Datasheet

20 | Keysight | InniiVision Oscilloscope Probes and Accessories - Data Sheet
Bandwidth* dc to 100 kHz (–3 dB)
Current range* 100 mV/A:100 mA to
10 A peak
10 mV/A:1 to 100 A peak
Output signal 1000 mV peak max
AC current
accuracy*
Range 100 mV/A
(50 mA to 10 A peak)
Accuracy
Range
3% of reading ±50 mA
10 mV/A
(500 mA to 40 A peak)
Accuracy
Range
4% of reading ±50 mA
10 mV/A
(40 A to 100 A peak)
Accuracy (40 A to 100 A peak)
15% max at 100 A
Noise Range 10 mV/A: 480 μV
Range 100 mV/A: 3 mV
Insertion impedance 0.01 Ω (50/60 Hz)
Maximum working
voltage
600 Vrms CAT II or
300 Vrms CAT III
Maximum common
mode voltage
600 Vrms CAT II or
300 Vrms CAT III
Inuence of adjacent
conductor
< 0.2 mA/A AC
Inuence of
conductor position
< 0.2 mA/A AC
Battery 9 V alkaline (NEDA
1604A, IEC 6LR61)
Low battery Green LED on when
≤ 6.5 V
Battery life 55 hours typical
Bandwidth
(–3 dB)
dc to 50 MHz (1147B)
dc to 100 MHz (N2893A)
Risetime 7 ns or less (1147B), 3.5 ns
or less (N2893A)
Maximum
current
(continuous)
15 A peak, 15 A DC,
10 Arms (when two probes
are used with InniiVision
scope)
30A peak, 30 A DC, 24 Arms
(when one probe is used
with InniiVision 3000X,
5000/6000/7000 scope)
Maximum peak
current
(non-continuous)
30 A peak (when two probes
are used with InniiVision
scope) 32 A peak (when one
probe is used with
InniiVision scope)
Output voltage
rate
0.1 V/A
Amplitude
accuracy
±1% rdg, ±10 mA (dc and 45
to 66 Hz, rated current)
Noise Equivalent to 2.5 mArms or
less (for 20 MHz bandwidth
measuring instrument)
Temperature
coefcient for
sensitivity
±2% or less (within a range
of 0 to 40 °C or 32 to 104 °F)
Effect of
external
magnetic elds
Equivalent to a
maximum of 20 mA (in a dc
to 60 Hz, 400 A/m magnetic
eld)
Maximum rated
power
3 VA (with rated
current)
Maximum input
voltage
300 V CAT I
Diameter of
measurable
:conductors
5 mm dia. (0.2 in dia.)
Probe interface AutoProbe interface
Cable lengths Sensor cable:
Appox. 1.5 m (59.0n in)
Power supply cable:
Appox. 1 m (39.4 in)
Maximum
number of
probes
supported
2
Bandwidth (–3 dB) dc to 2 MHz (N2780B)
dc to 10 MHz (N2781B)
dc to 50 MHz (N2782B)
dc to 80 MHz (N2783L)
dc to 100 MHz (N2783A)
Maximum current
(continuous)
500 A (N2780B)
150 A (N2781B)
30 A (N2782B/83B/83L)
Maximum peak
current (non-
continuous)
700 A peak (N2780B)
300 A peak (N2781B)
50 A peak
(N2782B/83B/83L)
Maximum input
voltage
300 V CAT I
(N2782B/83B/83L)
300 V CAT III, 600 V
CAT II (N2780B/81B)
Output voltage rate 0.01 V/A
(N2780B/81B)
0.1 V/A
(N2782B/83B/83L)
Amplitude accuracy ±1.0 % rdg ±500 mA
(N2780B)
±1.0 % rdg ±100 mA
(N2781B)
±1.0 % rdg ±10 mA
( N2782B)
±1.0 % rdg ±10 mA
(N2783B/83L)
1146B 100-kHz current probe
1147B 50-MHz current probe with
AutoProbe interface
N2893A 100-MHz current probe with
AutoProbe interface
N2780B 2-MHz current probe
N2781B 10-MHz current probe
N2782B 50-MHz current probe
N2783L 80-MHz current probe with
5 m long cable
N2783B 100-MHz current probe
N2779A 3-channel power supply for
N2780B/81B/82B/83B/83L
Clamp-on Current Probes (continued)
Characteristics of the 1146B
current probe
Characteristics of the 1147B/
N2893A current probe
Characteristics of N2780B
Series current probes
Ordering information for
Keysight current probes
Note: Reference conditions 23 ±5 °C,
(73.4 ±41 °C) 20 to 75% relative humidity, dc to 1
kHz, probe zeroed, 1-minute warmup, batteries at
9 V + 0.1 V, external magnetic eld <40 A/m, no dc
component, no external current carrying
conductor, 1 MΩ/100 pF load, conductor centered
in jaw.
* Characteristics marked with asterisks are
specied performance. Others are typical
characteristics.
For more information about the N2780B
Series current probes, refer to the Keysight
N2780B Series current probe data sheet,
literature number 5989-6432EN.
Note: The above specications are guaranteed at
23 ±3 °C (or 73 ±5 °F)