Owner's manual

40
15 SERVICE MODE
To access service mode, you need to wake up LOGBOX SD from sleep mode. You will type CTRL+BREAK
on your serial line once. If there is no response after about 5 seconds, try again (data logger is serial
line sensitive only for about 50% of time due to low power operation). Then, your communication
software must be set for communication speed 115200bps, as service mode is accessible with this
speed only. You will receive following answer from the LOGBOX SD:
LOGBOX SD (C) Physicus. Version T2.07.217 Jun 19 2007
Service mode started.
>
Now all setup commands are available.
Summary of setup commands:
? displays current setup
?
LOGBOX SD (C) Physicus. Version T2.07.217 Jun 19 2007
LogMode: CIRCULAR
LogAddress: 91336
CircEndAddress: 131040
LogInterval: 5 minutes
SdWrittingInterval: 10 minutes
SdFileClose: 24 hours
Delay: 0
BaudRate: 115200
Temperature[mV]: 1083.8, AVcc[mV]: 3106.4
ReferenceVoltage: 2500.000 mV
DateTime: 2005-03-11 11:22:56
OK
T time setup with syntax HH MM SS (HH-hour, MM-minute, SS-second)
T
Enter HH MM SS
19 09 10
You've entered 19:09:10
<Y> = set, else = quit
Y
SET.
D date setup with syntax YYYY MM DD (YYYY-year, MM-month, DD-day)
D
Enter YYYY MM DD
2005 11 27
You've entered: 2005-11-27
<Y> = set, else = quit
Y
SET.
0
Clear all logging memory