DATA SHEE T Keysight U3606B Multimeter I DC Power Supply Convenient and Full Featured. One-box source-and-measure device.
Introduction Features –– Combination of a 5.5 digit digital multimeter and 30-W power supply in a single unit –– 10 DMM measurements, including 4-wire miliohm measurement –– 8 built-in math functions –– OVP and OCP for load protection –– Ramp and scan function, and built in square-wave output –– USB 2.
Take a closer look For Multimeter: For DC power supply: Full-featured 5 ½ digit DMM Full-featured 30-W DC supply – 120,000 counts resolution – Low error rate of up to 0.025% basic DCV accuracy – 10 measurement functions (DCV, ACV, DCI, ACI, 2- and 4-wire resistance, frequency, continuity, diode, capacitance) – 8 built-in math functions – 4 -wire milliohm measurement with 0.
Digital multimeter specifications Specification assumptions: – Specifications stated are after 60 minutes of warm-up and for 5½-digit resolution – One-year calibration cycle, with calibration temperature of 23 °C ± 2 °C – Operating temperature: 18 to 28 °C (64.4 to 82.4 °F) – Accuracy is expressed as ± (% of reading + % of range) – Temperature coefficient: Add [0.1 x (the applicable accuracy) / °C] for 0 to 18 °C and 28 to 55 °C. Unless stated otherwise.
Digital multimeter specifications (continued) DC specifications (continued) Table 1. DC accuracy specifications ± (% of reading + % of range) 24 hours2 23 ºC ± 1 ºC 90 days 23 ºC ± 5 ºC 1 year 23 ºC ± 5 ºC Temperature coefficient/ºC 0 to 18 ºC 28 to 55 ºC Function Range1 Resolution Test current or burden voltage Low-resistance5 100 mΩ 1000 mΩ 10 Ω 100 Ω 1000 Ω 0.01/0.001 mΩ 0.1/0.01 mΩ 1/0.1 mΩ 10/1 mΩ 0.1/10 mΩ 1,0000 A 100.00 mA 100.00 mA 10.00 mA 10.00 mA — — — — — — — — — — 0.25 + 0.05 0.
Digital multimeter specifications (continued) AC specifications Table 2. AC accuracy specifications ± (% of reading + % of range) Function Frequency range 1 year 23 ºC ± 5 ºC Temperature coefficient/ºC 0 to 18 ºC 28 to 55 ºC 20 to 45 Hz 45 Hz to 10 kHz 10 to 30 kHz 30 to 100 kHz4 1 + 0.1 0.2 + 0.1 1.5 + 0.3 5 + 0.3 0.02 + 0.02 0.02 + 0.02 0.02 + 0.02 0.02 + 0.02 1.00000 V to 750.00 V 5, 6 20 to 45 Hz7 45 Hz to 10 kHz 10 to 30 kHz 30 to 100 kHz4, 8 1 + 0.1 0.2 + 0.1 1 + 0.1 3 + 0.2 0.02 + 0.02 0.
Digital multimeter specifications (continued) Frequency specifications Table 3. Frequency accuracy specifications ± (% of reading + % of range) Function Range Frequency range Voltage path: 100 mV to 750 V < 2 Hz < 20 Hz 20 Hz to 100 kHz 100 to 300 kHz 0.18 + 0.003 0.04 + 0.003 0.02 + 0.003 0.02 + 0.003 0.005 0.005 0.005 0.005 Current path: 10 mA to 3 A < 2 Hz < 20 Hz 20 Hz to 10 kHz 0.18 + 0.003 0.04 + 0.003 0.02 + 0.003 0.005 0.005 0.005 Frequency1 1.
Digital multimeter specifications (continued) Duty cycle and pulse width specifications 3 Table 6. Duty cycle and pulse width resolution and accuracy Function Range Resolution Duty cycle 100.000% Pulse width 199.999 ms2 1999.99 ms2 Accuracy of full scale 0.001% 0.3% + 0.2% per kHz 0.001 ms 0.01 ms Duty cycle/frequency Duty cycle/frequency 1 1. The range is from {10 µs × frequency × 100%} to {[1 – (10 µs × frequency)] × 100%}. For example, a 1 kHz signal can be measured from 1% to 99%. 2.
Digital multimeter specifications (continued) Supplemental characteristics DC voltage Measurement method Sigma Delta A-to-D converter Maximum input voltage 1000 Vdc on all ranges Input impedance 10 MΩ ± 2% range (typical) in parallel with capacitance < 120 pF Input protection 1000 Vrms on all ranges Response time Approximately 0.15 s when the displayed reading reaches 99.
Digital multimeter specifications (continued) Supplemental characteristics (continued) AC current Measurement method AC coupled true rms Maximum input current 10 mA to 3.0 A DC or AC rms1 Burden voltage and shunt resistance – – – – Input protection Protected with 3.15 A/500 V, FF fuse Crest factor For < 5:1 errors included. Limited by the peak input and 100 kHz bandwidth. Maximum 3.0 at full scale. Peak input 300% of range. Limited by maximum input. Response time Approximately 2.
Digital multimeter specifications (continued) Supplemental characteristics (continued) Diode Measurement method 0.83 mA ± 0.2% constant current source Open circuit voltage < +5.0 Vdc Audible tone – Continuous beep when level is below +50 mV DC – Single tone for normal forward-biased diode or semiconductor junction where 0.3 V ≤ reading ≤ 0.8 V Input protection 1000 Vrms on all ranges, < 0.
DC power supply specifications Safety considerations The U3606B is a safety class I instrument, which means it has a protective earth terminal. The terminal must be connected to an earth ground through a power source with a 3-wire ground receptacle. The DC power supply performance specifications are listed in the following pages. Specifications are warranted in the temperature range of 0 to 55 °C with a fix resistive load.
DC power supply specifications (continued) Performance specifications (continued) Table 8. DC power supply performance specifications Specifications Parameter S1S2 S1 S1m S2 S2m Output ratings AUTO 30 V/1 A 100 mA/30 V 8 V/3 A 1000 mV/3 A Voltage < 3 mV + (6 mV/A) < 3 mV + (6 mV/A) < 3 mV + (6 mV/A) < 3 mV + (6 mV/A) < 0.3 mV + (6 mV/A) Current < 0.03% + 0.3 mA < 0.03% + 0.3 mA < 0.03% + 0.03 mA < 0.03% + 0.3 mA < 0.03% + 0.3 mA Voltage < 0.01% + 3 mV < 0.01% + 3 mV < 0.
DC power supply specifications (continued) Supplemental characteristics Table 9. DC power supply supplemental characteristics Characteristics Parameter S1 S1m S2 S2m Output ratings 30 V/1 A 100 mA/30 V 8 V/3 A 1000 mV/3 A Voltage CV: 31.500 V OC: 1.05 A OCP: 1.1 A — CV: 8.4 V OC: 3.15 A OCP: 3.3 A CV: 1050 mV OC: 3.15 A OCP: 3.3 A Current CC: 1.05 A OV: 31.500 V OCP: 33.000 V CC: 105 mA OV: 31.500 V OCP: 33.000 V CC: 3.15 A OV: 8.4 V OCP: 8.8 V — Voltage 0.005% + 0.5 mV — 0.005% + 0.
DC power supply specifications (continued) Square-wave output characteristics Table 10. Square-wave output characteristics Parameter Range Characteristics Amplitude accuracy ± (offset) S1 (30 V/1 A) and S1S2 (AUTO) 0.2 V S2 (8 V/3 A) and S1S2 (AUTO) 0.2 V S1 (30 V/1 A) 1 mV Amplitude resolution S2 (8 V/3 A) 1 mV Frequency accuracy ± (% of frequency setting + offset) (27 steps1) 0.005% + 0.01 Hz Frequency resolution — 0.01 Hz Duty cycle accuracy ± (% of duty cycle setting) (256 steps: 0.
DC power supply specifications (continued) Sweep characteristics Table 11. Scan output characteristics Scan Constant voltage Range S1 S2 S2m S1 S1m S2 Maximum amplitude1 31.500 V 8.400 V 1050.0 mV 1.0500 A 105.00 mA 3.1500 A Step Constant current 1 step to 100 steps 1 step to 100 steps 1 s to 99 s 1 s to 99 s Dwelling time 1. Amplitude start position is fixed at 0 (V or A) by default. Table 12.
Product characteristics Power supply – Universal 100 Vac to 240 Vac ±10% – AC line frequency of 45 to 66 Hz; 360 to 440 Hz for 100/120 V operation Power consumption 150 VA maximum Current input fuse 3.15 A, 500 V FF fuse (on front panel) Display Highly visible vacuum-fluorescent display (VFD) Operating environment – – – – – Storage compliance –40 to 70 °C Safety compliance – IEC 61010-1:2001/EN 61010-1:2001 (2nd Edition) – Canada: CAN/CSA-C22.2 No.
Optional accessories U8201A Combo Test Lead Kit 34190A Rack Mount Kit U8202A Electronic Test Lead Kit (for DMM function) 34133A Precision Electronic Test Leads (for DMM function) 34330A Current Shunt (30 A) (for DMM function) 34136A 40 kV high-voltage probe (for DMM function) 11059A Kelvin Probe Set and 11062A Kelvin Clip Set (for DMM function) E3600A-100 Test Lead Kit (for DC power supply function) Learn more at: www.keysight.