Datasheet

Page 9
Find us at www.keysight.com
Digital multimeter specifications
Accuracy is given as ± (% of reading + counts of least significant digit) at 23 ± 5 °C, with relative humidity < 80 RH.
AC V specifications are AC coupled, true RMS and are valid from 5% to 100% of range.
Temperature coefficient is given as 0.1 × (specified accuracy) / °C (from 0 to 18 °C or 28 to 50 °C).
Common mode rejection ratio (CMRR) is > 90 dB at DC, 50/60 Hz ± 0.1% (1 kΩ unbalanced).
Normal mode rejection ratio (NMRR) is > 60 dB at 50/60 Hz ± 0.1%.
Maximum reading 10,000 counts with automatic polarity indication
Voltage
1
CAT II 1000 V or CAT III 600 V
Function Range Resolution Accuracy
Input impedance
(nominal) Test current
DCV 100.00 mV
2
0.01 mV 0.1% + 5 > 1 GΩ
1000.0 mV 0.1 mV 0.09% + 5 11.11 MΩ
10.000 V 0.001 V
0.09% + 2
10.10 MΩ
100.00 V 0.01 V
10.01 MΩ
1000.0 V
3
0.1 V 0.15% + 5
ACV 100.0 mV 0.01 mV 1% + 5 (40 Hz to 2 kHz) > 1 GΩ
1000.0 mV 0.1 mV
1% + 5 (40 to 500 Hz)
10.00 MΩ
2% + 5 (500 Hz to 1 kHz)
10.000 V
100.00 V
0.001 V
0.01V
1% + 5 (40 to 500 Hz)
1% + 5 (500 Hz to 1 kHz)
2% + 5 (1 to 2 kHz)
1000.0 V
3
0.1 V
1% + 5 (40 to 500 Hz)
1% + 5 (500 Hz to 1 kHz)
ACV + DC V 100.0 mV
2
0.01 mV 1.1% + 5 (40 Hz to 2 kHz) > 1 GΩ
1000.0 mV 0.1 mV
1.1% + 10 (40 to 500 Hz)
10.00 MΩ
2.1% + 10 (500 Hz to 1 kHz)
10.000 V
100.00 V
0.001 V
0.01 V
1.1% + 7 (40 to 500 Hz)
1.1% + 7 (500 Hz to 1 kHz)
2% + 5 (1 to 2 kHz)
1000.00 V
3
0.1 V
1.2% + 10 (40 to 500 Hz)
1.2% + 10 (500 Hz to 1 kHz)
Diode
4
1 V 0.001 V
0.3% + 2
~0.5 mA
Beeper < ~50 mV, single tone for normal forward-biased diode or semiconductor junction of
0.3 V ≤ reading ≤ 0.8 V
5
Overload protection: 1000 Vrms for short circuit with < 0.3 A
Open voltage: < +2.8 VDC
1. Only allowed to measure up to CAT III 600 V if referring to GND.
2. In an open connection, the reading shown on the display is noise pickup due to the high input impedance at the input terminal.
3. Only allowed for oating voltage
4. Denotes typical speciations, all others are warranted.
5. Denotes characteristics.
6. The accuracy is specied after the Null function is used to subtract the test lead resistance and thermal effect.
7. RH is specied for < 60%. The temperature coefcient is 0.15 × specied accuracy as > 50 MΩ.
8. The accuracy is based on lm capacitors or better and uses the Relative mode for residual values.
NOTE: Keysight recommends using the U1586B temperature adapter for temperature measurement.
Refer to http://literature.cdn.keysight.com/litweb/pdf/5990-9523EN for more information on the U1586B specifications.