User's Manual
13
14
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Group 1
Class B
N.A.
N.A.
Guidance and manufacturer’s declaration–electromagnetic
emissions
Emissions
Compliance
Electromagnetic
environment-guidance
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic
emissions
IEC 61000-3-2
Appendix 1 Guidance and Manufacturer Declaration Tables
The Model PG-IRT1603 is intended for use in
the electromagnetic environment specified below. The customer or
the user of the Model PG-IRT1603 Infrared Thermometer should
assure that it is used in such an environment.
Infrared Thermometer
The
Thermometer uses RF energy only
for its internal function. Therefore, its
RF emissions are very low and are
not likely to cause any interference in
nearby electronic equipment.
Model PG-IRT1603 Infrared
The
Thermometer is used in home and
it’s powered by DC 3V
Model PG-IRT1603 Infrared
Guidance and manufacturer’s declaration – electromagnetic
immunity
The is intended for use in
the electromagnetic environment specified below. The customer or
the user of the Model PG-IRT1603 Infrared Thermometer should
assure that it is used in such an environment.
Model PG-IRT1603 Infrared Thermometer
Immunity
test
Electrostatic
discharge
(ESD)IEC
61000-4-2
IEC 60601
test level
±8 kV contact
±2 kV, ±4 kV,
±8 kV, ±15KV
air
Power
frequency
(50/60 Hz)
magnetic
field IEC
61000-4-8
30 A/m,
50/60Hz
NOTE U is the a.c. mains voltage prior to application of the test level
T
Compliance
level
Electromagnetic
environment-guidance
±8 kV contact
±2 kV, ±4 kV,
±8 kV, ±15 KV
air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
30 A/m,
50/60Hz
Power frequency
magnetic fields should
be at levels characteristic
of a typical location in a
typical commercial or
hospital environment.
Specifications of the product
Product name:
Model number: PG-IRT1603
Product appearance dimensions: 31×175×72mm
Product weight: about 77g (except battery)
Measuring range: 34.0°C-43.0 .(93.2-109.4 )
Object temperature: 0- 93.2 (32-199.7 )
Resolution ratio: 0.1 /
Measurement Location: laboratory
Accuracy: (35.0 ~42.0 )±0.2 ,(95.0~107.9 )
±0.4 ,other temperature±0.3 .
Operation temperature: 10.0 ~40.0 (50.0 ~104.0 ),
relative maximum humidity 15%RH~93%RH
Atmospheric pressure:70kPa~106kPa
Transportation/storage temperature: -25℃~55
13 ~131 , relative maximum humidity 0%RH~93%RH
Atmospheric pressure: 50kPa~106kPa
Display screen: LCD display screen, 4 bit numbers and
special icons.
Sound: when you turn on the product and ready to measure,
short beep will be heard.
Infrared Ear/Forehead Thermometer
°C °F
°C °F
°C °F
°C °C °C °F
°F °C
°C °C °F °F
°C
(- °F °F)
a
The measurement is finished with a long beep.
System error or fault: short beeps for three times.
Fever alert: short beeps for ten times come with
urgency.
Memory: in memory mode, it can record nine temperature
numbers.
Automatically shut down: if no operation for 30 seconds,
will shutdown automatically.
Battery: two pieces of 1.5V AAA (number seven)batteries
(alkaline batteries are recommended to use).
Period of use: five years
it
Packing parts list:
1. Main body
2. Product manual
1603 Forehead mode:
Clinical bias, Dcb: 0.078
Limits of Agreement, LA:0.243
Clinical Repeatability, σr:0.069
The reference body site: forehead
Measuring site: forehead