Datasheet
Page 7
Find us at www.keysight.com
Capacitance specifications
Range Resolution Accuracy Measuring rate at
full scale
Maximum display
10.000 nF 0.001 nF 1% + 8 4 times/sec. 11000 counts
100.00 nF 0.01 nF 1% + 5
1000.0 nF 0.1 nF
10.000 µF 0.001 µF
100.00 µF 0.01 µF
1000.0 µF 0.1 µF 1 time/sec.
10.000 mF 0.001 mF 0.1 times/sec.
100.00 mF 0.01 mF 3% + 10 0.01 times/sec
U1251B/U1252B temperature specifications
Thermocouple type Range Resolution Accuracy
ºC ºF ºC ºF
K –200 ~ 1372 ºC –328 ~ 2502 ºF 0.1 ºC/0.1 ºF 0.3% + 3 ºC 0.3% + 6 ºF
J
(for U1252A/U1252B)
–210 ~ 1200 ºC –346 ~ 2192 ºF 0.1 ºC/0.1 ºF 0.3% + 3 ºC 0.3% + 6 ºF
U1253B temperature specifications
1 2 3
Thermocouple type Range Resolution Accuracy
ºC ºF ºC ºF
K –200 ~ –40 °C –104 ~ –40 °F 0.1 ºC/0.1 ºF 1% + 3 °C 1% + 5.4 °F
–40 ~1372 °C –40 ~ 2502 °F 0.1 °C /0.1 °F 1% + 1 °C 1% + 1.8 °F
J –210 ~ –40 °C –346 ~ –40 °F 0.1 °C /0.1 °F 1% + 3 °C 1% + 5.4 °F
–40 ~ 1200 °C –40 ~ 2192 °F 0.1 °C /0.1 °F 1% + 1 °C 1% + 1.8 °F
1. The accuracy does not include the tolerance of the thermocouple probe. The thermal sensor plugged into the meter should be placed in the operating
environment for at least an hour.
2. Use the Null function to reduce the thermal effect. Before using the Null function, set the meter to no ambient compensation (0C) mode and keep the
thermocouple probe as close to the meter as possible, avoiding contact with any surface that has a different temperature from the ambient temperature.
3. When measuring temperature with respect to any temperature calibrator, try to set both the calibrator and the meter with an external reference (without
internal ambient compensation). If both the calibrator and the meter are set with internal reference (with internal ambient compensation), deviation may
show between the readings of the calibrator and the meter, due to differences in ambient compensation between the calibrator and the meter.