User's Manual

4-12 Using the Carrier Station Controller
12. Click on the Enable Measure Delay check box to allow setting a measurement delay.
Click on the down arrow on the drop-down box and select the amount of delay desired
before the measurement starts. You can only select the fixed values (from 1 minute to 1
hour) in the drop-down list, typing a custom time value will not work.
Figure 4-16. Setting Measurement Delay
NOTE: Using the "Measurement Delay" prevents data recording until you have time to physically
place the Integral Wafer into the system to be measured. No data is recorded during the
"Measurement Delay" period. However, the battery is still running.
13. Click the Start Survey button.
14. The Start Survey button will change to read Abort Survey. Pressing this button before
the wafer has left the dock will cancel the survey and reset to the Start Survey. The
parameters for the survey may then be changed and the survey started once again.
Once the wafer has left the carrier station, the button will dim and no longer function. If
the carrier station is unplugged from computer, the program screen reverts to the
message ‘Looking for iWafer System…”.
15. Move the carrier station into position for transferring the wafer into the test chamber.
Non-contact Integral Wafer User Manual