchronous trigger while in the Normal trigger mode
Trigger coupling
DC: DC coupled trigger
AC: AC coupled trigger, cutoff frequency: ~ 10 Hz
HF reject: High frequency reject, cutoff frequency ~ 50 kHz
LF reject: Low frequency reject, cutoff frequency ~ 50 kHz
Noise reject: Selectable OFF or ON, decreases trigger sensitivity 2X
Trigger holdoff range
60 ns to 10 s
Trigger sensitivity
Trigger level range
6. Line trigger to ≤ 60 Hz.
7. Denotes warranted specifications; All others are typical. Specifications are valid after a 30-minute warm-up period and from ± 10 °C firmware calibration temperature.
8. Input voltage must remain within these limits for proper operation.
EDUX1052A/EDUX1052G
DSOX1202A/DSOX1202G
DSOX1204A/DSOX1204G
Internal
7
Greater of:
0.6 div or 2.5 mV (≤ 10
MHz)
0.9 div or 3.8 mV (10 to 50
MHz)
Greater of:
0.6 div or 2.5 mV (≤ 10 MHz)
0.9 div or 3.8 mV (10 to 70 MHz)
1.2 div or 5 mV (70 to
200 MHz)
External
≤ 10 MHz: 250 mVpp
10 MHz:
2
0 mVpp (1.6 V range)
100 mVpp (8 V range)
10 to 50 MHz: 500 mVpp
10 to 200 MHz:
100 mVpp (1.6 V range)
500 mVpp (8 V range)
EDUX1052A/EDUX1052G
DSOX1202A/DSOX1202G
DSOX1204A/DSOX1204G
Internal
± 6 div from center-screen
± 6 div from center-screen
External
8
± 8 V
± 1.6 V or ± 8 V selectable