Datasheet
08 | Keysight | U1730C Series Handheld LCR Meters – Data Sheet
U1731C/U1732C/U1733C Electrical Specications
Source Impedance of Inductance Measurement
Typical source impedance
Range U1731C/U1732C/U1733C U1732C/U1733C U1733C
100 Hz 120 Hz 1 kHz 10 kHz 100 kHz
20 μH NA NA NA 100 Ω 100 Ω
200 μH NA NA 100 Ω 100 Ω 100 Ω
2000 μH 100 Ω 100 Ω 100 Ω 100 Ω 100 Ω
20 mH 100 Ω 100 Ω 100 Ω 100 Ω 100 Ω
200 mH 100 Ω 100 Ω 100 Ω 1 kΩ 1 kΩ
2000 mH 100 Ω 100 Ω 1 kΩ 10 kΩ 1 kΩ
20 H 1 kΩ 1 kΩ 10 kΩ 10 kΩ 1 kΩ
200 H 10 kΩ 10 kΩ 100 kΩ 100 kΩ NA
2000 H 100 kΩ 100 kΩ 100 kΩ NA NA