Why do you need to do a cleaning job after each
measurement?
Because infrared temperature adopts highly sensitive
technique to detect the temperature of the target object,
not only will any earwax and dust paste affect the
measurement accuracy, more likely to cause bacteria
infection. So we suggest that you should do cleaning
job well as the picture shows after you use every time.
Warm prompt:
12
Frequently Asked Questions and Solutions
1. (When the measurement probe is
not properly placed in the ear canal
or if the measurement distance is
too far during the measurement
process, the measurement result
may be low.)
2. When measurement probe is
dirty, the measurement value may
be low; we should use a cotton
swab to dip alcohol to wash the
measurement probe properly.
Screen
display
Reasons
Solutions
HI
When the target object
temperature is higher than
measurement range which
the ear temperature is more
than 43.0℃, the LCD screen
will show "HI" indication.
Lo
When the target object
temperature is lower than
measurement range which
the ear temperature is less
than 34℃, the LCD screen
will show "Lo" indication.
Er.H
This product has an upper
limit of operating temperature
of 40°C. When the ambient
temperature exceeds this
temperature point, the LCD
screen will display an error
message “Er.H".
When you operate this product,
the environment temperature
cannot be greater than 40℃.
Er.L
This product has a lower limit
of operating temperature of
10°C. When the ambient
temperature exceeds this
temperature point, the LCD
screen will display an error
message “Er.L”.
When you operate this product,
the environment temperature can
not be lower than 10℃.
The measurement
temperature is low.
The measurement
position is not
correct.
There are dirt stuff
in the sensor or ear
canal.
Please clear the
dirt before the
measurement.
Measure the
temperature correctly
according to the
instructions.
Big temperature
fluctuations with
continuous
measurement.
The measurement
interval is too small.
The interval for each
measurement should
be above 10 seconds.
Troubleshooting
Phenomenons Reasons
Solutions
When the power
is turned on, the
screen can not
be displayed.
The battery is
exhausted.
Replace the new
battery.
The battery
polarity is wrong.
The battery polarity
is the same as the
battery case.
11
When the operation is wrong, if the sound is turned on, then the sound will
the value and remind at this moment.(A short beep will be heard.)read
Err
When the environment
temperature changes rapidly
about 5 degrees, it will show
“ Err ” during measurement
in object temperature mode,
then shutdown automatically.
When it shows “Err”, please place
this product and keep steady in
current environment for 30 minutes
before next measurement.