Datasheet

20 | Keysight | Streamline Series P924xA USB Oscilloscopes - Data Sheet
General and environmental
P9241A P9242A P9243A
Power consumption 35 W (typical) power from included external power adapter (no power is used from the USB connection)
+ 3.3 V 3.2 A (typical)
+ 12 V 2.0 A (typical)
Operating Storage
Temperature 0 to 55 °C 40 to 70 °C
Altitude Up to 3000 m Up to 4500 m
Humidity Type tested 95% RH @ 40C non-condensing (on second line) decreasing linearly to 50% RH at 55 °C
Dimensions 177 mm X 335 mm X 50 mm (WxDxH)
Weight 2.5 Kg
Safety UL61010-1 3rd edition, CAN/CSA-C22.2 No. 61010-1-12
Electromagnetic compatibility Meets EMC directive (2004/108/EC), meets or exceeds IEC 61326-1:2005/EN61326-1:2013 (basic)
IEC 61000-4-2/EN 61000-4-2
IEC 61000-4-3/EN 61000-4-3
IEC 61000-4-4/EN 61000-4-4
IEC 61000-4-5/EN 61000-4-5
IEC 61000-4-6/EN 61000-4-6
IEC 61000-4-8/EN 61000-4-8
IEC 61000-4-11/EN 61000-4-11
Canada: ICES/NMB-001:2006
Australia/New Zealand: AS/NZS CISPER 11:2011
System requirements
P9241A P9242A P9243A
Operating systems Windows 10 (64-bit)
Windows 7 SP1 (64-bit)
Processor speed 1 GHz 64-bit (x64)
Memory Available memory 1.5 GB minimum
Disk Available disk space 2.5 GB available hard disk space
Required application Keysight IO Libraries Suite 2018 Update 1.0
Microsoft .NET Framework
Display minimum 1024 x 768, 96 or 120 DPI
Included standard with oscilloscope
P9241A P9242A P9243A
Calibration 2 year
General and Environmental Characteristics
Publication title Publication number
Mask/Waveform Limit Testing For InfiniiVision Series Oscilloscopes - Data Sheet 5990-3269EN
Serial Bus Options for InfiniiVision X-Series Oscilloscopes - Data Sheet 5990-6677EN
Triggering on Infrequent Anomalies and Complex Signals using Zone Trigger - Application Note 5991-1107EN
Using an Oscilloscope Time Gated Fast Fourier Transforms for Time Correlated Mixed Domain Analysis - Application Note 5992-0244EN
Related literature
www.keysight.com/find/usb-scope