76
52 Various problems and countermeasures
Ifthefollowingproblemsoccurwithanyofthesedevices,consulttherespectivereferencesforcountermeasures.
IMPORTANT
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
• Afterperformingcountermeasures,iftheproblemdoesnotgoaway,contactKonicaMinoltatechnicalrepresentatives.
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
• Whenanerrormessagehasbeendisplayedintheimageprocessingcontroller,checktheerrordescriptionandcountermea-
sureslistedinthe"OperationManual"oftheimageprocessingcontroller.
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
521 AeroDR Detector
Status Error description Corrective actions
TheAeroDRDetectordoesnot
startup.
Powerdoesnotgoonevenwhenthepower
switchispressedfor2secondsorlonger.
Thebatterymightbeout.Chargeitasfollows.
Then,startit.
• InsertingitintotheAeroDRBatteryCharger
formorethan2minutes
• InsertingitintotheAeroDRBatteryCharger2
formorethan6minutes
• Connectingawiredcableformorethan6
minutes
TheAeroDRDetectordoesnot
shutdown.
Powerdoesnotgooffevenwhenthepower
switchispressedfor5secondsorlonger.
Itisnotpossibletoshutdownwhileexposing.
Shutdownafterexposureisended.
ThestatusLED(green)islit,
andthebusy/errorLED(orange)
ashesrapidly.
("Ready"isnotdisplayedonthe
imageprocessingcontroller)
Systemerrorisoccurring.
Ifabusy/errorLED(orange)continuestoash
after10minuteshaselapsed,shutdownthe
AeroDRDetector.
Or,when"Ready"isnotdisplayedontheim-
ageprocessingcontroller,restartit.
WhentheAeroDRDetectoris
placedonasmoothsurface,the
AeroDRDetectorisnotstablyat-
tachedtothesmoothsurface.
AeroDRDetectoriswarped.
IftheAeroDRDetectorisstillwarpedeven
afterthedetectorisplacedonasmooth
surface,contactKonicaMinoltatechnical
representatives.
Theprotectivecoverisdeformed.
ContactKonicaMinoltatechnical
representatives.
TheAeroDRDetectorwillnotgo
intothewallstandortable.
AeroDRDetectoriswarped.
IftheAeroDRDetectorisstillwarpedeven
afterthedetectorisplacedonasmooth
surface,contactKonicaMinoltatechnical
representatives.
Theprotectivecoverisdeformed.
ContactKonicaMinoltatechnical
representatives.
TheAeroDRDetectorcannot
beinsertedintheAeroDR
BatteryCharger/AeroDRBattery
Charger2.
Theprotectivecoverisdeformed.
Foreignmaterialisinthewiredconnection
connectoroftheAeroDRDetector.
Referto"6.1.2Cleaning"andremovethe
foreignmaterial.
Foreignmaterialisinthebottomofthe
AeroDRBatteryCharger/AeroDRBattery
Charger2.
ContactKonicaMinoltatechnical
representatives.
TheAeroDRI/FCablecannot
beconnectedtotheAeroDR
Detector.
Partofthewiredconnectionconnectorofthe
AeroDRDetectorisdeformed.
ThespringconnectoroftheAeroDRI/FCable
isdeformed.