Datasheet
03 | Keysight | U2500A Series USB Modular Simultaneous Sampling Multifunction DAQ Devices - Data Sheet
Product characteristics and general specifications
REMOTE INTERFACE
– Hi-Speed USB 2.0
– USBTMC-USB488
1
POWER REQUIREMENT
– +12 VDC (TYPICAL)
– 2 A (MAX) input rated current
– Installation Category II
POWER CONSUMPTION
+12 VDC, 480 mA maximum
OPERATING ENVIRONMENT
– Operating temperature from 0 °C to +55 °C
– Relative humidity at 15% to 85% RH (non-condensing)
– Altitude up to 2000 meters
– Pollution Degree 2
– For indoor use only
STORAGE COMPLIANCE
–20 °C to 70 °C
SAFETY COMPLIANCE
Certied with:
– IEC 61010-1:2001/EN 61010-1:2001 (2nd Edition)
– USA: ANSI/UL 61010-1:2004
– Canada: CSA C22.2 No.61010-1:2004
EMC COMPLIANCE
– IEC 61326-1:2002/EN 61326-1:1997+A2:2001+A3:2003
– CISPR 11: 1990/EN 55011:1990-Group 1 Class A
– Canada: ICES-001:2004
– Australia/New Zealand: AS/NZS CISPR 11:2004
SHOCK AND VIBRATION
Tested to IEC/EN 60068-2
IO CONNECTOR
68-pin female VHDCI Type
DIMENSION (W × D × H)
Module dimension:
– 120.00 mm × 182.40 mm × 44.00 mm (with plastic casing)
– 105.00 mm × 174.54 mm × 25.00 mm (without plastic casing)
Terminal block dimension:
– 103.00 mm × 85.20 mm × 42.96 mm
WEIGHT
– 565 g (with plastic casing)
– 400 g (without plastic casing)
Product outlook and
dimensions
Rear view
Top view
Standard shipped accessories
– AC/DC Power adapter
– Power cord
– USB extension cable
– L-Mount kit (used with modular
product chassis)
– Keysight USB Modular Products Quick
Start Guide
– Keysight Measurement Manager for
U2500A Series Quick Start Guide
– Keysight USB Modular Products
Reference CD-ROM
– Keysight Automation-Ready CD-ROM
(contains the Keysight IO Libraries
Suite)
– Certicate of Calibration
Optional accessories
– U2901A Terminal block and SCSI-II
68-pin connector with 1-meter cable
– U2902A Terminal block and SCSI-II
68-pin connector with 2-meter cable
120.00 mm
182.40 mm
44.00 mm
1. Compatible with Microsoft Windows operating systems only. Requires a direct USB connection to the
PC so the appropriate driver can be installed in the USB DAQ module.
Front
view