Instructions / Assembly

15
TEMPERATURE
1. Remove leads from meter and slide the temperature probe input
switch
18
to close jacks and expose the K-Type thermocouple
ports.
2. Insert K-type thermocouple into port T1
17
(observe polarity
markings on thermocouple and meter), and rotate function
selector switch
2
to the Temperature setting.
NOTE: The meter defaults to Fahrenheit scale in this mode. To
enter Celsius scale, long-press the SEL button
9
. Ensure that the
appropriate icon (either
°F
or
°C
) appears on the display.
NOTE: The meter may be set to default to the Celsius scale by
powering-ON the meter from the OFF position with the Data Hold
& Backlight button
6
depressed. To re-set the default to the
Fahrenheit scale repeat the power-ON sequence.
3. To measure temperature, make contact between the
thermocouple tip and the object being measured. When
thermocouple tip and object are in thermal equilibrium, the
measurement on the display will stabilize. The meter will auto-
range to display the measurement in the most appropriate range.
Remove thermocouple before switching meter to other
measurement functions.
T
he thermocouple included with the original purchase
is suitable for temperatures below 356°F / 180°C only.
To measure higher temperatures, a K-type thermocouple
with the appropriate measurement range should be used.
NOTE: The main display shows Frequency and the secondary display
shows Duty-Cycle.
2. Measure by connecting test leads across the circuit.
OPERATING INSTRUCTIONS
Black lead Red lead
K-Type Thermocouple