User Guide

27
Appendix
Guidance and manufactures declaration - electromagnetic immunity
The Oxygen Saturation Monitor Model PULSOX-300 is intended for use in the electromagnetic
environment specified below. The customer or the user of the Model PULSOX-300 should assure
that it is used in such anenvironment.
Immunity test IEC 60601
Test level
Compliance
level
Electromagnetic environment -
guidance
Radiated RF
IEC61000-4-3
3 V/m
80MHz to 2.5GHz
3 Vrms
150kHz to 80MHz
Conducted RF
IEC61000-4-6
3 V/m
3 Vrms
NOTE 1 At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephone and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the PULSOX-300 is used exceeds
the applicable RF compliance level above, the PULSOX-300 should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the
PULSOX-300.
b
Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.
Portable and mobile RF communications
equipment should be used no closer to
any part of the PULSOX-300, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
80MHz to 800MHz
800MHz to 2.5GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation distance
inmeters (m).
Field strength from fixed RF
transmitters, as determined by an
electromagnetic site survey
a
, should be
less than the compliance level in each
frequency range
b
.
Interference may occur in the vicinity
of equipment marked with the following
symbol:
Recommended separation distance
E󲎌1
E󲎌1