User's Manual

Table Of Contents
3.2OperationofAeroDRDetector
53
Chapter 3
Exposure switch
• Iftheexposureswitchremainshelddown,itmay
resultinanafterimagebecomingvisible.Once
exposureisnished,releasetheexposureswitch
immediately.
Grid
• Usethefollowinggridwhenexposing.
Griddensity
40lp/cmor
34lp/cm
Gridratio Variety
Convergence
distance
Variety
Angleerror
40lp/cm:1.0°orless
34lp/cm:1.7°orless
IMPORTANT
• ForanAeroDR1417Swithoutthe"1417S"identica-
tion,useagridwithagriddensityof34lp/cmandan
angleerrorof0.5°orless.
1417S
AeroDR P-12
( )
Identication
• Whenthe"AeroSyncmode"isused,useagridwitha
gridratioof8orless.
• Acappedgridisrecommendedwhenlayingthegrid
overtheAeroDRDetectoronthetabletoporforexpo-
surewiththeportableX-raydiagnosticdevice.
Exposure area
• ThesizelinesofAeroDRDetectorindicatethe
exposuresize.
• Duringexposure,placetheexposurebodypart
withinthewhitelinesonthefourcorners.
Whitelines
Whitelines
Sizelines
S-SRM connection
• WhenS-SRMconnectionisadopted,makesure
toperformexposureusingthehandswitchcon-
nectedtotheAeroDRInterfaceUnit2andAeroDR
GeneratorInterfaceUnit.
• Ifexposureisperformedusingthehandswitchon
theoperationpaneloftheX-raydevice,exposure
synchronizedtotheAeroDRSYSTEMisnotpos-
sible.
IMPORTANT
• Ifyourestartthe imageprocessing controller,also
restarttheAeroDR InterfaceUnit/AeroDR Interface
Unit2.