Datasheet

Page 6
Find us at www.keysight.com
Electrical Characteristics
Model number Bandwidth
(–3 dB)
Attenuation
ratio
Input C Input R
(Scope and probe)
Max input voltage
(AC RMS)
Scope input
coupling
Scope comp
range
N2870A 35 MHz 1:1 39 pF
(+oscilloscope)
1 MΩ 30 Vrms mains isolated,
60 Vdc
1 MΩ
N2871A 200 MHz 10:1 9.5 pF 10 MΩ 400 V mains isolated
2
300 V CAT II
1 MΩ 10-25 pF
N2872A 350 MHz 10:1 9.5 pF 10 MΩ 400 V mains isolated
2
300 V CAT II
1 MΩ 10-25 pF
N2873A 500 MHz 10:1 9.5 pF 10 MΩ 400 V mains isolated
1, 2
300 V CAT II
1 MΩ 10-25 pF
N2874A 1.5 GHz 10:1 1.8 pF 500 Ω 8.5 V mains isolated 50 Ω
N2875A 500 MHz 20:1 5.6 pF 20 M 400 V mains isolated
2
300 V CAT II
1 MΩ 7-20 pF
N2876A 1.5 GHz 100:1 2.2 pF 5 kΩ 21 V mains isolated 50 Ω
1. 300 Vrms, 400 V (dc + peak ac) mains isolated; 0 V transient overvoltage with InfiniiVision 6000X, Infiniium S and 9000 Series.
2. Maximum pulse rating = 1,650 V (step 0 V to 1,650 V), mains isolated, no overshoot permitted, not applicable to InfiniiVision 6000X, Infiniium S and 9000
Series.
Common to all
Probe ID readout: Compatible with Keysight’s InfiniiVision and Infiniium Series oscilloscopes.
Mechanical Characteristics
Weight (probe only): 48 g
Cable length: 1.3 m
Ground sleeve diameter: 2.5 mm
Environmental characteristics
Temperature
Operating: 0 °C to +50 °C
Non-operating: –40 °C to +70 °C
Altitude
Operating: 2,000 m (6,561 ft)
Non-operating: 15,000 (49,212 ft)
Humidity
Operating: 80% room humidity for temperatures up to 31 °C, decreasing linearly to 40% at 50 °C
Pollution degree: 2