Datasheet

6
Electrical Specifications
Measuring rate
Function Times/second
ACV 7
DCV (V or mV) 7
14
Diode 14
Capacitance 4 (< 100 F)
DCA (µA, mA, A) 7
ACA (µA, mA, A) 7
Temperature 7 (single)
Frequency 1 (> 10 Hz)
General Specifications
Display
Dual display (secondary display is intended for temperature function display only)
consists of 4-digit liquid crystal display (LCD) with maximum reading of 11,000 counts.
Automatic polarity indication.
Power consumption
0.22 VA maximum
Battery type and life
Four single standard 1.5 V AAA batteries (Alkaline or Zinc Chloride type); 300 hours
typical
Operating environment
Full accuracy at –10 to 55 °C; and to 80% RH for temperatures up to 30 °C, decreasing
linearly to 50% RH at 55 °C
0 – 2000 meters per IEC 61010-1 2nd Edition CAT III, 1000 V / CAT IV, 600 V IEC
61010-1 2
nd
Edition
Storage compliance
–20 to 70 °C
Safety compliance
IEC 61010-1:2001 / EN61010-1:2001
Canada: CSA C22.2 No. 61010-1:2004
Measurement category
CAT III 1000 V / CAT IV 600 V Overvoltage Protection, Pollution Degree 2
EMC compliance
CISPR 11:1990/EN55011:1990
Canada: ICES-001:2004
Australia/New Zealand: AS/NZS CISPR11:2004

> 90 dB at DC, 50/60 Hz ± 0.1% (1 kΩ unbalanced)

> 60 dB at 50/60 Hz ± 0.1%
Crest factor
< 3.0

0.1 × (specified accuracy) / °C (from –10 to 18 °C or 28 to 55 °C)
Shock and vibration
Tested to IEC/EN 60068-2

193.8 mm x 92.2 mm x 58.0 mm
Weight
450 g with batteries
400 g without batteries
Warranty
Three years for main unit
Three months for standard shipped accessories