User's Manual

2-4 System Setup
Place maintenance module back into case, flip RF shield back over module and replace lid. Save
case for future use.
Charging the Equipment Before Use
Place the Integral Wafer into the carrier station shipped with the unit to maintain the charge on the
internal batteries. Please see Using the Charge and Communication Carrier Station for more
information on inserting and removing the wafer to and from the carrier stations.
Plug the laptop into the AC power supply and fully charge the laptop battery.
Setting Up the Software
The Integral Wafer system normally includes a SensArray qualified laptop computer that has been
prepared with all needed software to setup, run, and retrieve data from the Integral Wafer.
However, if you are upgrading an existing Thermal MAP 3 system, please follow these directions.
Before using the Integral Wafer, the USB driver for the carrier stations and wafer, the Integral
Wafer Controller Software, and the wafer calibration files must be loaded onto the host computer.
The software has been tested under Windows 2000 and Windows XP.
1. Place the Integral Wafer CD in the host computer CD drive.
2. Using Windows Explorer or My Computer, navigate to the
\INTEGRATEDWAFER.INSTALL folder and run INtegratedWaferController v2.x.x.exe.
Follow the onscreen directions of the installer program. Do not change any default
directories; this could cause problems when the program is run.
3. Connect the USB cable between the host computer and the carrier station. Wait until the
Found New Hardware dialog begins.
4. Follow the Windows directions on screen to complete the new hardware setup. When
requested to locate the driver for the new hardware, select the CD-ROM drives option
and click Next. The wizard will automatically locate the correct driver. There are two
drivers that will load, just follow the onscreen directions. If you are using Windows XP,
ignore the “this is not a Windows certified driver” warning.
If you are installing Thermal MAP on your computer for the first time, follow
the directions in step 5.
If you are installing the Integral Wafer on a system that already has the full
Thermal MAP program, do not install the Analysis only version found on the
Integral Wafer CD. This will cause your existing version to lose the Acquisition
feature. Skip step 5 and use the provided upgrade Thermal MAP CD to install
the full version. Insert the CD and follow the onscreen directions. Then continue
with Step 6.