Chapter 3
By registering the AeroDR detector to the CS-7, it is
possible to move it between different radiology rooms.
3.4.1 Registering the AeroDR
Detector
1
Turn the power of the AeroDR detector
ON.
2
Insert the AeroDR detector into the regis-
tration AeroDR battery charger.
Registration process will start.
3
Once
the reg
istration is completed, a
buzzer will sound on the AeroDR battery
charger and the status LED will go from
slow ashing to lit (green).
HINT
Charging will also start when the AeroDR detector
is inserted into the AeroDR battery charger, and the
AeroDR battery charger feeding LED will light (blue).
CAUTION
Once it is registered in a new radiology room, it will
not be usable in the previous radiology room. Perform
the registration operation again when returning to a
previous radiology room.
The AeroDR detector cannot be registered with the
charging AeroDR battery charger.
4
Registration is completed when "Registra-
tion completed" is displayed on the CS-7.
3.4.2 Selecting the AeroDR
Detector
The AeroDR detector to be used is selected
as follows depending on the number of reg-
istered AeroDR detectors.
Number of regis-
tered AeroDR
detectors
Selection method
1
AeroDR detector is selected auto-
matically when that AeroDR detector
is ready to be selected (there is no
problem with the battery, wireless
connection, etc.).
The AeroDR detector is not selected
if it is not inserted into a bucky that
matches the order information.
Multiple
The AeroDR detector that was
selected immediately beforehand
will be selected automatically. Also,
if there is an AeroDR detector
inserted in the bucky that matches
the order information, that AeroDR
detector will be selected. If there is
no AeroDR detector that matches
the order information, no AeroDR
detector will be selected.
To change the selected AeroDR detector to another
AeroDR detector, turning the selection switch of the
AeroDR detector ON enables manual selection.
HINT
A selected AeroDR detector will have its status LED lit
(green).
The AeroDR detector can also be selected manually
from the CS-7. For details, refer to the "CS-7 Operation
Manual".