Datasheet

13 | Keysight | U1280 Series Handheld Digital Multimeters - Data Sheet
Temperature specification for U1281A / U1282A
2,3,4,5
Thermal type Range Resolution Accuracy ± (% of reading +
as specified below)
K
–200 °C to 1372 °C 0.1 °C 1 % + 1 °C
–328 °F to 2502 °F 0.1 °F 1 % + 1.8 °F
J
1
–210 °C to 1200 °C 0.1 °C 1 % + 1 °C
–346 °F to 2192 °F 0.1 °F 1 % + 1.8 °F
Notes
1. Only for U1282A
2. The specification above is specified after 60 minutes of warm-up time.
3. The accuracy does not include the tolerance of the thermocouple probe.
4. Do not allow the temperature sensor to contact a surface that is energized about 30 Vrms or 60 V DC.
Such voltages pose a shock hazard.
5. The temperature calculation is specified according to the safety standards of EN/IEC-60548-1 and
NIST 175.
Capacitance specification for U1281A / U1282A
1,2,3
Range Resolution Accuracy ± (% of reading +
counts of least significant digit)
10 nF 0.001 nF 1.0 % + 5
100 nF 0.01 nF 1.0 % + 5
1 µF 0.0001 µF 1.0 % + 5
10 µF 0.001 µF 1.0 % + 5
100 µF 0.01 µF 1.0 % + 5
1 mF 0.0001 mF 1.0 % + 5
10 mF 0.001 mF 1.0 % + 5
100 mF 0.01 mF 2.5 % + 10
Notes
1. Overload protection: 1000 Vrms for short circuits with < 0.3 A current.
2. The accuracy for all ranges is specified based on a film capacitor or better, and after the Null
function is used to subtract the test lead resistance and thermal effect (by opening the test leads).
3. The maximum display is 1200 counts / 12000 counts selectable