3�4 Registration and selection of the DR Detector
By registering to the image processing controller, the
DR Detector can be moved between exposure rooms.
IMPORTANT
When using ImagePilot as the image processing controller,
it is not possible to move the DR Detector between di󰀨erent
exposure rooms.
Once it is registered in a new exposure room, it will not be
usable in the previous exposure room. When returning to a
previous exposure room, perform the registration operation
again.
3�4�1 Registration with AeroDR
Battery Charger
IMPORTANT
AeroDR 1012HQ and AeroDR 3 1012HQ cannot be regis-
tered with the AeroDR Battery Charger.
When an AeroDR Battery Charger exclusively for charge is
used, conrm that the battery charger in the destination has
a "Charge and registration" label.
1
Make sure that all the devices in the desti-
nation exposure room are usable�
2
Insert the DR Detector into AeroDR Bat-
tery Charger in the destination exposure
room�
Registration process will start.
Exposure
side
3
Oncetheregistrationiscompleted,a
buzzerwillsoundontheAeroDRBattery
Charger,andtheLED(green)willchange
fromslowashingtolit.
LED (green)
4
ConrmthattheDRDetectoriconisdis-
played on the image processing control-
ler
HINT
Charging will also start when the DR Detector (AeroDR
1417HQ, AeroDR 1417S, AeroDR 1717HQ, AeroDR 2
1417HQ) is inserted into the AeroDR Battery Charger, and
the AeroDR Battery Charger LED (blue) will light.
AeroDR 2 1417S, AeroDR 3 1417HD, AeroDR 3 1717HD,
AeroDR 3 1417HD2, and AeroDR 3 1717HD2 can be regis-
tered with the AeroDR Battery Charger.