Datasheet

Page 10
Find us at www.keysight.com
Digital multimeter specifications (continued)
Maximum reading 10,000 counts with automatic polarity indication
Voltage
1
CAT II 1000 V or CAT III 600 V
Function Range Resolution Accuracy
Input
impedance
(nominal) Test current
Instant continuity
4
Continuous beep when resistance < 10 Ω
5
Resistance 1000.00 Ω
6
0.1 Ω
0.3% + 3
0.5 mA
10.000 kΩ
6
0.001 kΩ 50 μA
100.00 kΩ 0.01 kΩ 4.91 μA
1000.0 kΩ 0.1 kΩ 447 nA
10.000 MΩ 0.001 MΩ 0.8% + 3 112 nA
100.00 MΩ
7
0.01 MΩ 1.5% + 3 112 nA
Capacitance 1000.0 nF 0.1 nF
1.2% + 4
8
10.000 μF 0.001 μF
100.00 μF 0.01 μF
1000.0 μF 0.1 μF
2% + 4
8
10.000 mF 0.001 mF
Frequency
4
100.00 Hz 0.01 Hz
0.03% + 3
1000.0 Hz 0.1 Hz
10.000 kHz 0.001 kHz
100.00 kHz 0.01 kHz
1000.0 kHz 0.1 kHz
1. Only allowed to measure up to CAT III 600 V if referring to GND.
2. In an open connection, the reading shown on the display is noise pickup due to the high input impedance at the input terminal.
3. Only allowed for oating voltage.
4. Denotes typical specications, all others are warranted.
5. Denotes characteristics.
6. The accuracy is specied after the Null function is used to subtract the test lead resistance and thermal effect.
7. RH is specied for < 60%. The temperature coefcient is 0.15 × specied accuracy as > 50 MΩ.
8. The accuracy is based on lm capacitors or better and uses the Relative mode for residual values.
NOTE: Keysight recommends using the U1586B temperature adapter for temperature measurement.
Refer to http://literature.cdn.keysight.com/litweb/pdf/5990-9523EN for more information on the U1586B specifications.
Data logger specifications
Scope and meter logger
Range 1 s/div – 86400 s/div (1 day/div)
Recording time span 8 days
Memory depth 691200 points
Recording mode Continuous (Range will change according to the time elapsed)
Sampling rate 1 sample/s