Universal Remote User Manual

Table Of Contents
87
Chapter
2
File
Menu
900
910
File – Import – Digitizer – Step Scan (When VIVID 900/910 is Selected)
11
Place the calibration chart on the rotat-
ing stage.
Position the chart with the black line facing
the VIVID digitizer, and place it on the rotat-
ing stage so that both entire side surfaces are
observed in the viewnder.
If “Chart” is not displayed in [Current angle],
click the left mouse button and select “Chart”
from the pull-down menu that appears.
M
emo
If it is not possible to perform calibration using calibra-
tion chart S, use calibration chart M.
12
Click the [OK] button.
The calibration chart will be scanned, and the
“Save the chart data ?” message dialog box
will appear.
To save the calibration chart data, click the
[OK] button.
M
emo
Saving the calibration chart data helps you reduce work
time when scanning different objects under the same
conditions.
Note
• Ifaccuratedataisrequired,itisrecommendedthat
calibrationbeperformedeachtime.
• Makesurethattheacquiredrangeimagelookslike
thegoodexamplegivenbelow.Ifitresemblesthebad
example,clickthe[Scan]buttonagaintore-scanthe
calibrationchart.
Calibration chart S
#ALIBRATION#HART-
Good example Bad example