Datasheet

Electrical Specifications
Temperature specifications
1 - 4
Thermocouple
type
Range Resolution
Accuracy ± (% of reading +
as specified below)
U1271A U1272A U1273A/U1273AX
K
–200 to 1372 °C 0.1 °C
1% of reading
+ 1 °C
1% of reading
+ 1 °C
1% of reading
+ 1 °C
–328 to 2502 °F 0.1 °F
1% of reading
+ 1.8 °F
1% of reading
+ 1.8 °F
1% of reading
+ 1.8 °F
J
–210 to 1200 °C 0.1 °C
1% of reading
+ 1 °C
1% of reading
+ 1 °C
–346 to 2192 °F 0.1 °F
1% of reading
+ 1.8 °F
1% of reading
+ 1.8 °F
1. The specifications above is specified after 60 minutes of warm-up time.
2. The accuracy does not include the tolerance of the thermocouple probe.
3. Do not allow the temperature sensor to contact a surface that is energized above 30 Vrms or 60 V DC. Such voltages pose a shock hazard.
4. The temperature calculation is specified according to the safety standards of EN/IEC-60548-1 and NIST175.
Capacitance specifications
5, 6
Range Resolution
Accuracy ± (% of reading +
counts of least significant digit)
U1271A U1272A U1273A/U1273AX
10 nF 0.001 nF 1 + 5 1 + 5 1 + 5
100 nF 0.01 nF 1 + 2 1 + 2 1 + 2
1000 nF 0.1 nF 1 + 2 1 + 2 1 + 2
10 μF 0.001 μF 1 + 2 1 + 2 1 + 2
100 μF 0.01 μF 1 + 2 1 + 2 1 + 2
1000 μF 0.1 μF 1 + 2 1 + 2 1 + 2
10 mF 0.001 mF 1 + 2 1 + 2 1 + 2
5. Overload protection: 1000 Vrms for short circuits with < 0.3 A current.
6. The accuracy for all ranges is specified based on a film capacitor or better, and after the Null function is used to subtract the test lead resistance and
thermal effect (by opening the test leads).
15 | Keysight | U1270 Series Handheld Digital Multimeters - Data Sheet